background image

ORTHOPANTOMOGRAPH

 OP 3D

CEPH upgrade kit installation manual

222078-PTU rev. 6 | 0.805.5728

ENGLISH

Dental Excellence

OP 3D

Summary of Contents for ORTHOPANTOMOGRAPH OP 3D

Page 1: ...ORTHOPANTOMOGRAPH OP 3D CEPH upgrade kit installation manual 222078 PTU rev 6 0 805 5728 ENGLISH Dental Excellence OP 3D...

Page 2: ......

Page 3: ...ving the tubehead front cover 12 3 4 7 Removing the tubehead back cover 13 3 4 8 Removing the rotating unit side covers 13 3 4 9 Removing the sensor covers 14 3 5 Remove 3D sensor 16 3 6 Remove sensor...

Page 4: ...ed photocopied reproduced translated or reduced to any electronic medium or machine readable form in whole or part without the prior written permission of PaloDEx Group Oy The original language of thi...

Page 5: ...cient space for the CEPH arm and all CEPH movements Use the supplied installation template as a guide CEPH upgrade may require you to move the Column in a different location To move the column the car...

Page 6: ...e workstation SW drivers and KaVo driver if needed 5 Go to device settings on the GUI 6 Select Service mode menu and enter PIN code 1917 7 Select the Service menu from the menu list 8 Click Select Fil...

Page 7: ...4 Select Drive to packing position 5 Press and hold down the exposure switch to drive the unit to the correct height 6 Power the device off 3 3 Isolate the device from mains Disconnect the mains powe...

Page 8: ...3 Device upgrade 3 Remove the back panel cover screw and lift the cover off 4 Disconnect all cables from the device carriage 8 ORTHOPANTOMOGRAPH OP 3D...

Page 9: ...tating unit side covers 9 Sensor collars and sensor covers Sensor collars and covers are different between 3D and PAN only devices 3 4 1 Removing the lower shelf tray 1 Insert a screwdriver or similar...

Page 10: ...ndle 2 Remove the handle 3 4 3 Removing the mirror 1 Open the mirror 2 Lift the mirror slightly so that the guiding pins 4 pcs unlock from the keyholes on the mirror plate 3 Remove the mirror from its...

Page 11: ...from the cover halves 2 Pull the covers apart turn them outwards and lift them off 3 4 5 Removing the upper shelf covers Required tools Hex key 3mm 1 Remove the screws 2 pcs from top of the upper she...

Page 12: ...CE Be careful not to damage the clamps that hold the covers together 3 4 6 Removing the tubehead front cover Required tools Hex key 3mm 1 Detach the indicated screws 4 pcs that hold the cover on the t...

Page 13: ...s a screw cover cap over it 2 Remove the tubehead back cover 3 4 8 Removing the rotating unit side covers Required tools Hex key 3mm NOTICE Turn the rotating unit carefully so that it does not collide...

Page 14: ...sor covers Required tools Hex key 3mm 2D ONLY DEVICES 1 Detach the 2D sensor cover screw 2 Pull the sensor collar halves apart to detach them from the sensor 3D DEVICES 1 Pull the 2D sensor collar cov...

Page 15: ...eft side cover upwards and pulling the covers apart 3 Remove the screw on the 2D sensor cover 4 Pull the sensor covers apart to detach them To help unlocking the cover halves squeeze the left side cov...

Page 16: ...the rotating unit NOTICE If the device is PAN only and does not have a 3D sensor skip this step 1 Remove the sensor connectors and their EMC clamps from i2000 circuit board 2 Remove the sensor power c...

Page 17: ...r in place NOTICE Hold on to the sensor when you remove the screws When all screws are off the sensor can easily fall down and break 5 Remove the sensor and sensor cables from the device Both will be...

Page 18: ...NOTICE A new CEPH imaging compatible sensor holder assembly is supplied with the upgrade kit What to do next If the device is serial number KR1902594 or older upgrade the linear movement mechanism an...

Page 19: ...upper shelf You must remove the rotating unit from the device to do this 1 Detach all main cable set connectors from the i3000 main control board and tubehead cable connectors from the upper shelf J30...

Page 20: ...ead must be in home position on the right hand side of the device 4 Push the cable clamp lid down and pull it open to release the main cable 5 Remove the screws on the linear movement sledge that hold...

Page 21: ...and slide it off from upper shelf Lay the rotating unit carefully on the floor NOTICE 2 persons are recommended for this task 8 Remove the linear wiper from the linear movement sledge 9 Remove the re...

Page 22: ...t of the coupler NOTICE Do not remove the linear movement screw from the linear movement sledge 13 Remove the linear movement sledge with the linear movement screw from the upper shelf 14 Loosen the s...

Page 23: ...etaining nut 20 Check that the linear movement screw is able to move a bit in horizontal direction 21 Move the linear movement back and forth by hand to make sure that the movement is smooth 22 Tighte...

Page 24: ...nt sledge by pushing it in NOTICE Make sure that the wiper is locked firmly to its place 26 Lift and slide the rotating unit to the rails on the linear movement sledge NOTICE The Linear movement sledg...

Page 25: ...ly 2 screw holes then attach the rotating unit with only 2 screws 28 Route the main cable set through the cable clamps and attach the clamps The main cable set has markings that indicate the correct c...

Page 26: ...spring to the rotating unit and connect the motor cable 5 Push the motor mount inwards and lift the belt up around the sprocket NOTICE Make sure that the belt is tight and has not fallen off the larg...

Page 27: ...rriage from the column safely 1 Disconnect the Column control and Emergency switch connectors connected to top of the column 2 Remove the front support bar screws 2 pcs 3 Pull the front support bar ou...

Page 28: ...otating unit against the carriage and attach the rotating unit support to the carriage with 4 screws M8x16 6 Attach the supplied plywood support on top of the upper shelf with 4 screws M8x25 and washe...

Page 29: ...e the device package cardboard sheets This prevents unwanted damage to the floor during the device disassembly 1 Make sure that the spring latch in the column is securely in place 2 Loosen the carriag...

Page 30: ...the floor against the front support bar 6 Pull the carriage off the column support hinge 7 Turn the carriage upside down so that it rests on top of the plywood support 8 Remove the wall bracket screw...

Page 31: ...all with CEPH wall bracket 3 Fix the column to the floor with a pivot screw 4 Install the carriage back on the column 5 Remove the supports installed during the device disassembly 3 12 Install the CEP...

Page 32: ...s M4x16 2 pcs on both sides of the rotating unit 4 Attach the device cables to i2000 image capture board and the cable clamp size 3N with M4x8 and grounding cables M4x8 with star washers 2 pcs to the...

Page 33: ...hrough the cable holder and through the opening on the sensor holder assembly 8 Turn the 2D sensor movement screw counter clockwise with a flat head screwdriver or 3mm hex key to move the 2D sensor as...

Page 34: ...x20 1 on both sides of the rod 12 Attach the previously removed 3D sensor to the 3D sensor mount with 4 screws NOTICE The black surface of the sensor must face the tubehead 13 Connect the 3D sensor da...

Page 35: ...install 3 14 Install CEPH secondary collimator Install the CEPH secondary collimator to the PAN 3D tubehead and connect all remaining cables to i2000 circuit board 1 Remove the 2 indicated screws and...

Page 36: ...ews on the collimator assembly 4 Install the secondary collimator frame to the tubehead frame with 2 screws M4x16 5 Tighten the previously loosened screws 6 Attach the collimator plate in place with 2...

Page 37: ...ondary collimator Connect the connector J1005 to CEPH laser 8 Test that the child collimator plate can move freely and that it does not hit the tubehead 9 Route the collimator cable to the i2000 circu...

Page 38: ...does not interfere with the cover installation NOTICE Make sure that the collimator cable and the cable labels do not interfere the X ray beam 11 Connect the collimator cable connector J2008 and all...

Page 39: ...that the GUI shows CEPH imaging modality This means that the device identifies the new CEPH arm and that it is functional NOTICE If the new imaging modality is not shown on the GUI check that all dev...

Page 40: ...up Oy representative for the most current information Copyright by PaloDEx Group Oy All rights reserved Headquarters PaloDEx Group Oy Nahkelantie 160 FI 04300 Tuusula FINLAND Tel 358 10 270 2000 https...

Reviews: