Instruction Manual
HI 9147
Portable, Water-resistant
Dissolved Oxygen Meter
with
Galvanic Probe
and
Battery Recharging System
w w w . h a n n a i n s t . c o m
WARRANTY
HI 9147 meters are guaranteed for two years
against
defects in workmanship and materials when used for their
intended purpose and maintained according to instructions.
The probes are guaranteed for one year.
This warranty is limited to repair or replacement free of charge.
Damages due to accident, misuse, tampering or lack of pre-
scribed maintenance are not covered.
If service is required, contact the dealer from whom you
purchased the instrument. If under warranty, report the model
number, date of purchase, serial number and the nature of
the failure.
If the repair is not covered by the warranty, you will be
notified of the charges incurred.
If the instrument is to be returned to Hanna Instruments, first
obtain a Returned Goods Authorization Number from the
Customer Service department and then send it with shipment
costs prepaid. When shipping any instrument, make sure it is
properly packaged for complete protection.
Dear Customer,
Thank you for choosing a Hanna product.
Please read carefully this instruction manual before using the
meter. If you need additional technical information, do not
hesitate to e-mail us at
tech@hannainst.com
.
This instrument is in compliance with the CE directives.
PRELIMINARY EXAMINATION
Remove the instrument from the packing material and exam-
ine it carefully to make sure that no damage has occurred
during shipment. If there is any damage, notify your Dealer.
The meter is supplied complete with
•
D.O. probe, fixed:
HI 76409/4
with 4 m cable for
HI 9147-04
HI 76409/10
with 10 m cable for
HI 9147-10
•
Spare membranes (
HI 76409A/P
) with O-rings, 5 pcs.
•
HI 7042S
electrolyte solution (30 ml bottle)
•
HI 710016
shockproof rubber boot
• 9V Ni-MH rechargeable battery
• Screwdriver and instructions.
Note:
Conserve all packing material until the instrument has
been observed to function correctly. Any defective item
must be returned in its original packing.
1)
HI 76409/4
or
HI 76409/10
galvanic D.O. probe (fixed)
2) Liquid Crystal Display
3) BL key, turn the backlight ON (as long as it is pressed)
4) °C key, display temperature measurement
5) % key, display O
2
measurement in % of saturation
6) % trimmer, for calibration adjustment
7) SAL trimmer, for setting salinity compensation value
8) SAL key, display setting of salinity factor (active only in
mg/L measurement mode)
9) mg/L key, display O
2
measurement in mg/L (ppm)
10) OFF key, turn the instrument OFF
Note: To turn the instrument ON, press any of the
measurement keys (°C, %, mg/L)
Range
0.0 to 50.0 mg/L (ppm) O
2
0 to 600 % O
2
-5.0 to 50.0°C
Resolution
0.1 mg/L or 1% (O
2
) / 0.1°C
Accuracy
±1% of reading (O
2
) / ±0.2°C
Typical EMC Deviation
±2% of reading (O
2
) / ±1°C
Calibration
Manual, in saturated air
Temperature Compensation
Automatic, from -5 to 50°C (23 to 122°F)
Salinity Compensation
0 to 51 g/L (resolution 1 g/L)
Probe
(
fixed
)
HI 76409/4
with 4 m cable or
HI 76409/10
with 10 m cable
Battery Type
1 x 9V Ni-MH, rechargeable
Battery Charge Life
200 hours of use (BL off)
Battery Life
Approx. 5 years
Auto-off
After approx. 6 minutes
Environment
0 to 50°C (32 to 122°F);
max 95% RH non-condensing
Dimensions
164 x 76 x 45 mm
(6.5 x 3.0 x 1.8")
Weight
(with HI 76409/4 probe)
511 g (1.12 lb.)
ACCESSORIES
HI 76409/4*
Galvanic D.O. probe with built-in temperature
sensor and 4 m cable
HI 76409/10*
Galvanic D.O. probe with built-in temperature
sensor and 10 m cable
HI 76409-0
Protective sleeve for
HI 76409
probes series
HI 76409A/P
Membrane for
HI 76409
probes series, 5 pcs.
HI 7042S
Electrolyte solution, 30 ml bottle
HI 710015
Shockproof rubber boot, blue color
HI 710016
Shockproof rubber boot, orange color
HI 731326
Calibration screwdriver (20 pcs)
HI 710040
Battery recharger
*
To be substituted by authorized technical personnel only
GENERAL DESCRIPTION
Oxycheck
is a water-resistant Dissolved Oxygen meter with
manual calibration, automatic temperature compensation (ATC),
salinity compensation and direct probe.
The instrument is equipped with an inductive system for battery
recharge, with no external contact, to ensure a watertight seal.
Dissolved Oxygen is indicated in mg/L (ppm) or in % of saturation.
The temperature can be measured in the range from -5 to 50°C.
The Dissolved Oxygen readings are automatically compensated for
the temperature effects on the oxygen solubility and membrane
permeability. Moreover, the salinity compensation feature allows
determination of Dissolved Oxygen even in salty waters.
The meter is housed in a rugged water-resistant case for maximum
protection in the field as well as in the laboratory.
The D.O. probe is provided with a membrane covering the galvanic
sensors and a built-in thermistor for temperature measurement and
compensation. The thin permeable membrane isolates the sensor
elements from the testing solution, but allows oxygen to enter.
Oxygen that passes through the membrane causes a current flow,
from which the oxygen concentration is determined.
Two models are available:
•
HI 9147-04
with
HI 76409/4
probe (4 m cable)
•
HI 9147-10
with
HI 76409/10
probe (10 m cable)
IST9147R2 05/05
Hanna Instruments reserves the right to modify the design, construction
and appearance of its products without advance notice.
FUNCTIONAL DESCRIPTION
SPECIFICATIONS