Digital Multimeter
Instruction Manual

CAT II

MAX

IMPORTANT SAFETY INFORMATION
Read

and 

understand

all of the instructions and

safety information in these operating instructions
before using this meter. Use the meter only as
specified in this manual; otherwise, the protection
provided by the meter may be impaired.

A        

WARNING

statement identifies hazardous

conditions and actions that could cause bodily
harm or death.

WARNINGS

To avoid possible electric shock, personal injury or
death follow these guidelines:
• Do not use if meter appears damaged.  Visually

inspect the meter to ensure case is not cracked
and back case is securely in place.

• Inspect and replace leads if insulation is dam-

aged, metal is exposed, or probes are cracked.

• Do not use meter if it operates abnormally as

protection maybe impaired.  

• Do not use during electrical storms or in wet

weather.

• Do not use around explosive gas, dust, or vapor.
• Do not apply more than the rated voltage to the

meter. 

• Do not use without the battery and the back case

properly installed.

• Remove the test leads from the meter before

removing battery cap.

• Do not attempt to repair this unit as it has no

user-serviceable parts.

•  Disconnect power and discharge capacitors

before testing resistance, continuity, diodes,
capacitance or temperature.

• Replace battery as soon as low battery indicator    

appears.

• Use the proper terminals, function and range for

your measurements.

To protect yourself, think "Safety First":
• Voltages exceeding 30VAC or 60VDC pose a

shock hazard so use caution.

• Use appropriate personal protective equipment

such as safety glasses, face shields, insulating
gloves, insulating boots, and/or insulating mats.

CAT II

MAX

• Before each use:

- Perform a continuity test by touching the

test leads together to verify the functionality
of the battery and test leads.

- Use the 3 Point Safety Method.  (1) Verify

meter operation by measuring a known
voltage.  (2) Apply meter to circuit under
test.  (3) Return to the known live voltage
again to ensure proper operation.

• Never ground yourself when taking electrical

measurements.

• Connect the black common lead to ground

before applying the red test lead to voltage.
Disconnect the red test lead from the voltage
first.

• Always work with a partner.
• When using the probes, keep fingers as far

behind the probe tips as possible.

Ranges & Accuracies:  

AC Converter:  

61-312 model is averaging

sensing, rms calibrated; 61-314 model is true rms
sensing. 

Accuracy:  

Accuracy is specified as +/-(a

percentage of the r a fixed amount) at
23°C±5°C (73.4°F ± 9°F), less than 75% relative
humidity.

Temperature Coefficient:

0.1 times the

applicable accuracy specification from 32°F to
64°F and  82°F to 122°F (0°C to 18°C ; 28°C to
50°C).

~

Symbols

Risk of electric shock

See instruction card

DC measurement

Equipment protected by double or 
reinforced insulation

Battery

Earth

AC measurement

Conforms to EU directives

Ranges & Accuracies

Function

Range &

R

esolution

A

ccuracy

Overload Protection

DC

V

oltage

400.0m/4.000/40.00/400.0/600 V

±(0.8%+4)

50~60Hz

40~400Hz

AC V

oltage

4.000/40.00/400.0/600 V

±(1.2%+5) for 312

±(1.5%+5) for 312

600V DC/AC rms.

±(1.2%+8) for 314*

±(1.5%+8) for 314*

DC

Current

400.0/4000 µA; 40.00/400.0 mA

±(1.2%+5)

500mA/250V  Fast Fuse

4.000/10.00 A

±(1.5%+5)

10A/250V  Fast Fuse

AC

Current

400.0/4000 µA; 40.00/400.0 mA

±(1.5%+5) 

500mA/250V  Fast Fuse

(40~400Hz)

4.000/10.00 A

±(2.0%+5) 

10A/250V Fast Fuse

400.0/4.000k/40.00k/400.0k 

Ω

±(1.0%+4) 

Resistance

4.000M

Ω

±(1.2%+4)

250V DC/AC rms.

40.00M

Ω

±(1.5%+5)

Capacitance**

**40.00nF/400.0n/4.000µ/40.00µ/100.0µ

F

±(3.0%+5) 

250V DC/AC rms.

Frequency

10.00/100.0/1.000k/100.0k/1.000M/10.00M Hz

±(0.3%+4) 

250V DC/AC rms.

Sensitivity: 1V rms ; 5V rms (>1MHz)

Diode Check

Test current : 0.6mA and then open circuit voltage is 1.5VDC typical.

250V DC/AC rms.

Continuity

The buzzer sounds with less than 30

Ω

.

Te

mperature***

-4~1382°F

±(

1.

0%

+4

@

-4~572°F ; ±(3.0%+5) 

@

573~1382°F

-20~750°C

±(

1.

0%

+4

-20~300°C ; ±(3.0%+5) 

301~750°C

*CF<2

**Accuracy not available for <10nF capacitance.

***Accuracy is stated for meter only

.  Thermocouple accuracy adds another ±2% to reading.

Input Impedance :  10M

Ω

for A

V

C &

DCV

DC Voltage (V )

CAT II

MAX

Toggle for DC

ND 6296-1 61-312 61-314 Ins  5/7/07  11:16 AM  Page 1

Reviews: