TEE Transducer Care 

WARNINGS

Be sure to use the appropriate Personal Protective Equipment (PPE) whenever necessary. 

DO NOT

 skip any steps or abbreviate the cleaning process in any way.

Cautions

DO NOT

 allow bodily fluids to dry on the transducer. Always clean the transducer immediately after use. 

DO NOT

 bend the endoscopic shaft smaller than a 20 cm (8 inch) curve. 

DO NOT

 use unapproved cleaning agents such as alcohol or bleach (for example: Sani-Cloth) as these can damage the 

transducer and void the warranty. For more information about approved cleaners and disinfectants, see 

www.sonosite.com/

support/cleaners-disinfectants

DO NOT

 spray or immerse the controls, the cord, or the connector with any fluid. Be careful to prevent moisture from getting 

on the connector.

DO NOT

 soak the transducer longer than recommended by the chemical manufacturer. Follow the manufacturer’s 

recommendations for concentration, temperature, and duration. 

S

TEP

 1

 

ALWAYS

 use a bite guard. 

Remove any dentures before placing the 

bite guard in the patient’s mouth. 

S

TEP

 2

 

DO NOT

 

allow bodily fluids to 

dry on the transducer. 

Remove all visible 

biological material by 

wiping the entire 

transducer, 

except

 the connector, with a 

cloth or wipe moistened with water. 

Connector

Cable

Controller

Endoscopic shaft Scan head

Can be submerged

Cannot be submerged

90 cm

Refer to the following diagram when caring for your transducer

S

TEP

 3

 

Place up to 

90 cm of the 

endoscopic shaft in an 

approved

 cleaner. 

Make sure the 

controller cannot fall 

into the solution. 

CHECK

 the expiration date 

of the cleaner, and follow 

ALL

 chemical 

manufacturer’s instructions for soak time and 

temperature.

S

TEP

 4 

During the 

soak: 

A Scrub

 the shaft for at 

least 

3 min

 using a soft 

brush or single-use 

endoscopic sponge 

moistened with cleaner. 

B Wipe

 the controller 

and cable with a clean, 

non-linting cloth or 

single-use endoscopic 

sponge moistened with 

cleaner. 

Remove

 

cleaner by wiping again 

with a fresh cloth or sponge moistened with 

clean water. 

S

TEP

 5

 

Rinse

 the 

shaft for at least 

3

 

min 

in a large volume of 

clean, lukewarm water 

(for example, 8 L). 

Inspect 

the transducer 

for any remaining biological material. 

S

TEP

 6

 

Dry

 the 

transducer with a 

clean, non-linting 

towel. 

Examine 

the 

transducer and cable 

for any cracks. 

DO 

NOT

 use the transducer if any damage is 

found.

 

Cleaning 

Preventing damage

For a list of approved cleaners and disinfectants, refer to the cleaners and disinfection 

tool available at 

www.sonosite.com/support/cleaners-disinfectants