background image

HOBO®

 

MX

 

Soil

 

Moisture

 

(MX2306)

 

and

  

MX

 

Soil

 

Moisture

 

and

 

Temperature

 

(MX2307)

 

Data

 

Logger

 

Manual

 

26456

A

  

The

 

HOBO

 

MX

 

Soil

 

Moisture

 

(MX2306)

 

and

 

HOBO

 

MX

 

Soil

 

Moisture

 

and

 

Temperature

 

(MX2307)

 

data

 

loggers

 

measure

 

soil

 

moisture

 

and

 

temperature

 

in

 

numerous

 

soil

 

types,

 

including

 

high

 

salinity

 

and

 

sandy

 

soil.

 

These

 

Bluetooth®

enabled

 

loggers

 

are

 

designed

 

for

 

wireless

 

communication

 

with

 

a

 

phone,

 

tablet,

 

or

 

computer.

 

Using

 

the

 

HOBOconnect®

 

app,

 

the

 

user

 

can

 

easily

 

configure

 

the

 

logger

 

and

 

then

 

download

 

logged

 

data

 

to

 

view

 

or

 

export

 

for

 

further

 

analysis.

 

The

 

loggers

 

can

 

calculate

 

minimum,

 

maximum,

 

average,

 

and

 

standard

 

deviation

 

statistics

 

and

 

be

 

configured

 

to

 

indicate

 

alarms

 

at

 

user

specified

 

thresholds.

 

The

 

loggers

 

also

 

support

 

burst

 

logging,

 

in

 

which

 

data

 

is

 

logged

 

at

 

a

 

faster

 

interval

 

when

 

sensor

 

readings

 

are

 

above

 

or

 

below

 

certain

 

limits.

 

Specifications

 

 

Measurement

 

Range

 

(select

 

one

 

at

 

configuration)

 

0.00

 

to

 

0.64

 

m³/m³

 

VWC

 

in

 

mineral

 

soils

 

0.00

 

to

 

0.70

 

m³/m³

 

VWC

 

in

 

soilless

 

media

 

Apparent

 

Dielectric

 

Permittivity:

 

1

 

(air)

 

to

 

80

 

(water)

  

Accuracy

 

*

 

±0.031

 

m³/m³

 

typical

 

in

 

mineral

 

soils

 

that

 

have

 

solution

 

EC

 

<8

 

dS/m

 

±0.051

 

m³/m³

 

typical

 

in

 

soilless

 

media

 

Apparent

 

Dielectric

 

Permittivity:

 

+/

‐ 

1

 Ɛ

a

 

for

 

the

 

range

 

1

 

to

 

40

 

(soil

 

range);

 

+/

15%

 

for

 

over

 

40

  

Resolution

 

0.001

 

m³/m³

 

Apparent

 

Dielectric

 

Permittivity:

 

0.1

 Ɛ

a

 

Dielectric

 

Measurement

 

Frequency

 

70

 

MHz

 

Volume

 

of

 

Influence

 

430

 

ml

  

  

Percent

 

of

  

  

Field

 

Capacity

 

  

(for

 

VWC

 

measurements

      

only)

 

Range:

 

0

 

to

 

200%

 

Resolution:

 

0.1%

 

 

Temperature

 

Sensor

 

(MX2307)

 

 

Range

 

40

 

to

 

100°C

 

(

40

 

to

 

212°F),

 

with

 

tip

 

and

 

cable

 

immersion

 

in

 

fresh

 

water

 

up

 

to

 

50°C

 

(122°F)

 

for

 

one

 

year

 

Accuracy

 

±0.25°C

 

from

 ‐

40

 

to

 

0°C

 

(±0.45

 

from

 ‐

40

 

to

 

32°F)

 

±0.2°C

 

from

 

0

 

to

 

70°C

 

(±0.36

 

from

 

32

 

to

 

158°F)

 

±0.25°C

 

from

 

70

 

to

 

100°C

 

(±0.45

 

from

 

158

 

to

 

212°F)

 

Resolution

 

0.04°C

 

(0.072°F)

 

Drift

 

<0.01°C

 

(0.018°F)

 

per

 

year

 

Response

 

Time

 

3

 

minutes

 

in

 

air

 

moving

 

1

 

m/s;

 

20

 

seconds

 

in

 

stirred

 

water

 

4

 

minutes

 

in

 

air

 

moving

 

1

 

m/s

 

with

 

RS3

B

 

Solar

 

Radiation

 

Shield

 

Logger

 

 

Operating

 

Range

 

25°

 

to

 

60°C

 

(

13°

 

to

 

140°F)

 

Radio

 

Power

 

1

 

mW

 

(0

 

dBm)

 

Transmission

 

Range

 

Approximately

 

30.5

 

m

 

(100

 

ft)

 

line

of

sight

 

Wireless

 

Data

 

Standard

 

Bluetooth

 

Low

 

Energy

 

(Bluetooth

 

Smart)

 

Logging

 

Rate

 

1

 

second

 

to

 

18

 

hours

 

Logging

 

Modes

 

Fixed

 

interval

 

(normal,

 

statistics)

 

or

 

burst

 

Memory

 

Modes

 

Wrap

 

when

 

full

 

or

 

stop

 

when

 

full

 

Start

 

Modes

 

Immediate,

 

push

 

button,

 

date

 

&

 

time,

 

or

 

next

 

interval

 

Stop

 

Modes

 

When

 

memory

 

full,

 

push

 

button,

 

date

 

&

 

time,

 

or

 

after

 

a

 

set

 

logging

 

period

 

HOBO

 

MX2306

 

and

  

MX2307

 

Data

 

Loggers

 

Models:

 

 

Soil

 

Moisture

 

(MX2306)

 

 

Soil

 

Moisture

 

and

 

Temperature

 

(MX2307)

 

Included

 

Items:

 

 

Screws

 

 

Cable

 

ties

 

Required

 

Items:

 

 

HOBOconnect

 

app

 

 

Mobile

 

device

 

with

 

Bluetooth

 

and

 

iOS,

 

iPadOS®,

 

or

 

Android™,

 

or

 

a

 

Windows

 

computer

 

with

 

a

 

native

 

BLE

 

adapter

 

or

 

supported

 

BLE

 

dongle

 

Accessories

 

(sold

 

separately):

 

 

Solar

 

radiation

 

shield

  

(RS3

B

 

for

 

use

 

with

 

MX2307)

 

 

Soil

 

moisture

 

verification

 

clip

 

(TEROS

CLIP)

 

 

Replacement

 

battery

  

(HRB

2/3AA)

 

 

 

MX2307

 

Model

 

Shown

 

Summary of Contents for MX2306

Page 1: ...Hz Volume of Influence 430 ml Percent of Field Capacity for VWC measurements only Range 0 to 200 Resolution 0 1 Temperature Sensor MX2307 Range 40 to 100 C 40 to 212 F with tip and cable immersion in...

Page 2: ...NEMA 6 and IP67 Soil Moisture sensor and cable 40 to 60 C 40 to 140 F Temperature sensor and cable MX2307 Immersion in water up to 50 C 122 F for 1 year The CE Marking identifies this product as comp...

Page 3: ...com products software hoboconnect 2 Open the app and enable Bluetooth in the device settings if prompted 3 Press the button on the logger to wake it up 4 Tap Devices and then tap the logger in the ap...

Page 4: ...ion settings are loaded on the logger The data will start with this indetermined time and at the selected logging interval from there On Next Logging Interval This is the most common start option Logg...

Page 5: ...n any time during logging For example if the low alarm is set to 0 05 m m for a soil moisture alarm and the duration is set to 30 minutes the alarm will trip once the sensor readings have been below 0...

Page 6: ...g During fixed logging the logger records data for enabled sensors and or selected statistics at the logging interval selected Statistics are calculated at a sampling rate you specify with the results...

Page 7: ...ger records the following internal events to track logger operation and status You can view events in exported files or plot events in the app To plot events tap HOBO Files and select a file to open T...

Page 8: ...the sensor body in a horizontal position will provide measurements at a more discreet depth Avoid any metal located between the sensor and the ferrite core because it can interfere with VWC measuremen...

Page 9: ...not cut open incinerate heat above 85 C 185 F or recharge the lithium battery The battery may explode if the logger is exposed to extreme heat or conditions that could damage or destroy the battery c...

Reviews: