background image

LDU 179.1 Technical Manual rev. 1.10 - February 2019                                                                       Page 1 of 45

 

 

          

                                              

 

 
 
 
 
 

Load Cell Digitizing Unit 

 
 
 

Type LDU 179.1 

 

 

 

T

ECHNICAL 

M

ANUAL

 

 
 
 

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 

 
 
 
 
 
 
 
 
 

Firmware Version 179.181.v.1.10 or higher 

Hardware Version 179.101.5.v.1.01 

Document No. X179 Rev 1.1 EN 

 

 

 
 
 

Hauch & Bach ApS 

Femstykket 6 

DK-3540 Lynge 

Denmark 

www.haubac.com

 

Summary of Contents for LDU 179.1

Page 1: ...2019 Page 1 of 45 Load Cell Digitizing Unit Type LDU 179 1 TECHNICAL MANUAL Firmware Version 179 181 v 1 10 or higher Hardware Version 179 101 5 v 1 01 Document No X179 Rev 1 1 EN Hauch Bach ApS Femst...

Page 2: ...w 21 8 Commands Description 23 8 1 System Diagnosis Commands ID IV IS SR RS 23 8 1 1 ID Get Device Identity 23 8 1 2 IV Get Firmware Version 23 8 1 3 IS Get Device Status 23 8 1 4 SR Software Reset 23...

Page 3: ...mple Value continuously 34 8 7 4 SA Send Triggered Average Value automatically 34 8 7 5 SL Send Data String Average Gross and Status automatically 34 8 8 Commands for External I O Control IN IO and OM...

Page 4: ...13 1 RW Trigger Window for Re Trigger Function 43 8 13 2 TT Averaging Time for Re trigger Function 43 8 13 3 TS Stop Value for Re trigger Function 43 8 13 4 DT Short time Averaging Period 43 8 13 5 TW...

Page 5: ...Hauch Bach shall not be liable against any damages or problems arising from the use of any options or any consumable products other than those designated as Original Hauch Bach Products NOTICE The co...

Page 6: ...laws of the Member States relating to electromagnetic compatibility and repealing Directive 2004 108 EC Diese Erkl rungbescheinigt die bereinstimmung mit den genannten Richtlinien beinhaltetjedochkein...

Page 7: ...ls n 10 000 Minimum input voltage per VSI V 0 2 Measuring range FS mV V 3 3 Maximum resolution at FS approx incr 880000 Measuring rate Hz 9 4 to 1200 Digital filter cut off frequency 3dB Hz 0 25 to 18...

Page 8: ...his manual means Space _ and Enter CR LF Master PC SPS sends Slave LDU 179 1 responds Meaning ID D 1790 Identity of the device IV V 0107 Firmware version of the device GN N 1234 56 Net weight with alg...

Page 9: ...Load cell Pin no 0 Shield 1 Excitation 2 Sense 3 Signal 4 Signal 5 Sense 6 Excitation 7 Shield 8 NC 9 NC LDU 179 1 Pin Function Pin no 10 TxD RS232 TxD 11 RxD RS232 RxD 12 CANL CAN line low 13 CANH C...

Page 10: ...dbus mode and back to ASCII mode In Modbus mode this interface can run Modbus RTU protocol The procedure to switch between the modes and the Modbus RTU communication is described in the sepperate manu...

Page 11: ...is TDPO2 contains the module number the module status and the dosed weight In checkweigher applications the TPDO2 is used to send triggered measurements With RDPO1 frames you can send simple commands...

Page 12: ...average weight if it is a TPDO2 and tare value if it is a TPDO3 Then status follows as a 16 bit field with the following values defined 0001 Under range 0002 Over range 0008 Center zero 0010 No motio...

Page 13: ...the LDU in the PRE OPERATIONAL state When a SDO has been received by the controller no further communication takes place until the service has been acknowledged or a timeout occurs SDO s Are only ava...

Page 14: ...t mapping Abbreviations used in Tables ro read only rw read write wo write only read will not be regarded as an error but returns undefined results UI8 Unsigned 8 UI16 Unsigned 16 UI32 Unsigned 32 I32...

Page 15: ...NMT Master or the configuration tool 1014 0 COB ID Emergency Message UI32 ro 80H Node ID COB ID of the Emergency Object 1017 0 Heartbeat Time UI16 rw 0 Producer Heartbeat time in ms If index 1017h is...

Page 16: ...DO 1 1st Object Cmd Byte UI8 UI32 ro ro 2 20060308H Mapping parameters of the 1st Receive PDO Object is a bitwise command Bit7 Bit6 Bit5 Bit4 Bit3 Bit2 Bit1 Bit0 Cmd SnG SnN ST RT SZ RZ 1601 0 1 Entri...

Page 17: ...cation parameters of 4th Transmit PDO Determined using the CANopen minimum system ID assignment procedure Asynchronous communication Not used will not be transmitted 1A00 0 1 2 Number of mapped Entrie...

Page 18: ...11 12 13 14 15 16 17 18 19 20 21 22 23 Number of entries Dummy Dummy Dummy Filter setting Dummy Logic outputs Logic inputs Measure Time Filter Mode No motion Range No motion Time Output Mask Dummy St...

Page 19: ...AC protected Not used Not used Calibrate max 1 TAC protected Calibrate min TAC protected Not used Calibrate zero TAC protected Decimal point TAC protected Display step size TAC protect Multi range mul...

Page 20: ...oint 1 Set point 2 UI8 UI8 UI8 ro rw rw 2 0 0 Number of Set point parameters Set point 1 source Set point 2 source 2900 0 1 2 3 4 5 6 7 8 9 10 11 12 13 Number of entries Gross weight Net Weight Tare D...

Page 21: ...Frequency 0 8 30 GA Output Get Triggered Average Value None 33 41 GG Output Get Gross Value None 32 GI Retrieves an image file from the LDU s EEPROM None 40 GL Output Get Data String Average Gross Sta...

Page 22: ...ing Net Gross Status continuously SX Auto transmit Send ADC Sample Value continuously None 34 SZ Scale function System Zero Point None 31 TD Transmit delay For compatibility only 0 255 ms TE Trigger f...

Page 23: ...this request gives the firmware version of the device 8 1 3 IS Get Device Status SDO 2900 sub 0A Master PC SPS sends Slave LDU 179 1 responds IS S 067000 example The response to this request comprises...

Page 24: ...nt in multi range applications Permitted values are from 0 to 999999 Master PC SPS sends Slave LDU 179 1 responds Meaning CM1 M 030000 Request CM1 30000 d CE E 00017 example Request TAC counter CE17 C...

Page 25: ...aster PC SPS sends Slave LDU 179 1 responds Meaning DS S 00002 Request Step size 2 CE E 00017 example Request TAC counter CE17 CE17 OK Calibration sequence active DS50 OK Setup Step size 50 Legal for...

Page 26: ...1 Request ZT status CE E 00017 example Request TAC counter CE17 CE17 OK Calibration sequence active ZT0 OK Setup ZT Disable Zero tracking will be performed only on results less than ZT range at a rate...

Page 27: ...S sends Slave LDU 179 1 responds Meaning CE E 00017 example Request TAC counter CE17 CE17 OK Calibration sequence active ZI100 OK Setup Initial Zero range 100 d 8 2 14 TM Tare mode SDO 2300 sub 12 Thi...

Page 28: ...ve AZ_00500 OK New Zero point 0 0500 mV V Factory default 00 000d 0 0000mV V input signal 8 2 18 AG Absolute gain calibration eCal SDO 2300 sub 01 The command AG is used as absolute gain or measuring...

Page 29: ...to be stable or not The weight signal has to vary by less than NR divisions over the period of time NT to be considered stable Permitted values are from 0 to 65535 Master PC SPS sends Slave LDU 179 1...

Page 30: ...No filtering 1221 1 55 18 57 1221 2 122 8 78 1221 3 242 4 96 1221 4 322 3 104 1221 5 482 2 114 1221 6 963 1 132 1221 7 1923 0 5 149 1221 8 3847 0 25 164 1221 Output rate 1221 2UR samples s Prefilter...

Page 31: ...e required and the calibration zero differ by more than 2 of the CM value maximum allowable value See chapter 9 Used in Approved applications 8 5 1 SZ Set System Zero RPDO1 02 This command sets a new...

Page 32: ...d the weighing signal returns to gross mode Master PC SPS sends Slave LDU 179 1 responds Meaning RT OK Tare de activated Gross operation The LDU responds to the RT command with either OK or ERR If OK...

Page 33: ...lue of the sum of all ASCII values within the data string without the check sum itself 8 6 6 GA Get Triggered Average Value SDO 2900 sub 06 This command reads the measurement result of a measurement c...

Page 34: ...commands in this chapter n a 8 7 1 SG Send Gross Value continuously Master PC SPS sends Slave LDU 179 1 responds Meaning SG G 001 100 Gross value 1 100 d 8 7 2 SN Send Net Value continuously Master PC...

Page 35: ...he logic outputs is normally determined by the internal setpoints see section 10 9 2 and therefore setting the logic output status using the IO commands is not allowed Setting Master PC SPS sends Slav...

Page 36: ...eresis and Switching Action for a Setpoint H0 SDO 2700 sub 01 H1 SDO 2700 sub 02 The setpoint switching logic is defined by the numeric value and polarity of the hysteresis Examples of the switching a...

Page 37: ...e stored in the EEPROM using the SS command See chapter 8 11 3 8 9 4 HT Hold time for all Setpoints SDO 2500 sub 05 This command defines the hold time for the setpoint limit The signal has to exceed t...

Page 38: ...ings For Serial Channel and CAN Interface n a The command NS Interface Param New Value can display or set various communication parameters in the device The parameter Interface addresses the physical...

Page 39: ...x responds Meaning NS_0_0 D 1790 The device type is LDU 179 NS_0_1 B 115200 The serial channel baud rate is 115200 NS_0_2 A 000 The serial channel address is 0 NS_0_3 S 00000 The serial channel mode i...

Page 40: ...e command returns ERR and has no updating action unless it is preceded by the CE_XXXXX 8 11 2 WP Save the Setup Parameters SDO 2004 sub 03 With this command the settings of the Filter FL FM the No mot...

Page 41: ...MT 0 trigger function disabled time plot of a typical checkweigher cycle see below 8 12 3 GA Get Triggered Average Value SDO 2900 sub 06 This command reads the measurement result of a measurement cycl...

Page 42: ...up TL 1000 In the example a new measurement cycle would automatically start if the signal exceeds 1000 d e g 100 0 g trigger commands SD and TL Default setting TL 999999 trigger level disabled Note Al...

Page 43: ...d defines an averaging time for calculating the cycle average value If this time period has been elapsed the measurement cycle will be finished at the latest The setting TT 0 disables the re trigger f...

Page 44: ...aged tare value falls outside this window then the tare value will not be updated Default setting TW 0 automatic taring disabled Master PC SPS sends Slave LDU 179 1 responds Meaning TW T 00000 Request...

Page 45: ...s Usually a discussion with the Weighing Equipment Approvals Officers at the relevant National Weights Measures Office will then reveal the extent of any pattern testing that may be necessary to ensur...

Reviews: