Operator Quick Guide

ORBISPHERE 410

Revision E - 24/04/2008

Summary of Contents for ORBISPHERE 410

Page 1: ...Operator Quick Guide ORBISPHERE 410 Revision E 24 04 2008...

Page 2: ...interface software is Windows CE based providing easy selection through menus All the measurement configuration calibration and standard service routines can be called by pressing buttons and menu bar...

Page 3: ...f information displayed depends on the gas being measured and the channel configuration The statistical view offers statistical data that matches with Total Quality management tools Statistics is a to...

Page 4: ...lved Gas unit The list of available units depends on unit type selected above Liquid When medium is liquid select water or a liquid with a different solubility if available Display resolution Maximum...

Page 5: ...ntration values are above High alarms or below Low alarms Set this to a minimum but enough to avoid alarms for non representative peaks beyond the set level Measurement Filter Configuration The filter...

Page 6: ...the buffer is full Adjust the parameters for recording and storing measured data Storage mode selection No storage Store once When volatile memory is full 1 000 positions measurement recording stops R...

Page 7: ...the Modify button Last calibration parameters are memorized so this step can be ignored if correct parameters are already set Modify Calibration Parameter Calibration mode 2 types available depending...

Page 8: ...saturated air to provide a known oxygen reference against which to calibrate Dry the sensor thoroughly before placing the sensor storage cap under tap water Shake off any excess water but leave a few...

Page 9: ...tion temperature barometric pressure current the date and time the user name and ID Barometric Pressure Calibration Upper box shows the barometric pressure measured by the instrument Using a precision...

Page 10: ...cing and maintenance For more information on this please refer to the corresponding sensor maintenance manual Language Check the language as required and restart the instrument to apply the change The...

Page 11: ...urement is frozen Alarm low low The gas concentration is below the Alarm LowLow limit Alarm low The gas concentration is below the Alarm Low limit Alarm high The gas concentration is above the Alarm H...

Page 12: ......