Model:

 

PM-20 

Sensor type: 

Differential 

Operating range: 

130mbar 

 

Connector Type:

  

Push fit connector 

Units: 

 

mbar  

Operating Temperature:

 

-10°C to +50°C

Accuracy of Unit Only

:    

+

20 to +50°C:  0.10% of RDG +0.10% of FS + 1 LSD

-10 to +20°C:  0.15% of RDG +0.15% of FS + 1 LSD
All ± 1 LSD 

Digit ( 0.1˚C Low scale; 1˚C High scale )

Negative pressure: 0.3% r 0.3% full scale

Note: FS=Full Scale LSD= Least Significant Digit (Full range of the unit to be used for both low & high scale calculations) 

Particular care should be taken not to over pressure the device as this may rupture the sensor membrane. This is not covered by the 
manufacturers warranty. 

Note: The differential/gauge instrument will only measure relative pressures i.e positive pressure applied to the positive port or 
negative pressure applied to the negative port. 

Instrument Application 

The PM-20 is suitable for the measurement of differential pressure. 

Instrument Operation 

The instrument requires two AA or equivalent cells. These are not supplied with the unit. 
When the batteries require changing the following symbol will be displayed:  
Please ensure the screws holding the battery compartment are firmly tightened when inserting or replacing batteries to ensure the 
IP67 rating is maintained. 
Press ON key to turn unit ON. Press OFF key to turn unit OFF. 
The unit will switch OFF automatically after 12 minutes, unless any key is activated. 
To disable automatic switch off, hold      Key when the unit is being switched on. This will disable the automatic switch-off function until 
the unit is switched off. 

Out of Range 

If pressure exceeds the scale range 

Out

 will appear on the display. 

Zeroing 

In order to achieve maximum accuracy, it is recommended to zero the instrument in the orientation it will be used before taking any 
measurements. 
Ensure both ports are open to atmospheric pressure, if indicated pressure is within 3% of full scale of the factory calibrated zero, press 

0.0

 key which will adjust the reading to zero. 

Hold Function 

To enable the Hold Function, press the “Hold”

      button, press again to release the Hold Function. 

Hose fitting 

The input port(s) on the instrument are designed to accept a hose with 8mm Interior Diameter and 1mm wall. The hose is fitted by 
pushing the hose onto the port, an additional clamp fitting may be used to ensure secure fitting. 

Back Light 

Press and hold 

ON

 key. 

Time Test Facility Mode 

The instrument’s Timer Mode indicates the measured pressure and time showing alternating on the display. The 

time advances in 

minutes and seconds to a maximum time of 99mins 59sec. The timer can be paused, continued or reset as required during the test 
period. The timer mode needs to be selected to be viewed on the display. 