Technical
Publications

2337741–100

Revision 5

Innova 2000 Cardiovascular Imaging
System
Pre–Installation Manual
pim

do not duplicate

Copyright

 2002, 2003, 2004 and 2005 by General Electric Co.

Summary of Contents for Innova 2000

Page 1: ...Technical Publications 2337741 100 Revision 5 Innova 2000 Cardiovascular Imaging System Pre Installation Manual pim do not duplicate Copyright 2002 2003 2004 and 2005 by General Electric Co...

Page 2: ...al standards ATENCION LOS APARATOS DE RAYOS X SON PELIGROSOS PARA EL PACIENTE Y EL MANIPULADOR CUANDO LAS NORMAS DE PROTECCION NO ESTAN OBSERVADAS Aunque este aparato est construido seg n las normas d...

Page 3: ...12 2 10 Injectors 12 3 INNOVA LC PRODUCT IDENTIFICATION 13 3 1 Innova LC positioner 13 3 2 Product Options 20 4 RESPONSIBILITIES OF THE PURCHASER CUSTOMER 21 5 PRE INSTALLATION PROCESS 23 CHAPTER 2 SY...

Page 4: ...URVES 74 3 MOUNTING REQUIREMENTS 79 3 1 Floor Loading and Recommended Mounting Methods 79 3 2 Positioner and Table Floor Mounting 80 3 3 Innova LC Positioner and Omega Table Floor Preparation Kits GEM...

Page 5: ...18 7 2 Pedestal Injector 118 7 3 Injector L F ILLUMINA 118 8 RECOMMENDED POWER DISTRIBUTION SYSTEM 119 8 1 Main functions 119 8 2 Typical PDB 120 8 3 Interlocks 121 9 EMERGENCY 121 10 MAC LAB SYSTEM E...

Page 6: ...n Manual REV 5 pim 2337741 100 6 5 PROCESS ORDER SELECT 167 5 1 Generality 167 5 2 XT Stationary Rails 167 5 3 Cable and Component Select 168 5 4 Pre Installation kit S18721PN BASIC 168 6 PRE INSTALLA...

Page 7: ...tra ner chez le technicien l op rateur ou le patient des blessures dues des dangers lectriques m caniques ou autres ATENCION NO SE DEBER DAR SERVICIO T CNICO AL EQUIPO SIN HABER CONSULTADO Y COMPRENDI...

Page 8: ...por em perigo a seguran a do t cnico operador ou paciente devido a choques el tricos mec nicos ou outros AVVERTENZA SI PROCEDA ALLA MANUTENZIONE DELL APPARECCHIATURA SOLO DOPO AVER CONSULTATO IL PRESE...

Page 9: ...THE PIM IS NOT INTENDED TO REFLECT ALL OF THEM 2 DESCRIPTION OF INNOVA 2000 SYSTEM An Innova 2000 system is used for cardiography cases with several options An Innova 2000 system is first offered in...

Page 10: ...panel MPPU5 Microgrid H V tank mounted on the Maxiray150 X Ray tube housing 2 4 Description of the X Ray head The Innova 2000 System uses a MAXIRAY 150 tube housing including an X Ray tube a SIEMENS...

Page 11: ...or optional 2 6 Description of the TV monitor Rad Shield suspensions The common type of this suspension is an XT inboard monitor bridge A monitor frame support receiving 3 CRT monitors 2x Hi Brite mon...

Page 12: ...e the frame is fastened to a boom As an option 3 monitors dual 21 Hi Brite monitors and an additional monitoring screen are supplied mounted on the TV frame support with tray The frame is fastened to...

Page 13: ...Name Plate 2243600 2243600 2 LCA H1C POSITIONER with HV cables 24 m 80 feet LCA H1C POSITIONER with HV cables 30 m 100 feet Positioner lower left side 2243594 POS TIT CABINET Left top side 2237456 3...

Page 14: ...Innova 2000 Cardiovascular Imaging System GE Medical Systems Pre Installation Manual REV 5 pim 2337741 100 14 Illustration 1 POSTIT CABINET...

Page 15: ...41 100 15 Illustration 2 INNOVA LC BASIC PRODUCT Innova LC Positioner Innova LC positioner including L arm Item 1 Pivot Item 2 C arc Item 3 motorized elevator Item 4 for the Revolution Digital Detecto...

Page 16: ...GE Medical Systems Pre Installation Manual REV 5 pim 2337741 100 16 Illustration 3 OMEGA PATIENT TABLE BASIC PRODUCT TABLE SIDE CONTROL OR TSSC WITH CONTOUR FILTER MULTIAXIS CONTROL HANDLE OR SMART H...

Page 17: ...System GE Medical Systems Pre Installation Manual REV 5 pim 2337741 100 17 Illustration 4 OMEGA PATIENT TABLE OMEGA PATIENT TABLE Illustration 5 X RAY TUBE CHILLER DIGITAL DETECTOR CHILLER X Ray tube...

Page 18: ...Innova 2000 Cardiovascular Imaging System GE Medical Systems Pre Installation Manual REV 5 pim 2337741 100 18 Illustration 6 ACAB CABINET...

Page 19: ...ular Imaging System GE Medical Systems Pre Installation Manual REV 5 pim 2337741 100 19 Illustration 7 INNOVA USER AREA 17 Hi Brite Image Monitor Color flat panel 15 keyboard and mouse Keypad Innova c...

Page 20: ...Innova 2000 Cardiovascular Imaging System GE Medical Systems Pre Installation Manual REV 5 pim 2337741 100 20 3 2 Product Options Illustration 8 PRODUCT OPTIONS...

Page 21: ...responsibility for the various items during the system installation process To help you in determining these responsibilities review the following checklists with the customer and assign responsibilit...

Page 22: ...Innova 2000 Cardiovascular Imaging System GE Medical Systems Pre Installation Manual REV 5 pim 2337741 100 22 Blank page...

Page 23: ...s in Chapters 5 6 and 7 of this manual They represent an important part of the pre installation process The checklists summarize the required preparations and allow to verify the proper completion of...

Page 24: ...Innova 2000 Cardiovascular Imaging System GE Medical Systems Pre Installation Manual REV 5 pim 2337741 100 24 Blank page...

Page 25: ...7 3 This document Optional Second MAC TSSC 2237459 3 This document Optional Second MAC Table 4 SYSTEM COMPATIBILITIES CROSS REFERENCE TABLE OMEGA TABLES SUB SYSTEM PRODUCT NAME MODEL NUMBER PRE INSTAL...

Page 26: ...pplicable Chiller COOLIX 2200 2210405 9 2137573 100 Low noise or High ambient Table 7 SYSTEM COMPATIBILITY CROSS REFERENCE TABLE INNOVA LC IMAGING AND X RAY CONTROL TABLE SUB SYSTEM PRODUCT NAME MODEL...

Page 27: ...m Inboard Monitor Bridge or 7 ft 9 in 2 4 m Inboard Monitor Bridge 2270678 46 182884G2 or 46 193992G2 2208876 100 B2057AE B2057AG Note 1 By S O I process the B2057JB sleeve selection can be ordered w...

Page 28: ...ontour Filter Table 11 SYSTEM COMPATIBILITIES CROSS REFERENCE TABLE WALL MOUNT MONITOR PRODUCT NAME MODEL NUMBER PRE INSTALLA TION DOCUMENT NUMBER CATALOG NUMBER Wall Mount for 17 or 21 43 or 53 cm 23...

Page 29: ...Innova 2000 Cardiovascular Imaging System GE Medical Systems Pre Installation Manual REV 5 pim 2337741 100 29 CHAPTER 3 PHYSICAL REQUIREMENTS OF ROOM...

Page 30: ...Innova 2000 Cardiovascular Imaging System GE Medical Systems Pre Installation Manual REV 5 pim 2337741 100 30 Blank page...

Page 31: ...38 C 40 C 70 C POSTIT Cabinet 10 90 10 80 5 C 35 C 40 C 70 C Omega Patient table 5 90 5 90 5 C 35 C 40 C 70 C Maxiray 150 V5 X Ray tube assy N A N A N A N A 5 C 40 C 20 C 80 C Innova Digital Detector...

Page 32: ...1 The Detector chiller HEAT DRY 1 is shipped within GEMS packaging Note 2 The detector should be stored at 10 40 C and less than or equal to 90 RH in the plastic wrapped shipping box This should incl...

Page 33: ...U HR HEAT OUTPUT WATTS PRODUCT OR COM PONENT STANDBY IN USE STANDBY IN USE Innova LC Positioner negligible 1166 negligible 342 POSTIT Cabinet 5600 11123 1640 3260 Omega Patient table negligible neglig...

Page 34: ...the specified electromagnetic environment The purchaser or user of the Innova 2000 system should assure that it is used in an electromagnetic environment as described below Emissions Test Compliance...

Page 35: ...nt Surge IEC 61000 4 5 1 kV line s to lines s 2 kV line s to earth 1 kV line s to lines s 2 kV line s to earth Mains power quality is that of a typical commercial and or hospital environment Voltage d...

Page 36: ...ty At that disturbance amplitude the system is fully operational At that disturbance amplitude the monitor image might present some slight flicker For image quality improve ment the monitor might be r...

Page 37: ...lumn where P is the power rating of the transmitter in watts W according to the transmitter manufacturer Note These are guidelines Actual conditions may vary 1 3 4 LIMITATIONS MANAGEMENT Adhering to t...

Page 38: ...rated AC power dis tribution panel separated power line and an X ray shielded room Subsystem accessories Power supply distribution All components accessories subsystems systems which are electrically...

Page 39: ...d is specified less than 1 Gauss in Examination room and in the Control Area Static field is specified less than 3 Gauss in the Technical Room Electrostaticdischarges environment recommendations In or...

Page 40: ...hospital wall that could disturb the Mobile Innova X Ray CT MR Other system behind the hospital wall that could be disturbed by the Mobile Innova nursery emergency room Any other equipment which may d...

Page 41: ...00 Cardiovascular Imaging System GE Medical Systems Pre Installation Manual REV 5 pim 2337741 100 41 1 8 Preference Cabinet location Illustration 10 PREFERRED SYSTEM CABINET LOCATION IN ALL CASES ACAB...

Page 42: ...Innova 2000 Cardiovascular Imaging System GE Medical Systems Pre Installation Manual REV 5 pim 2337741 100 42 Blank page...

Page 43: ...7 Planning Aids Door Height The minimum door height to accommodate Innova positioner cabinet on its dolly is 2 09 m 82 in If the height is limited to 2 m 79 in you will need a fall over cabinet With...

Page 44: ...2 hex nuts Floor Furnished through boltswill accommodate200 to 430 mm 8I to 17I thick floor FOR EXAMPLE Floor requirements when using provided floor anchors The maximum pullout force per provided anch...

Page 45: ...C POSITIONER OMEGA Table 4 bolts with a pull out of 5650 N 445 3 17 53 336 13 23 TO BOLT HOLES Isocenter 1242 6 1 5 48 92 0 06 COMPACT LONG 1125 7 1 5 44 32 0 06 LONG optional For alternative table bo...

Page 46: ...cables from the sharp edges of conduits it is necessary to place the cable conduit inside the table cable access opening but the height of the outcoming conduit plus bushing is limited to 1 2 in 12 mm...

Page 47: ...ELECTRICAL CABLE TROUGH OR CONDUIT 5 0 127mm 3 6 91mm WATER LINES THRU FLOOR FITTING FLOOR TABLE SIDE 4 0 102mm 9 or 225 mm 1 OPENING THRU FLOOR ABOVE GRADE I D Internal Diameter Suitable External di...

Page 48: ...urvature with a minimum radius of 150 mm 6 In any other cases i e flush floor no such curvature is allowed Mini radius 150 mm 6 inches Note In case of on grade cabling because of the minimum curvature...

Page 49: ...tionary rails to rigid vertical members hung from ceiling slab Securing a supplementary channel to the bottom of the vertical members and mounting the stationary rails to this channel can greatly redu...

Page 50: ...ation 17 Recommended locations for 3 TV monitor suspension and rad shield 2500 98 42 660 4 26 660 4 26 660 4 26 660 4 26 660 4 26 660 4 26 660 4 26 4724 185 98 4500 177 1 330 13 500 19 7 1422 55 98 88...

Page 51: ...mm 13 inches should be changed to adjust both beams to the new product support Illustration 18 CEILING REQUIREMENTS FOR COUNTERBALANCED RADIATION SHIED 2 STEEL BEAMS 3 M or 10 ft LENGTH MOUNTED TO ADD...

Page 52: ...the main power demand to the X Ray generator X Ray generator is feed by a 126 kVA three phase power line For Generator Power Supply Features see ref 60601 1 IEC Table 21 ELECTRICAL REQUIREMENTS System...

Page 53: ...to a modem will preferably be located close to the cabinet The phone outlet must be located less than 1 meter 3 feet from the ACAB cabinet A modem compliant to each country is supplied with the Innova...

Page 54: ...Innova 2000 Cardiovascular Imaging System GE Medical Systems Pre Installation Manual REV 5 pim 2337741 100 54 Blank page...

Page 55: ...OMEGA IV COMPACT PATIENT TABLE DIMENSIONS Illustration 20 59 OMEGA V LONG PATIENT TABLE DIMENSIONS Illustration 21 60 POSTIT CABINET DIMENSIONS Illustration 22 61 ACAB CABINET DIMENSIONS Illustration...

Page 56: ...IONER INTERFERENCE REGIONS Illustration 36 75 OMEGA IV COMPACT PATIENT TABLE INTERFERENCE REGIONS Illustration 37 76 OMEGA V LONG PATIENT TABLE INTERFERENCE REGIONS Illustration 38 77 INNOVA LC POSITI...

Page 57: ...m GE Medical Systems Pre Installation Manual REV 5 pim 2337741 100 57 Illustration 19 INNOVA LC POSITIONER DIMENSIONS 87 95 50 12 27 36 27 75 4 72 42 12 13 5 45 98 25 2 Center of gravity 3 07 24 25 2...

Page 58: ...2000 Cardiovascular Imaging System GE Medical Systems Pre Installation Manual REV 5 pim 2337741 100 58 Illustration 19 b INNOVA LC POSITIONER DIMENSIONS 87 95 3 07 42 12 14 76 All dimensions are in mm...

Page 59: ...118 0 12 1504 59 21 1395 54 92 OMEGA IV COMPACT PATIENT TABLE PIVOT POINT OMEGA IV COMPACT PATIENT TABLE PIVOT POINT 780 3 30 71 0 12 467 18 38 1080 3 42 52 0 12 2593 13 102 10 0 5 SIDE VIEW TOP VIEW...

Page 60: ...TABLE DIMENSIONS FRONT VIEW FOOT END SIDE VIEW TOP VIEW 1 080 3 42 52 12 30 71 0 12 780 3 55 43 0 236 1 408 6 LCAISOCENTER 1278 3 3 108 3 1 928 3 75 9 0 12 TABLE TOP 3 336 3 131 34 0 12 1007 10 PATIEN...

Page 61: ...vascular Imaging System GE Medical Systems Pre Installation Manual REV 5 pim 2337741 100 61 Illustration 22 POSTIT CABINET DIMENSIONS 230 9 06 885 34 85 460 18 11 Gravity Center 165 6 50 All dimension...

Page 62: ...NET DIMENSIONS 1490 58 66 74 2 91 WARNING A free space up from the floor is required in front of the cabinet to open the door When the door is closed a free space about 15 cm 6 is required in front of...

Page 63: ...Exact dimensions of the chiller depends on the model type Required floor space depends on ambient room temperatures When in doubt allow for maximum floor space 140 cm MINIMUM 55 in 165 cm MINIMUM 65 i...

Page 64: ...OR CHILLER DIMENSIONS 141 REF 5 55 351 13 82 283 11 14 20 0 79 40 1 57 20 0 79 311 12 24 4 6 35 Dia 0 25 203 7 99 DISPLAY RESERVOIR CAP 340 13 39 359 14 13 378 14 88 HANDLE RS232 ON OFF IEC 320 WATER...

Page 65: ...0 Cardiovascular Imaging System GE Medical Systems Pre Installation Manual REV 5 pim 2337741 100 65 Illustration 26 DETECTOR CHILLER CONFIGURATION ORIENTATION CLEAR AIR ZONE AIR FLOW BULKHEAD LOCATION...

Page 66: ...aging System GE Medical Systems Pre Installation Manual REV 5 pim 2337741 100 66 Illustration 27 INNOVA LC POSITIONER AND OMEGA IV COMPACT PATIENT TABLE RELATIVE POSITIONS SIDE VIEW SIDE VIEW 54 92 Al...

Page 67: ...em GE Medical Systems Pre Installation Manual REV 5 pim 2337741 100 67 Illustration 28 INNOVA LC POSITIONER AND OMEGA IV COMPACT PATIENT TABLE RELATIVE POSITIONS TOP VIEW TOP VIEW 54 92 14 ft 8 in 139...

Page 68: ...Cardiovascular Imaging System GE Medical Systems Pre Installation Manual REV 5 pim 2337741 100 68 Illustration 29 INNOVA LC POSITIONER AND OMEGA IV COMPACT PATIENT TABLE RELATIVE POSITIONS FRONT VIEW...

Page 69: ...ging System GE Medical Systems Pre Installation Manual REV 5 pim 2337741 100 69 Illustration 30 INNOVA LC POSITIONER AND OMEGA V LONG PATIENT TABLE RELATIVE POSITIONS SIDE VIEW SIDE VIEW All dimension...

Page 70: ...GE Medical Systems Pre Installation Manual REV 5 pim 2337741 100 70 Illustration 31 INNOVA LC POSITIONER AND OMEGA V LONG PATIENT TABLE RELATIVE POSITIONS TOP VIEW TOP VIEW All dimensions are in mm in...

Page 71: ...Innova 2000 Cardiovascular Imaging System GE Medical Systems Pre Installation Manual REV 5 pim 2337741 100 71 Illustration 32 DL KEYPAD AND FLAT PANEL DIMENSIONS All dimensions are in mm inches...

Page 72: ...Installation Manual REV 5 pim 2337741 100 72 Illustration 33 DL IMAGE MONITOR DIMENSIONS 325 2 12 80 394 6 15 53 183 5 7 22 259 9 10 23 323 6 12 74 456 17 95 275 10 82 126 4 96 364 14 33 324 12 75 406...

Page 73: ...Innova 2000 Cardiovascular Imaging System GE Medical Systems Pre Installation Manual REV 5 pim 2337741 100 73 Illustration 34 VCR VIDEO STATION OPTIONAL All dimensions are in mm inches...

Page 74: ...x the VCR The mounting holes accommodate M3 x 10 mm screws The holes are 0 4 inches 10 mm deep All dimensions are in mm inches 2 SWEPT VOLUME CURVES Refer to this section for details on the mechanical...

Page 75: ...lar Imaging System GE Medical Systems Pre Installation Manual REV 5 pim 2337741 100 75 Illustration 36 INNOVA LC POSITIONER INTERFERENCE REGIONS 88 3 90 61 1 24 SIDE VIEW FRONT VIEW TOP VIEW All dimen...

Page 76: ...ems Pre Installation Manual REV 5 pim 2337741 100 76 Illustration 37 OMEGA IV COMPACT PATIENT TABLE INTERFERENCE REGIONS 1007 10 39 64 0 4 All dimensions are in mm inches OMEGA IV COMPACT PATIENT TABL...

Page 77: ...IENT TABLE INTERFERENCE REGIONS SIDE VIEW TOP VIEW 1 080 3 42 52 12 30 71 0 12 780 3 55 43 0 236 1 408 6 LCAISOCENTER 1278 3 3 108 3 1 928 3 75 9 0 12 TABLE TOP 3 336 3 131 34 0 12 1007 10 PATIENT COV...

Page 78: ...Innova 2000 Cardiovascular Imaging System GE Medical Systems Pre Installation Manual REV 5 pim 2337741 100 78 Blank page...

Page 79: ...oner 700 1543 See Illustration 19 and Illustration 19 b Circle diameter 23 62 600 mm Recommended Through Bolts 12 Alternates On Grade 5 8 in Anchors 12 Above Grade 3 4 in Anchors 12 See Illustration 3...

Page 80: ...d with four supplied 3 4 19 mm diameter x 4 1 4 108 mm floor anchors or 3 4 10 x 20 grade 5 thru bolts for slab or pan construction 7 With 2 real time disks or 121 3 kg 325 lbs and one real time disk...

Page 81: ...dical Systems Pre Installation Manual REV 5 pim 2337741 100 81 Illustration 39 INNOVA LC POSITIONER FLOOR MOUNTING METHODS 2 35 11 2 2 11 11 2 11 2 11 2 11 11 8 11 8 2 11 2 2 11 11 2 2 11 11 2 11 2 2...

Page 82: ...System GE Medical Systems Pre Installation Manual REV 5 pim 2337741 100 82 14 14 33 33 14 33 33 33 33 14 33 14 14 33 33 14 33 14 33 14 14 33 33 14 14 33 33 0 57 90 89 4 7 57 8 62 9 144 7 14 57 33 0 57...

Page 83: ...BLE CONDUIT FOR ON GRADE FLOOR ANCHOR KIT Note The key hole has been redisigned to permit a 60 mm 2 5 in water conduit for a separate access for water pipes used with a MAXIRAY 150 Push 76 3 0 hole in...

Page 84: ...Innova 2000 Cardiovascular Imaging System GE Medical Systems Pre Installation Manual REV 5 pim 2337741 100 84 Illustration 41 INNER BASE PLATE FOR ABOVE GRADE FLOOR ANCHOR KIT...

Page 85: ...G TEMPLATE PREFERRED 127 6 4 5 0 25 1 148 1 5 45 2 0 06 1 125 7 1 5 44 32 0 06 473 18 62 445 3 17 53 355 6 6 4 14 0 0 25 2 058 6 4 81 02 0 25 262 3 6 4 10 33 0 25 195 6 4 7 68 0 25 R 20 0 8 119 5 4 7...

Page 86: ...stems Pre Installation Manual REV 5 pim 2337741 100 86 Illustration 43 INNOVA LC OMEGA IV COMPACT OR OMEGA V LONG TABLE SYSTEM FLOOR MOUNTING TEMPLATE On Innova 2000 use only hole patern A or B See Il...

Page 87: ...ough Bolts 2285646 Assembly Separation Select 2285050 Base Plate Assembly 2285048 All GE supplied vascular system floor preparation are contained in catalog There are some additional gantry table moun...

Page 88: ...1 98 54 9 2 8 06604 6 22 9 95 65809054 7 522 57 9 2 6580905404 6 4 8 62 9 08 3 970 9 7 8 7 8 09 0 3 7 8 2 70 49 6 Doc 2230112 100 8 2 7 497 8 62 9 4 2 6 Doc 2229297 100 89 3 8 62 9 4 3 2 7 6 7 9054 6...

Page 89: ...6 418 5 7 2 7 4 98 98 1 3 8 4 46 418 5 7 2 7 4 6498 77 0 4 6 468 6 6 1 43 9 6 7 22 2 0 3 8 5 7 5 3 11 4 2 0 8 7 3 437869 8 3 47 8 43 6 6498 2 4 1 44 4 1 22 2 8 6 5 2286398 On grade Floor anchor 1446 3...

Page 90: ...8 5 66 3 9 C 95 37 6 5A 7 0 9 37 8 37 8 3 387 4 A 285 8 03 5 3 5 87 3 80 87 3 80 43 5 3 5 7 87 3 9 37 2 3B 55 387 2285053 Through bolts with insert 85 2 8 12 85 85 08 8 3 387 7 85 08 5 9 2 8 12 85 68...

Page 91: ...285054 Templates 2285056 A C N Templates 0 0 0 4 0 2186109 EP Template 4 0 Note 1 Old Kit B5079 KS is replaced by 2285048 and B5079 KT is split into 2285056 and 2186109 2 The kit 2285632 Floor plate B...

Page 92: ...IGHT DIMENSIONS LOAD BEARING AREA WEIGHT OCCUPIED MOUNTING OR COMPO NENT WEIGHT kg lbs LENGTH BEARING AREA INCHES MM OCCUPIED AREA MOUNTING METHOD 3 5 Wall Loading and Recommended Mounting Methods See...

Page 93: ...Innova 2000 Cardiovascular Imaging System GE Medical Systems Pre Installation Manual REV 5 pim 2337741 100 93 CHAPTER 5 ROOM LAYOUTS...

Page 94: ...Innova 2000 Cardiovascular Imaging System GE Medical Systems Pre Installation Manual REV 5 pim 2337741 100 94 Blank page...

Page 95: ...e unimpeded conduct of CPR Cardiac Pulmonary Resuscitation Clinicians at the patient table must be able to communicate with assistants in the control area There must be an unrestricted view of the vid...

Page 96: ...Innova 2000 Cardiovascular Imaging System GE Medical Systems Pre Installation Manual REV 5 pim 2337741 100 96 Blank page...

Page 97: ...ng Requirement for lighting concern the following general light technique characteristics Illuminator level Lighting distribution Preventing the operator from being dazzled by the light by direct ligh...

Page 98: ...ging System GE Medical Systems Pre Installation Manual REV 5 pim 2337741 100 98 Illustration 45 RECOMMENDED ROOM LAYOUT FOR INNOVA 2000 SYSTEM 6000 19 ft 8 in 0 25 10 1100 3 ft 7 in 6000 19 ft 8 in 97...

Page 99: ...ng screen 15 Injector on Pedestal 16 VCR 17 Bookcase or bookshelves required for Technical Publications and special parts and tools 18 GEMnet CRS Station Note The phone outlet must be located less tha...

Page 100: ...ging System GE Medical Systems Pre Installation Manual REV 5 pim 2337741 100 100 Illustration 46 RECOMMENDED ROOM LAYOUT FOR INNOVA 2000 SYSTEM TV Monitor suspension shown on this illustration is not...

Page 101: ...ing screen 15 Injector on Pedestal 16 VCR 17 Bookcase or bookshelves required for Technical Publications and special parts and tools 18 GEMnet CRS Station Note The phone outlet must be located less th...

Page 102: ...ing System GE Medical Systems Pre Installation Manual REV 5 pim 2337741 100 102 Illustration 47 RECOMMENDED ROOM LAYOUT FOR INNOVA 2000 SYSTEM 6000 19 ft 8 in 0 25 10 1100 3 ft 7 in 5988 19 ft 65 in 9...

Page 103: ...g screen 15 Injector on Pedestal 16 VCR 17 Bookcase or bookshelves required for Technical Publications and special parts and tools 18 GEMnet CRS Station Note The phone outlet must be located less than...

Page 104: ...ing System GE Medical Systems Pre Installation Manual REV 5 pim 2337741 100 104 Illustration 48 RECOMMENDED ROOM LAYOUT FOR INNOVA 2000 SYSTEM 6000 19 ft 8 in 0 25 10 1100 3 ft 7 in 5988 19 ft 65 in 9...

Page 105: ...g screen 15 Injector on Pedestal 16 VCR 17 Bookcase or bookshelves required for Technical Publications and special parts and tools 18 GEMnet CRS Station Note The phone outlet must be located less than...

Page 106: ...GE Medical Systems Pre Installation Manual REV 5 pim 2337741 100 106 Illustration 49 RECOMMENDED ROOM LIGHTING FOR INNOVA 2000 SYSTEM ZONE 1 ZONE 2 ZONE 3 6000 19 ft 8 in 0 25 10 1100 3 ft 7 in 6000 1...

Page 107: ...or bookshelves required for Technical Publications and special parts and tools 18 GEMnet CRS Station Note The phone outlet must be located less than 1 meter 3 feet from the ACAB Cabinet 3 In Examinat...

Page 108: ...a 2000 Cardiovascular Imaging System GE Medical Systems Pre Installation Manual REV 5 pim 2337741 100 108 Illustration 50 INNOVA 2000 SYSTEM WITH DIGITAL MOBILE 9800 Refer to Illustration 45 for the l...

Page 109: ...a 2000 Cardiovascular Imaging System GE Medical Systems Pre Installation Manual REV 5 pim 2337741 100 109 Illustration 51 INNOVA 2000 SYSTEM WITH DIGITAL MOBILE 9800 Refer to Illustration 46 for the l...

Page 110: ...Innova 2000 Cardiovascular Imaging System GE Medical Systems Pre Installation Manual REV 5 pim 2337741 100 110 Blank page...

Page 111: ...Innova 2000 Cardiovascular Imaging System GE Medical Systems Pre Installation Manual REV 5 pim 2337741 100 111 CHAPTER 6 ELECTRICAL CONNECTIONS...

Page 112: ...Innova 2000 Cardiovascular Imaging System GE Medical Systems Pre Installation Manual REV 5 pim 2337741 100 112 Blank page...

Page 113: ...n provides the shortest cable path 1 2 Floor Duct Floor duct has advantages when used with a single room or two adjacent rooms Floor duct combines a neat functional appearance with accessibility and r...

Page 114: ...Innova 2000 Cardiovascular Imaging System GE Medical Systems Pre Installation Manual REV 5 pim 2337741 100 114 2 POWER DISTRIBUTION Illustration 52 AC INTERCONNECT...

Page 115: ...irement is not in contradiction with local regulation In case of conflict follow local regulation No other measures employed for radiation protection should cause the interruption of irradiation and a...

Page 116: ...Innova 2000 Cardiovascular Imaging System GE Medical Systems Pre Installation Manual REV 5 pim 2337741 100 116 Illustration 53 GROUNDING INTERCONNECT...

Page 117: ...hield of any shielded cable coming from the distribution cannot replace the ground wire Reference For specific Vascular system grounding maps and connection details refer to the MisMap and mis chart l...

Page 118: ...it to a wall outlet fed by the main disconnect room device 7 2 Pedestal Injector A power cord is supplied with the injector In all cases 110 Vac or 230 Vac it will be connected to a wall outlet near...

Page 119: ...y CM R1 and R Main disconnect device by DMD1 and DMD2 Low voltage created by TR1 applied on remote L1 and L2 Separated injection circuit by TR2 and BR L is mandatory for French standards Protect the R...

Page 120: ...Innova 2000 Cardiovascular Imaging System GE Medical Systems Pre Installation Manual REV 5 pim 2337741 100 120 8 2 Typical PDB WITH AUXILIARY CONTACT...

Page 121: ...al contingency An interconnect diagram is defined as follows Hospital main vault UPS 300 kVA PDB Innova 2000 System In cooperation with GE Digital Energy a new UPS is under validation In all cases con...

Page 122: ...Innova 2000 Cardiovascular Imaging System GE Medical Systems Pre Installation Manual REV 5 pim 2337741 100 122 Blank page...

Page 123: ...dical Systems Pre Installation Manual REV 5 pim 2337741 100 123 11 PHYSICAL RUNS 11 1 Synoptic Illustration 54 PHYSICAL RUN SYNOPTIC DISTANCES EXPRESSED IN TOTAL LENGTH OPTIONS DISTANCES EXPRESSED IN...

Page 124: ...tion Manual REV 5 pim 2337741 100 124 11 2 System Core Matrix All lengths of cable are in useable meter when you look at group level or in meters connector to connector when you look at the cable leve...

Page 125: ...Innova 2000 Cardiovascular Imaging System GE Medical Systems Pre Installation Manual REV 5 pim 2337741 100 125 Illustration 56 CABLE GROUP 2 FROM TECHNICAL AREA TO CONTROL ROOM...

Page 126: ...on 57 CABLE GROUP 3 FROM TECHNICAL AREA TO TECHNICAL AREA This group enables to split the cabinets in 2 areas one for VAMP cabinet and one for POSTIT and ACAB cabinets These 2 areas can be either 3 or...

Page 127: ...337741 100 127 Illustration 58 CABLE GROUP 4 WITH STANDARD SUSPENSION Suspension are always pre cabled with 24 meter cables of connector to connector Monitor options includes their 24 meters length fo...

Page 128: ...Innova 2000 Cardiovascular Imaging System GE Medical Systems Pre Installation Manual REV 5 pim 2337741 100 128 Illustration 59 CABLE GROUP 4 WITH LCD SUSPENSION...

Page 129: ...enudage cosse ronde 12 27509A 2231647 27510A 2261135 WATER LINE 27511A 2261135 WATER LINE 27512A 2264440 OPTIC FIBER 27513A 2250247 2464 300 7 CEI 320 Fiche HBL4570C 38 6 27514A 2230590 2789 30 7 3 Su...

Page 130: ...ses CEI320 Prise HBL4570C Cosse ronde 39 27491A 2258467 2789 30 10 9 SCSI 68 pts 64 8 27492A 2264834 1354 27493A 2258471 Categorie 5 Low Volt age SCSI 5 5 RJ45 14 1 27494A 2258472 2464 300 6 SubD 9 pt...

Page 131: ...1 27455A 2249475 2464 300 7 HE15 Prise secteur HUB BELL 38 6 27465A 2258192 1019 600 9 1 Cosses rondes 12 27467A 2258191 1019 600 9 1 Pre denudage Cosses rondes 12 27517A 2258294 1019 600 7 3 Pre denu...

Page 132: ...300 16 CEE22 cosse ronde BNC50 BNC50 cosse ronde Fiche HBL4570C 38 6 Group n 4 3 From ACAB POSTIT to Monitors Standard suspension 27502A 2269591 2343 300 16 CEE22 cosse ronde BNC50 BNC50 cosse ronde F...

Page 133: ...e diam eter mm Max Voltage carried V Voltage rating V UL Style Cable Assembly MIS number 27993 2385196 300 Power 28042 2385212 2 2789 30 SubD 9pts 28043 2385213 2 2789 30 SubD 9pts 28175 2385197 300 P...

Page 134: ...Innova 2000 Cardiovascular Imaging System GE Medical Systems Pre Installation Manual REV 5 pim 2337741 100 134 Blank page...

Page 135: ...GE Medical Systems Pre Installation Manual REV 5 pim 2337741 100 135 JOB CARD PIST044 007 System facility power Time 0 h 15 min Personnel 1 field engineer 1 1 of 8 1 SUPPLIES None 2 TOOLS None 3 SAFE...

Page 136: ...and for any other medical devices powered in the examination room refer to Illustration 2 in the Atlas Cabinet for Innova 2000 subsystems All room outlets and emergency power sources in the room shal...

Page 137: ...7 System facility power 3 3 of 8 Illustration 1 Room ground point description 07 2 0 5 402 0 4 02 02 0 42 402 511 2 07 205 2 8 07 2 0 42 402 511 53 2 33 02 26 30 4 20 0 5 4 0 4 00 3 0 4 02 08 4 3 49 4...

Page 138: ...ce shall do a physical walk through to the facility to insure that the following are correct 1 The ground wires are the same size as the power feeder or AWG 1 0 whichever is larger 2 The grounds at ju...

Page 139: ...t all customer contractor supplied equipment se definitionin NEC article 100 and wiring conductors required have been installed will provide power as specified in GE Medical System preinstalltion draw...

Page 140: ...00 Cardiovascular Imaging System GE Medical Systems Pre Installation Manual REV 5 pim 2337741 100 140 JOB CARD PIST044 007 System facility power 6 6 of 8 Illustration 2 Ground connection at distributi...

Page 141: ...Innova 2000 Cardiovascular Imaging System GE Medical Systems Pre Installation Manual REV 5 pim 2337741 100 141 JOB CARD PIST044 007 System facility power 7 7 of 8 Illustration 3 Wire impedance test...

Page 142: ...nova 2000 Cardiovascular Imaging System GE Medical Systems Pre Installation Manual REV 5 pim 2337741 100 142 JOB CARD PIST044 007 System facility power 8 8 of 8 Illustration 4 Ground rod impedance tes...

Page 143: ...andard Tools screwdrivers drilling machine 3 SAFETY PRECAUTIONS Follow the procedure to unpack the spool Attach the power strip on wall to respect regulatory constraints 4 PREREQUISITES Gullets racewa...

Page 144: ...on two independent spools the group 3 3 is delivered in a separate box The pallet could be used as a tool to unspool the cables sets This tool 4 includes the axis 5 the two wooden forms 6 and the supp...

Page 145: ...ystems Pre Installation Manual REV 5 pim 2337741 100 145 JOB CARD PIST044 001 Innova 2000 System Interconnect 10 3 of 10 6 Use the two wooden forms as a slope to run the spool on the pallet 6 The cabl...

Page 146: ...nual REV 5 pim 2337741 100 146 JOB CARD PIST044 001 Innova 2000 System Interconnect 11 4 of 10 8 Put in place the two forms to build side support with anchor 8 Put the axis in the spool and raise the...

Page 147: ...e included in a long plastic bag 9 to limit the cables crossed This plastic bag can be kept on site in the gullet depending on local regulations 10 Pull out the cable set from the spool A specific pul...

Page 148: ...Note The power cable detector between Omega V and LC mischart 27722A is not part of spool but part of group 1 5 3 Run the cable Set Group 2 Group 2 links the Control booth area to the Technical area...

Page 149: ...g and the connector protections Run the cable set in the gullet from the 3 monitor suspension to the technical area Keep enough loop to enable the bridge movement Set of cables includes 7 cables for A...

Page 150: ...3 monitor suspension Cables are connected to the ACAB bulkhead 5 6 Run the cable Set Group 3 Group 3 links the technical area cabinets to the technical area cabinets The cable set is delivered in the...

Page 151: ...lation Manual REV 5 pim 2337741 100 151 JOB CARD PIST044 001 Innova 2000 System Interconnect 16 9 of 10 6 cables between POSTIT and ACAB cabinets 2 cables between ACAB cabinet and conditioner 1 cable...

Page 152: ...Innova 2000 Cardiovascular Imaging System GE Medical Systems Pre Installation Manual REV 5 pim 2337741 100 152 JOB CARD PIST044 001 Innova 2000 System Interconnect 17 10 of 10 Blank page...

Page 153: ...Innova 2000 Cardiovascular Imaging System GE Medical Systems Pre Installation Manual REV 5 pim 2337741 100 153 CHAPTER 7 ADDITIONAL PLANNING AIDS...

