background image

HOBO® MX CO

2

 Logger (MX1102A) Manual 

22504-F 

The HOBO MX CO

data logger records carbon dioxide, temperature, and relative humidity (RH) 

data in indoor environments using non-dispersive infrared (NDIR) self-calibrating CO

2

 sensor 

technology and integrated temperature and RH sensors. This Bluetooth® Low Energy-enabled 
logger is designed for wireless communication and also supports a USB connection. Using the 
HOBOconnect® app on your phone, tablet, or computer, or HOBOware® software on your 
computer, you can easily configure the logger, read it out, and view plotted data. The logger can 
calculate minimum, maximum, average, and standard deviation statistics and can be configured to 
trip audible or visual alarms at thresholds you specify. In addition, it supports burst logging in 
which data is logged at a different interval when sensor readings are above or below certain 
limits. This logger also has a built-in LCD screen to display the current CO

2

 level, temperature, RH, 

logging status, battery use, memory consumption, and more. 

Specifications 

Temperature Sensor 

 

Range 

0° to 50°C (32° to 122°F) 

Accuracy 

±0.21°C from 0° to 50°C (±0.38°F from 32° to 122°F), see Plot A 

Resolution 

0.024°C at 25°C (0.04°F at 77°F), see Plot A 

Drift 

<0.1°C (0.18°F) per year 

RH Sensor* 

 

Range 

1% to 90% RH (non-condensing) 

Accuracy 

±2% from 20% to 80% typical to a maximum of ±4.5% including 
hysteresis at 25°C (77°F); below 20% and above 80% ±6% typical 

Resolution 

0.01% 

Drift 

<1% per year typical 

CO

2

 Sensor 

 

Range 

0 to 5,000 ppm 

Accuracy 

±50 ppm ±5% of reading at 25°C (77°F), less than 90% RH (non-
condensing) and 1,013 mbar 

Warm-up Time 

15 seconds 

Calibration 

Auto or manual to 400 ppm 

Non-linearity 

<1% of FS 

Pressure Dependence 

0.13% of reading per mm Hg (corrected via user input for 
elevation/altitude) 

Operating Pressure Range 

950 to 1,050 mbar (use Altitude Compensation for outside of this 
range) 

Compensated Pressure 
Range 

-305 to 5,486 m (-1,000 to 18,000 ft) 

Sensing Method 

Non-dispersive infrared (NDIR) absorption 

Response Time 

 

Temperature 

12

 

minutes to 90% in airflow of 1 m/s (2.2 mph) 

RH 

1 minute to 90% in airflow of 1 m/s (2.2 mph) 

CO

2

 

1 minute to 90% in airflow of 1 m/s (2.2 mph) 

Logger 

 

Radio Power 

1 mW (0 dBm) 

Transmission Range 

Approximately 30.5 m (100 ft) line-of-sight 

Wireless Data Standard 

Bluetooth Low Energy (Bluetooth Smart) 

Logger Operating Range 

0° to 50°C (32° to 122°F); 0 to 95% RH (non-condensing) 

*Per RH sensor manufacturer data sheet 
Note: The HOBO U-Shuttle (U-DT-1) is not compatible with this logger. 

 

 

HOBO MX CO

2

 Logger 

MX1102A 

Included Items: 

 

Four AA 1.5 V alkaline 
batteries 

Required Items: 

 

Mobile device with 
HOBOconnect app, 
Bluetooth, OS, iPadOS®, or 
Android™, or a Windows 
computer with a native BLE 
adapter or supported BLE 
dongle 

OR 

 

HOBOware 3.7.3 or later 
and USB cable 

Accessories: 

 

Mounting kit with 
mounting brackets, screws, 
tie wraps, and Command™ 
strip 

 

 

www.

.com

information@itm.com

1.800.561.8187

Summary of Contents for MX1102A

Page 1: ...f 4 5 including hysteresis at 25 C 77 F below 20 and above 80 6 typical Resolution 0 01 Drift 1 per year typical CO2 Sensor Range 0 to 5 000 ppm Accuracy 50 ppm 5 of reading at 25 C 77 F less than 90 RH non condensing and 1 013 mbar Warm up Time 15 seconds Calibration Auto or manual to 400 ppm Non linearity 1 of FS Pressure Dependence 0 13 of reading per mm Hg corrected via user input for elevatio...

Page 2: ...e reduced by remaining connected excessive readouts audible alarms and paging Visual audible alarms and other events can have a marginal impact on battery life Memory 128 KB 84 650 measurements maximum Download Type USB 2 0 interface or via Bluetooth Smart Full Memory Download Time 20 seconds via USB approximately 60 seconds via Bluetooth Smart may take longer the further the device is from the lo...

Page 3: ...tooth The more bars there are the stronger the Bluetooth signal The logger is currently communicating with HOBOware via USB cable The logger is currently being powered by USB cable The battery indicator shows the approximate battery power remaining The logger has been configured to stop logging when memory fills The memory bar indicates the approximate space remaining in the logger to record data ...

Page 4: ...en in the software When the LCD is turned off for logging you can still temporarily view the LCD screen by pushing the Start Stop or Clear Next button The LCD will then remain on for 10 minutes When the logger is connected to the computer with the USB cable the LCD screen refreshes every second regardless of logging interval When the logger has stopped logging the LCD screen will remain on with ST...

Page 5: ...ice attempts to connect to it To reset a password simultaneously press both the Start Stop button and the Clear Next button on the top of the logger for 3 seconds or tap and tap Reset Mark the logger as a favorite You can then filter the list of devices to only show loggers marked as favorites Update the firmware on the logger A logger readout will be completed automatically at the beginning of th...

Page 6: ... rate See Statistics Logging for more information Show LCD Enable or disable to control whether the LCD remains illuminated while the logger is logging If you disable the LCD the logger will not show the current reading status or other information while the logger is logging You will however be able to temporarily turn the LCD screen on by pressing the Start Stop button on the logger for 1 second ...

Page 7: ...ect Use Carbon Dioxide sensor altitude compensation and enter the altitude above or below sea level where the logger will be deployed in either meters or feet 6 Click OK 7 Click Start in the Launch Logger window to load settings to the logger Note If both auto calibration and manual calibration are selected the logger will automatically calibrate within 24 hours after logging begins unless a manua...

Page 8: ...lution specifications When you read out the logger alarm events can be displayed on the plot or in the data file See Logger Events Once cleared an audible alarm will start beeping again if the sensor values go out of the normal range Even if an audible alarm is cleared a visual alarm may remain on the logger LCD and in the app if applicable depending on the settings selected for maintaining visual...

Page 9: ... following statistics can be logged for each sensor The maximum or highest sampled value The minimum or lowest sampled value An average of all sampled values and The standard deviation from the average for all sampled values For example a logger is configured with the temperature and CO2 sensors enabled and the logging interval set to 5 minutes The current reading and all four statistics are enabl...

Page 10: ...am are not automatically available in the other To open app files in HOBOware share the HOBO file via email and open it on the computer with HOBOware Files in HOBOware cannot be viewed in the app You can however export data in HOBOware to a text or Excel file that you can open on your mobile device See the app user s guide and HOBOware Help for details on sharing or exporting data Logger Events Th...

Page 11: ...es can impact battery life Estimates are not guaranteed due to uncertainties in initial battery conditions and operating environment The logger can also be powered by the USB cable if faster logging intervals are desired or when the remaining battery voltage is too low for it to continue logging Connect the logger to the computer click the Readout button on the toolbar in HOBOware and save the dat...

Reviews: