background image

Non-contact infrared thermometer

981C

 

Instruction manual 

1

Introduction 

Compact,  rugged  and  easy  to  use.  Just  aim  and  push  the 
button,  read  current  surface  temperatures  in  less  than  a 
second.  Safely  measure  surface  temperatures  of  hot, 
hazardous or hard-to-reach objects without contact. 

How it works 
Infrared thermometer measures the surface temperature of 
an  object.  The  unit’s  optics  sense  emitted,  reflected,  and 
transmitted  energy  which  is  collect  and  focused  onto  a 
detector.  The  unit’s  electronics  transmitted  energy  which 
is display on the unit. For increased ease and accuracy the 
laser pointer makes aiming even more precise. 

Cautions 

Infrared  thermometer  should  be  protected  for  the 
following: 
--EMF(electro-magnetic  fields)  from  arc  welders, 
induction heaters. 
--Thermal  shock(cause  by  large  or  abrupt  ambient 
temperature  changes  allow  1  hours  for  unit  to  stabilize 
before use). 
--Do  not  leave  the  unit  on  or  near  objects  of  high 
temperature. 

Warning 

Do  not  point  laser  at  eye  or  indirectly  off  reflective 
surfaces. 
1.

When  take  measurement,  point  thermometer  toward

the object to be measured and hold the yellow trigger. 
The  object  under  test  should  be  large  than  the  spot 
size calculated by the field of view diagram. 

2.

Distance 

  spot  size:  As  the  distance  from  the

object  increase,  the  spot  size  of  measuring  area
becomes large.

3.

Field of view: Make sure the target is larger than the
unit’s  spot  size.  The  smaller  the  target  the  close
measure  distance.  When  accuracy  is  critical,  make
sure  the  target  is  at  least  twice  as  large  as  the  spot
size.

4.

Emissivity:  Most  organic  materials  and  painted  or
oxidized  surfaces  have  an  emissivity  of  0.95.
Inaccurate  readings  will  result  from  measuring  shiny
or polished metal surfaces. To compensate, cover the
surface  to  be  measured  with  masking  tape  or  flat
black paint. Measure the tape or painted surface when
the tape or painted reach the same temperature as the
material underneath.

2

Quick start instruction 

1.

Press  battery  door  clip,  install  battery  correctly.  Pull
the  trigger,  LCD  display  reading 

  battery  icon.

Release  the  trigger  and  the  reading  will  hold  for  7
seconds

LCD display:   
A    LCD display                                    B    MAX 
C    Low temperature warning setting    D    Average   
E    High temperature warning setting    F    Emissivity 
G    MIN     

H    Storage date NO. 

I      Low battery indication     

  J    Laser Light   

K    High temperature warning 

L    Measure data 

M    Low temperature warning 
N    Temperature Unit     

O    Date HOLD 

2.

Locating  a  hot  spot  aim  the  thermometer  outside  the

(Figure 3) 

(Figure 1) 

(Figure 2) 

Reviews: