Scale Users Guide

ISHTAA SCALES INC., 

 INDIA

www.ishtaascales.com

ITA / ITB - AS / AP / AP SERIES

Electronic Table Top Weighing Balance

 

ITA Series – Table Top Precision Weighing Balance

ITA Series – Table Top Precision Weighing Balance

TM

14

Summary of Contents for ITA SERIES

Page 1: ...Guide ISHTAA SCALES INC INDIA www ishtaascales com ITA ITB AS AP AP SERIES Electronic Table Top Weighing Balance ITA Series Table Top Precision Weighing Balance ITA Series Table Top Precision Weighing Balance TM 14 ...

Page 2: ... and purposeful destruction are NOT covered by this warranty Technical Service MUST be performed by authorised staff or factory authorized personnel Manufacturer s obligation is limited to replacement of parts that have been so returned and are disclosed to Manufacturer s satisfaction to be defective The provisions of these warranty clauses are lieu of all other warranties expressed or implied and...

Page 3: ...ernal weight Weight 4 kg Approx Pan Size 230mm x197mm Front Display Type 0 8 Bright red LED display Rear Display Optional Available with 0 56 LED Display Communication Stabilization 3 seconds Power 9V AC Adaptor or Power Cord built in battery for backup to 30hrs 40 hrs with sleep mode active Operating Temp 50 to 98 F 10 to 35 C Dimensions 245 x 210 x 140 w x d x h SOLUTION ERROR DISPLAY ERROR 8 ER...

Page 4: ...rectly when the bubble scale is in the center of the circle The unit comes with 12V AC adaptor or 220V AC as per model specifications country of use The scale is fitted with a built in 6V 4Ah battery which provides backup of more than 35 50 hours depending on glow setting and sleep mode setting Put a known weight on the machine and make sure that the scale reads the correct weight on the display I...

Page 5: ...ATION Internal calibration also referred as service calibration requires opening the machine and changing the jumpers It is advisable to do this step only if you have required assistance or the service personnel assisting you to do this Open the battery cover from the bottom of the machine and remove the cover and change the calibration jumper marked as CAL to ON Make sure there is nothing kept on...

Page 6: ...e changes geographic gravity variations altitude changes and abuse are few reasons why a scale may need recalibration The scale can be calibrated using a known external weight 4 9 7 1 EXTERNAL USER CALIBRATION To start calibration turn the scale off and then turn it on During the count down press and hold down the Zero button do not release until the display shows CAL Make sure there is no weight ...

Page 7: ...ounting function to come in to the weighing mode It is important to know that SLEEP mode increases the battery back up time by double The Normal backup time when the sleep mode is switched off is 30 Hours and with the SLEEP mode switched on the battery back up time increases to more than 60 Hrs If the USER wants to switch off the SLEEP MODE for some special usage or function it can be done as foll...

Page 8: ...s 3d ITA Series Table Top Precision Weighing Balance ITA Series Table Top Precision Weighing Balance 6 7 6 3 LED GLOW The display GLOW strength can be increased or decreased depending upon the user requirement Increasing the GLOW of the display will effect the Battery back up time and the battery life The GLOW setting is set to minimum by Default If the user wants to increase the GLOW of the displ...