background image

USB

RH/TEMPERATURE/GPP

MINI DATA LOGGER

USER’S MANUAL

HT08

Please read this manual carefully and thoroughly before using this product.

HT08 - Manual-RR-FINAL_061914_awb  

Summary of Contents for HT08

Page 1: ...USB RH TEMPERATURE GPP MINI DATA LOGGER USER S MANUAL HT08 Please read this manual carefully and thoroughly before using this product HT08 Manual RR FINAL_061914_awb...

Page 2: ...7 12 Installing Replacing the Battery 7 Install Software and Drivers 7 9 Configure the Unit 9 12 Operating Instructions 13 16 Deploy the Unit 13 14 Stop Logging and Import Logs 14 Viewing Printing and...

Page 3: ...a job s drying phase Data logging can be started or stopped by pushing a button on the unit eliminating the need to bring a laptop to the job site After a data log has been captured and stored it can...

Page 4: ...uring Alerts provided by the low humidity alarm function of the HT can mitigate the risk of damaging high voltage static discharges KEY FEATURES One button starts stops data logging without a PC Plug...

Page 5: ...R PACK The HT08 comes in a blister pack that also contains One 1 2 AA battery Replacement batteries Part No BA10 are available from General www generaltools com A disc with software drivers for Window...

Page 6: ...rself with their names and functions before moving on to the Setup Instructions A Temperature and humidity sensor window B Battery compartment latch is on rear of unit C Two color green and red LEDs i...

Page 7: ...ery each of the two LEDs will flash red once and then green once Note If the battery is removed from a unit before it has been set up from software or while it is recording the unit must be reconfigur...

Page 8: ...t to install the software elsewhere click Browse and choose another location Within each window after selecting either the default location or another location click Next N to continue 7 When the Prep...

Page 9: ...le USB jack does not have enough vertical clearance to accommodate the HT08 T08 T08 without mechanically stressing its USB plug do not plug the HT directly into the jack You do not want the plug of th...

Page 10: ...o appear Click Yes to stop recording and import all data recorded to that point in time Click No to continue to Setup without importing recorded data meaning it will be lost Click Cancel to go back an...

Page 11: ...and low temperature alarm setpoints High and low humidity alarm setpoints To begin click on Set up the USB Data Logger in the window shown in the figure above Doing so will cause the following window...

Page 12: ...e meter to log unless you use it to log data for months rather than weeks or days Be aware that if you do manage to fill up the HT s memory by logging for a very long time at a very short sampling tim...

Page 13: ...dow Fig 1 Callout A is not covered The humidity and temperature values at the unit s location are representative of the room The HT with its protective cap on can be deployed either by laying it on a...

Page 14: ...uploading importing data logs is identical Just click on the middle selection Stop the USB Data Logger and download data to instantly execute both functions The data will be imported as a txt file A...

Page 15: ...ific period of time scale up and down the curves by dragging the slider at the bottom of the window to the left and right Hovering the cursor over a particular point of the curve displays all data rel...

Page 16: ...unit is ready to log data a quick press of the MODE button will cause both LEDs to flash green once If the HT08needs to be configured the two LEDs will alternately flash red and green twice If a quick...

Page 17: ...t Accuracy 3 from 20 to 80 RH 3 5 elsewhere RH Measurement Resolution 0 1 and Repeatability Internal Storage Capacity 16 300 pairs of RH Temp readings Temperature Response Time 5 30 seconds Humidity R...

Page 18: ...repair or replace this instrument if after examination the company determines it to be defective in material or workmanship This warranty does not apply to damages that General determines to be from...

Page 19: ...ntact our Customer Service to obtain an RGA Return Goods Authorization number before forwarding the unit via prepaid freight to the attention of our Service Center at this address General Tools Instru...

Page 20: ...2 431 6100 FAX 212 431 6499 TOLL FREE 800 697 8665 e mail sales generaltools com www generaltools com HT08 User s Manual Specifications subject to change without notice 2014 GENERAL TOOLS INSTRUMENTS...