Page 154: ...Innova 2000 Cardiovascular Imaging System GE Medical Systems Pre Installation Manual REV 5 pim 2337741 100 154 Blank page...

Page 155: ...POSTIT Cabinet 82 2 08 29 0 74 52 1 32 Shipping Dolly See Illustration 4 Omega Table Base Assembly 49 1 24 38 0 96 84 2 2 14 1 290 585 On pallet See Illustration 5 Omega Table Top Assembly 9 0 22 137...

Page 156: ...PRODUCT OR COMPONENT METHOD OF SHIPMENT WEIGHT POUNDS KILOGRAMS DIMENSIONS INCHES METERS PRODUCT OR COMPONENT METHOD OF SHIPMENT WEIGHT POUNDS KILOGRAMS Depth Width Height TV monitor susp bridge 25 2...

Page 157: ...Pre Installation Manual REV 5 pim 2337741 100 157 2 DETAIL OF INNOVA SHIPPING INFORMATION Illustration 2 INNOVA LC POSITIONER GANTRY ON SHIPPING DOLLY 300 mm 12 with caster All dimensions are in mm in...

Page 158: ...Innova 2000 Cardiovascular Imaging System GE Medical Systems Pre Installation Manual REV 5 pim 2337741 100 158 Illustration 3 INNOVA LC POSITIONER AIR SHIPMENT...

Page 159: ...2337741 100 159 Illustration 4 INNOVA LC POSITIONER CABINET ON SHIPPING DOLLY This cabinet is shipped strapped to two dollies The cabinet requires a door opening of about 82 inches 208 cm high and 29...

Page 160: ...Innova 2000 Cardiovascular Imaging System GE Medical Systems Pre Installation Manual REV 5 pim 2337741 100 160 Illustration 5 OMEGA SHIPMENT...

Page 161: ...Innova 2000 Cardiovascular Imaging System GE Medical Systems Pre Installation Manual REV 5 pim 2337741 100 161 Illustration 6 OMEGA USING POSITIONER DOLLIES...

Page 162: ...ovascular Imaging System GE Medical Systems Pre Installation Manual REV 5 pim 2337741 100 162 Illustration 7 OTHER ELEMENTS PACKAGE Note All OEM parts are shipped inside there original boxes group as...

Page 163: ...2 inch Ratchet Wrench 2 Raise and Lower Positioner shipping dolly Wrench Spanner 46 176584P1 High Voltage Cable Installation Laptop Computer MS DOS Windows Positioner Configura tion and Calibration S...

Page 164: ...nova LC Service Engineer s Tool Case Fill in any additional tools or test equipment as required Ethernet adaptation kit for laptop 2128794 General use to be ordered before kit for laptop 2128794 order...

Page 165: ...a The Route Survey is a form on which site data are listed step 2 Illustration 8 ROUTE SURVEY MAP 3 4 1 2 6 5 7 8 9 10 Elevator Floor 2 Floor 1 Ramp Destination 4 2 Step Two Survey Data concerning the...

Page 166: ...al Systems Pre Installation Manual REV 5 pim 2337741 100 166 Table 31 ROUTE SURVEY TABLE ROUTE SURVEY Ref No Area loading dock doorway hallway turn elevator obstruction etc Limitations loading capacit...

Page 167: ...4 216 13 10 6 660 4 3 962 152 51 B0166JA 4 318 14 2 6 660 4 3 962 254 51 B0170JA 4 420 14 6 6 660 4 3 962 356 51 B0174JA 4 521 14 10 6 660 4 3 962 457 51 B0178JA 4 623 15 2 6 660 4 3 962 559 51 B0182...

Page 168: ...information is received on time the RAD is shifted to an other date 5 4 Pre Installation kit S18721PN BASIC During site preparation Installation specialists will need different spare parts The pre ins...

Page 169: ...4 3 505 3 607 3 708 3 810 3 912 4 013 4 115 4 216 4 318 4 420 4 521 4 623 4 724 4 826 4 928 5 029 5 131 5 232 5 334 5 436 5 537 5 639 5 791 660 4 660 4 660 4 660 4 660 4 660 4 660 4 660 4 406 51 51 40...

Page 170: ...2243723 5 18 m Group 3 2258106 3 2 m 2258106 4 9 m Group 4 4 Group 4 4 2266051 3 6 m Pre cabled Pre cabled 24 m 24 m 2266051 3 6 m 2266051 4 12 m B0134JA 134 in B0138JA 138 in B0142JA 142 in B0146JA...

Page 171: ...is great 8 is poor modulo 2 O O O O O How are Cat Numbers coming from structure other than INNOVA 2 is great 8 is poor modulo 2 O O O O O Operator Manual language agreed by customer O O Application s...

Page 172: ...O O O O O Cable access ready for the 4 cabinets and the 2 conditioners O O O O O Humidity temperature is AC completed Be sure that the temperature in the technical area will not exceed 32 C for 7 5kW...

Page 173: ...dality When prompted select the option for obtaining an IP address GEMSE Use the new mail form called INSFORM xls or INSFORM txt for obtaining an IP Address If you have questions or need clarification...

Page 174: ...Innova 2000 Cardiovascular Imaging System GE Medical Systems Pre Installation Manual REV 5 pim 2337741 100 174 Blank page...

Page 175: ...m 2276550 100 154 1 October 30 2002 Manual updated for Heroic 2 174 2 October 28 2003 Manual updated for Heroic 2 VCP 176 3 March 30 2004 ACAB audible noise level modification spr EURge01205 176 4 Mar...

Page 176: ...Innova 2000 Cardiovascular Imaging System GE Medical Systems Pre Installation Manual REV 5 pim 2337741 100 176 Blank page...