background image

Tera TENS

Model GF-TT5

Operation Manual

Read this manual before operating 

your Tera TENS.

Save this manual for future use.

The most current version of this 

manual can be found online at 

www.grahamfield.com.

GF-TT5-INS-LAB-RevA18

Summary of Contents for GF-TT5

Page 1: ...del GF TT5 Operation Manual Read this manual before operating your Tera TENS Save this manual for future use The most current version of this manual can be found online at www grahamfield com GF TT5 INS LAB RevA18 ...

Page 2: ...EAR PANELS 11 ELECTRODE SNAP AND ADHESIVE SURFACES 12 4 SETUP 13 SKIN PREPARATION AND CARE 13 CHECK REPLACE THE BATTERY 14 BATTERY SAFETY 14 ELECTRODE PLACEMENT 15 ELECTRODE REMOVAL 16 5 OPERATION 17 TURN THE DEVICE ON 17 ADJUST THE TREATMENT INTENSITY 17 TURN THE DEVICE OFF 17 6 MAINTENANCE AND STORAGE 18 CARE AND CLEANING 18 STORAGE 19 7 TROUBLESHOOTING 20 8 SPECIFICATIONS AND SYMBOLS 21 GF TT5 ...

Page 3: ...al impulses are applied to nerves through electrode pads placed on the skin TENS is non invasive and does not use pharmaceuticals TENS uses tiny battery operated electrical impulses sent through the skin to nerves to modify your pain perception TENS does not cure any physiological problem it only helps control the pain TENS does not work for everyone however in most patients it is effective in red...

Page 4: ...apply stimulation over swollen infected inflamed areas or skin eruptions e g phlebitis thrombophlebitis varicose veins etc WARNING Avoid electrode placements that apply current to the carotid sinus region anterior neck or transcerebrally through the head WARNING Do not use this device over poorly enervated areas WARNING Do not use this device over an inguinal hernia WARNING Do not use this device ...

Page 5: ... that if not avoided could result in death or serious personal injury CAUTION Indicates a potential hazard or unsafe practice that if not avoided could result in minor or moderate personal injury s NOTICE Indicates a potential hazard or unsafe practice that if not avoided could result in product or property damage Info Provides application recommendations or other useful information to ensure that...

Page 6: ...the electrode on the front of the thorax marked by ribs and breastbone nor on the two large pectoral muscles The introduction of electrical current into the chest could cause rhythm disturbances to the heart which could cause death WARNING Do not apply stimulation in the presence of electronic monitoring equipment e g cardiac monitors ECG alarms which may not operate properly when electrical stimu...

Page 7: ...vice WARNING Discontinue use of this device and do not increase the intensity level if you feel discomfort during use WARNING The long term effects of electrical stimulation are unknown WARNING Electrical stimulation devices have no curative value WARNING Notice for California Customers California Proposition 65 WARNING This product can expose you to chemicals including DEHP which is known to the ...

Page 8: ...ersensitivity due to the electrical stimulation or electrical conductive medium silica gel CAUTION Use caution for patients with suspected or diagnosed heart problems if you have suspected or diagnosed heart disease you should follow precautions recommended by your healthcare professional CAUTION If you have suspected or diagnosed epilepsy you should follow precautions recommended by your healthca...

Page 9: ...ION If the stimulation levels are uncomfortable reduce the stimulation intensity to a comfortable level and contact your healthcare professional if problems persist NOTICE STATEMENTS s NOTICE Do not sterilize this device s NOTICE Do not use sharp objects such as pencil point or ballpoint pen to operate the buttons on the control panel s NOTICE Check the electrode connections before each use to ens...

Page 10: ...evA18 10 3 FEATURES COMPONENTS INCLUDED WITH DEVICE 2 1 3 4 ITEM DESCRIPTION QUANTITY 1 GF TT5 Tera TENS Unit 1 2 Electrode pad self adhesive 2 83 in x 2 24 in 72 mm x 57 mm 1 3 Battery 3V CR2032 1 4 Operation Manual 1 ...

Page 11: ...11 TENS UNIT FRONT AND REAR PANELS 1 front panel rear panel 3 4 5 4 6 2 1 ON increase intensity button 2 OFF decrease intensity button 3 Function display LED red LED 4 Contact plug snap 5 Battery compartment 6 Serial number ...

Page 12: ...LAB RevA18 12 ELECTRODE SNAP AND ADHESIVE SURFACES 1 2 5 3 3 4 snap surface adhesive surface 1 Electrode pad 2 Contact plug snap 3 Electrically conductive adhesive surface 4 Insulated surface 5 Protective backing ...

Page 13: ...kin where you will be placing the electrode with mild soap and water rinse soap off thoroughly and dry skin well 2 Clip excess hair with scissors do not shave stimulation area 3 Wipe the area with the skin preparation your healthcare professional has recommended and let it dry before applying the electrode 4 It may be helpful to apply skin lotion to electrode placement area after therapy when elec...

Page 14: ...ert a new 3V CR2032 Battery into the battery compartment 3 After installing the battery close the battery cover Ensure the battery cover is tightly closed BATTERY SAFETY Dispose of the battery in accordance with local guidelines and regulations WARNING If a battery is swallowed seek medical attention immediately WARNING Do not disassemble or short circuit the battery WARNING Do not place the batte...

Page 15: ...electrical stimulation and your needs may vary from the conventional settings suggested here If the initial results are not positive speak to your healthcare professional about alternative stimulation settings and or electrode placements Once an acceptable placement has been achieved record it for later use 1 Remove the protective backing from the electrode adhesive surface Do not throw away the p...

Page 16: ... for single patient use only s NOTICE Replace the electrode when it is no longer sticky 1 When removing the electrode remove by pulling in the direction of hair growth Lift the corner of the electrode and gently remove it from the skin 2 Apply the protective backing to the sticky side of the electrode for storage 3 It may be helpful to improve repeated application by spreading a few drops of cold ...

Page 17: ...r decreasing the intensity each click of the ON or OFF button increases or decreases the output intensity by one level which will show on the Function Display LED Info The device has ten levels of intensity Adjusting the intensity may cause a tingling sensation adjust the intensity to a level that is comfortable to you WARNING If the stimulation levels are uncomfortable or become uncomfortable red...

Page 18: ...or damage Replace worn items as required Clean the device after use with a soft clean slightly moistened cloth In case of more extreme soiling clean the device with a soft clean cloth slightly moistened with mild soapy water Electrode Care WARNING Use this device only with the electrode provided by GF Health Products Inc or suggested by your healthcare professional WARNING Use only the electrode p...

Page 19: ... in a cool dry well ventilated room and do not expose it to heat sunlight or moisture WARNING Do not dispose of the spent battery in household waste Dispose of the battery in accordance with local guidelines and regulations Battery operated devices are recyclable material and should not be disposed of in household waste Dispose of this device in accordance with local guidelines and regulations ...

Page 20: ...tly Insert battery again up Device not switched on Switch device on LED lights up weakly Battery low Replace battery Low output from device Electrode pad dirty Clean or replace electrode pad Device not applied properly Re apply device LED function normal but no power output Battery dead Replace battery Electrode pad dirty or short circuited Check electrode pad and replace if necessary Device switc...

Page 21: ...nditions Operating 50 F to 104 F 10 C to 40 C 40 90 Relative Humidity Storage 14 F to 140 F 10 C to 60 C 30 95 Relative Humidity Dimensions TENS Unit L x W x H 2 3 in x 1 2 in x 0 5 in 59 mm x 30 mm x 12 15 mm Electrode L x W 2 8 in x 2 2 in x 0 5 in 72 mm x 57 mm Weight including battery 0 5 oz 12 3 g KEY TO SYMBOLS The following symbols are used on this product s labels Consult instructions for ...

Page 22: ...ranty period or the expiration date OBTAINING WARRANTY SERVICE This limited warranty shall only apply to defects that are reported to the Distributor from whom the Customer purchased the product within the applicable warranty period If there is not a Distributor you must contact GF directly by calling 1 770 368 4700 sending a fax request to 1 770 368 2386 or by e mailing a request to cs grahamfiel...

Page 23: ...TIAL DAMAGES INCLUDING BUT NOT LIMITED TO DAMAGES FOR LOSS OF PROFITS OR INCOME LOSS OF USE DOWNTIME COVER OR EMPLOYEE OR INDEPENDENT CONTRACTOR WAGES PAYMENTS AND BENEFITS The warranties contained herein contain all the representations and warranties with respect to the subject matter of this document and supersede all prior negotiations agreements and understandings with respect thereto The reci...

Page 24: ... can be found on our website www grahamfield com 2018 GF Health Products Inc All Rights Reserved Graham Field and Grafco are trademarks of GF Health Products Inc GF Health Products Inc is an ISO 13485 Certified Company Manufactured for GF Health Products Inc One Graham Field Way Atlanta GA 30340 3140 Made in Taiwan ...

Reviews: