OPERATION MANUAL 

Revision: January 2021 

 
 

MOTORISED TREATMENT CHAIR 

 

 

 
 

Summary of Contents for TRIAD

Page 1: ...TRIAD IBIOM Instruments Ltd 670 Joseph Louis Mathieu Street Sherbrooke Quebec J1R 0X3 Tel 1 877 563 5777 Fax 450 445 9837 www ibiom com info ibiom com OPERATION MANUAL Revision January 2021 MOTORISED...

Page 2: ...RECTION LOCK 10 6 3 CHAIR FUNCTIONS 11 6 4 ARMREST 12 6 5 EMERGENCY POSITION ELECTRICAL 15 6 6 SIDE TABLE 16 6 7 RETRACTABLE SIDE TABLE 17 6 8 RECHARGEABLE BATTERY 18 6 9 EMERGENCY FOOT SWITCH 19 7 CL...

Page 3: ...the following information ready 7 digit serial number example 1234567 Model number or name example TRIAD This information is available on label located on base of the chair Phone 1 877 563 5777 Fax 4...

Page 4: ...ader about a situation which if not avoided could result in death or serious injury It may also describe potential serious adverse reactions and safety hazards Warning electrical shock hazard Operator...

Page 5: ...eight and accessories WARNING Never exceed safe working load of 100lb or 45kg on TRIAD s armrests Safe working load includes patient s weight and accessories optional retractable side table for exampl...

Page 6: ...reach WARNING Avoid placing hands and other body parts near moving parts when the chair is adjusted WARNING Always lower and close the armrests once the patient is in place on the chair WARNING Alway...

Page 7: ...ame and address Product code or name 7 digit serial number Product s weight Date of manufacture YYYY MM DD Maximum safe working load maximum patient weight Made in Canada QR code pointing to IBIOM s w...

Page 8: ...MRI environment WARNING Do not use in wet environment WARNING Do not use for CPR or X ray procedures or with defibrillator WARNING Do not roll on power cable or hand control cable WARNING Keep all cab...

Page 9: ...ANT Make sure you have enough space around the chair to avoid collisions between its surroundings in ALL POSITIONS 3 Once in place lock all wheels 4 Remove any remaining packaging from the product Exa...

Page 10: ...any other cable WARNING Always ensure there is sufficient clearance around the chair before activating the various movements to avoid collisions with the environment equipment and people Central lock...

Page 11: ...ended WARNING Medical staff should inform the patient on how to adjust the chair WARNING If patient is not able to adjust the chair the hand control should be placed out of the patient s reach WARNING...

Page 12: ...nd close the armrests once the patient is in place on the chair WARNING Always ensure that the rotation of the armrest is locked in the closed position once the patient is in place on the chair WARNIN...

Page 13: ...Page 13 of 30 Revision January 2021 Swivel armrest WARNING Always ensure that the rotation of the armrest is locked in the closed position once the patient is in place on the chair 1 Push and hold lev...

Page 14: ...ifting armrest WARNING Always lower and close the armrests once the patient is in place on the chair Each armrest can be lifted Lifting may be locked with combination of some options 1 Hold armrest un...

Page 15: ...stretcher to move the patient from one room or floor to another WARNING Always ensure there is sufficient clearance around the chair before activating the various movements to avoid collisions with t...

Page 16: ...e WARNING Always ensure that side table mechanism is tighten solidly before the patient is placed in the chair WARNING Take appropriate precautions to avoid injury when handling accessories arms and s...

Page 17: ...BIOM Instruments ltd 2021 Page 17 of 30 Revision January 2021 6 7 Retractable side table WARNING Never exceed safe working load of 10lb or 4 5kg on TRIAD s retractable side table 1 Slightly lift side...

Page 18: ...y develop excessive heat CAUTION NEVER discharge completely battery or it may be damaged and reduce its lifetime Always allow battery to settle to room temperature before use Do not exceed duty cycle...

Page 19: ...oid placing hands and other body parts near moving parts when the chair is adjusted Emergency foot switches are located on both side of the chair 1 To activate the emergency position press the pedal d...

Page 20: ...roducts or in concentration other than those specified in this section WARNING Do not use abrasive cleaning products WARNING Avoid placing hands and other body parts near moving parts when the chair i...

Page 21: ...ngs between the backrest and seat and those between the seat and leg section IMPORTANT Allow the surfaces to dry before raising the backrest or leg section o NOTE In case of major spill it is possible...

Page 22: ...position Foot switch for emergency position option Adjustable backrest 0 to 80 0 Trendelenburg Adjustable leg 0 to 50 Adjustable height 20 to 28 in 51 to 71 cm Adjustable legrest 6 in 15 cm Emergency...

Page 23: ...le sealed lead acid Battery capacity option 1 3 Ah Protection class IP Control box IPX6 Actuator IPX4 Lifting column IPX0 Hand control IPX6 Foot switch IPX6 Battery IPX4 Duty cycle Intermittent 2 18 M...

Page 24: ...truments ltd 2021 Page 24 of 30 Revision January 2021 8 4 Certification All IBIOM s treatment chairs are certified according to CSA standard for electrical medical device except contrary indication La...

Page 25: ...r by misuse of the user guide instructions and warnings Also it doesn t apply if modifications or repairs have been done by someone not authorized by IBIOM instruments ltd Electrical components are gu...

Page 26: ...ose of it properly in accordance with the regulations in force in your region Before disposal equipment must be clean properly to avoid contamination Equipment can be disposed of by separating the dif...

Page 27: ...OPERATION MANUAL TRIAD Copyright IBIOM Instruments ltd 2021 Page 27 of 30 Revision January 2021 11 MAINTENANCE See service manual...

Page 28: ...movements If problem persists go to next cause CAUTION Never allow the battery to discharge completely as this may damage the battery Green light on the control box with or without audible alarm Cont...

Page 29: ...OPERATION MANUAL TRIAD Copyright IBIOM Instruments ltd 2021 Page 29 of 30 Revision January 2021 13 SPARE PARTS See service manual...

Page 30: ...OPERATION MANUAL TRIAD Copyright IBIOM Instruments ltd 2021 Page 30 of 30 Revision January 2021 IBIOM INSTRUMENTS LTD 670 JOSEPH LOUIS MATHIEU STREET SHERBROOKE QUEBEC J1R 0X3 CANADA www ibiom com...