background image

INSTRUCTION MANUAL

KEW MATE 2000A/2001A

DIGITAL MULTIMETER 

WITH 

AC/DC CLAMP SENSOR

KYORITSU ELECTRICAL INSTRUMENTS

WORKS, LTD.

1.SAFETY WARNINGS

This instrument has been designed and tested according to IEC Publication 61010: 

Safety Requirements for Electronic Measuring Apparatus.  This instruction manual 

contains warnings and safety rules which must be observed by the user to ensure 

safe operation of the instrument and to retain it in safe condition.  Therefore, read 

through these operating instructions before starting using the instrument.

 

   

WARNING

● Read  through  and  understand 

instructions  contained  in  this  manual  before 

starting using the instrument.  

● Save and keep the manual handy to enable quick reference whenever necessary. 

● Be  sure  to  use  the  instrument  only  in  its  intended  applications  and  to  follow 

measurement procedures described in the manual. 

● Be sure to understand and follow all safety instructions contained in the manual.

The instrument is to be used only in its intended applications.

Understand and follow all the safety instructions contained in the manual.

Failure  to  follow  the  instructions  may  cause  injury,  instrument  damage  and/or 

damage  to  equipment  under  test.  Kyoritsu  is  by  no  means  liable  for  any  damage 

resulting from the instrument in contradiction to this cautionary note.

Failure  to  follow  the  above  instructions  may  cause  injury,  damage  to  the  instrument 

and/or damage to equipment under test.

The symbol   indicated on the instrument means that the user must refer to related 

parts of the manual for safe operation of the instrument.  Be sure to carefully read the 

instructions following each   symbol in this manual.

  

DANGER

   is reserved for conditions and actions that are likely to cause serious or fatal injury.

  

WARNING

 is reserved for conditions and actions that can cause serious or fatal injury.

  

CAUTION

   is reserved for conditions and actions that can cause minor injury or instrument damage.

Following symbols are used on the instrument and in the instruction manual. Attention 

should be paid to each symbol to ensure your safety.

Refer to the instructions in the manual.

This symbol is marked where the user must refer to the instruction manual so as not 

to cause personal injury or instrument damage.
Indicates an instrument with double or reinforced insulation.
Indicates that this instrument can clamp on bare conductors when measuring

a voltage corresponding to the applicable Measurement category, which is marked 

next to this symbol.
Indicates AC (Alternating Current).
Indicates DC (Direct Current).
Indicates AC and DC.

   

DANGER

● Never  make  measurement  on  circuits  with  a  maximum  voltage  difference  of 

600VAC/DC  or  greater  between  conductors  (300VAC/DC  or  greater  between  a 

conductor and ground). 

● Do  not  attempt  to  make  measurement  in  the  presence  of  flammable  gasses.  

Otherwise,  the  use  of  the  instrument  may  cause  sparking,  which  leads  to  an 

explosion.  

● Never attempt to use the instrument if its surface or your hand is wet.  

●Do not exceed the maximum allowable input of measuring ranges.

● Never open the battery compartment cover while making measurement.

● Never  try  to  make  measurement  if  any  abnormal  conditions,  such  as  broken 

Transformer jaws or case is noted.

● The  instrument  is  to  be  used  only  in  its  intended  applications  or  conditions. 

Otherwise,  safety  functions  equipped  with  the  instrument  doesn t  work,  and 

instrument damage or serious personal injury may be caused.

are  noted, 

to the 

 Return 

under test 

cover  for 

or color 

before 

e  making 

use. When 

after 

abrasives 

shes  safety 

, and called 

  electrical 

t  designed 

e designed 

pply. 

outlet by a 

tly  to  the 

tlets.

the power 

nel).

that  comes  

not require 

category 

3.SPECIFICATIONS

●Measuring Ranges and Accuracy (at 23℃±5℃, relative humidity75% or less)

※Electromagnetic RF field  ≦ 1 V/m
  compatibility     ACV/DCV/OHMS/FREQUENCY total accuracy = specified accuracy
  (IEC 61000-4-3)     ACA/DCA            total accuracy = specified a5dgt
          RF  transmitters  such  as  mobile  telephones  may  not  be  used  in  close 

proximity.

●Safety Standard 

IEC 61010-1

   

 

measurement CAT 

, 300V, pollution degree 2

   

 

measurement CAT 

, 600V, pollution degree 2

   

 

IEC 61010-031

   

 

IEC 61010-2-032, IEC 61010-2-033

   

 

IEC 61326-1 (EMC), EN 50581 (RoHS)

●Operating System 

Dual integration

●Display 

 

 Liquid  crystal  display  with  maximum  reading  of  3399  as 

well as units and annunciators

   

 

Bar graph with maximum points of 33

●Over Input Indication 

"OL" on the LCD (Ω ranges only)

●Auto-ranging Operation 

 Shifts to the next higher range when bar graph increases 

to 33 points

   

 

 Shifts to the next lower range when bar graph decreases 

to 3 points

●Sample Rate  Numeric reading: about 400ms, 

   

 

bar graph: about 20ms

MODEL

2000A

2001A

Range

 60A

100A

AC Current   A                                                 

DC Current   A                                                  

AC Voltage  V Input impedance: 10MΩ

DC Voltage  V  Input impedance: 10MΩ

Measuring Range

0-60.0A

0-100.0A

Accuracy

 ± 2.0%rdg ± 5dgt(50/60Hz)

 ± 2.0%rdg ± 5dgt(50/60Hz)

MODEL

2000A

2001A

Range

 60A

100A

Measuring Range

0- ± 60.0A

0- ± 100.0A

Accuracy

± 2.0% rdg ± 5dgt

± 2.0% rdg ± 5dgt

Range

 3.4V

 34V

 340V

 600V

Measuring Range

 0-600V

(Auto-ranging)

Accuracy

± 1.5%rdg ± 5dgt(50-400Hz)

Range

 340mV

 3.4V

 34V

 340V

 600V

Measuring Range

0- ± 600V

(Auto-ranging)

Accuracy

± 1.5%rdg ± 4dgt

Resistance Ω/ 

Range

 340 Ω

 3.4k Ω

 34k Ω

 340k Ω

 3.4M Ω

 34M Ω

Measuring Range

0-33.99M Ω

 (Auto-ranging)

Accuracy

± 1.0%rdg ± 3dgt

Buzzer beeps below 30 ± 10Ω

(Continuity  buzzer  works  on  340Ω  range 

only)

± 5%rdg ± 5dgt

± 15%rdg ± 5dgt

Frequency Hz 

Range

Current

Voltage

Measuring Range

0-3.399kHz

3.4kHz-10kHz

 (Auto-ranging)

0-3.399kHz

3.4kHz-33.99kHz

34kHz-300kHz

 (Auto-ranging)

Accuracy

± 0.1%rdg ± 1dgt

± 0.1%rdg ± 1dgt

●Location for use 

 Indoor use, Altitude up to 2000m

●Accuracy-insured 

23℃±5℃, relative humidity 75% or less 

 Temperature and 

(without condensation)

 Humidity Ranges

●Operating Temperature 

0-40℃, relative humidity 85% or less 

 and Humidity Range  

(without condensation)   

●Storage Temperature 

-20-60℃, relative humidity 85% or less

 and Humidity Range  

 (without condensation)

●Source 

 

Two 1.5VDC R03 (UM-4) batteries

●Current Consumption 

Approx. 10mA

●Power-save Function 

 Shifts to the power-save state about 10 minutes after the 

last switch operation

   

 

 (current consumption: approx. 10μA)   

●Overload Protection 

AC/DC current ranges:  MODEL2000A  AC/DC  72A  for 

10 seconds

            

 AC/DC current ranges:  MODEL2001A  AC/DC 

120

A  for 

10 seconds

            

 AC/DC voltage ranges: AC/DC 720V for 10 seconds

   

 

Resistance ranges: AC/DC 720V for 10 seconds

   

 

Frequency ranges: AC/DC 720V for 10 seconds

●Withstand Voltage 

 AC3470V  for  5  sec.  between  electrical  circuit  and 

housing case

●Insulation Resistance 

 10MΩ or greater at 1000V between electrical circuit and 

housing case

●Conductor Size 

MODEL2000A Approx.  6mm diameter max.

            

 MODEL2001A Approx. 10mm diameter max. 

●Dimensions   

MODEL2000A 128(L)×87(W)×24(D)mm

            

 MODEL2001A 128(L)×92(W)×27(D)mm 

●Weight 

 

MODEL2000A Approx. 210g

            

 MODEL2001A Approx. 220g

●Accessories  

Two R03 (UM-4) batteries

   

 

Instruction Manual

4.INSTRUMENT LAYOUT

Data Hold Button

MODEL2001 MODEL2000

Range Hold Button

Digital Multimeter

Function Selector Switch

Holster

Test Lead

LCD Display

AC/DC Clamp Sensor

Barrier

Protective fingerguard

DCA

Zero Adjust

Knob

5.PREPARATIONS FOR MEASUREMENT

(1)Checking battery voltage

Set the Function Selector Switch to any position other than the OFF position. 

 If the marks on the display is clearly legible without symbol 

"BATT"

 showing, battery 

voltage  is  OK.  If  the  display  blanks  or 

"BATT"

  is  indicated,  replace  the  batteries 

according to section 

8: Battery Replacement

.

NOTE

When the instrument is left powered on, the auto-power-save function automatically 

shut the power off; The display blanks even if the Function Selector Switch is set to 

a  position  other  than  the  OFF  position  in  this  state.    To  power  on  the  instrument, 

turn the Function Selector Switch or press the Data Hold Button.  If the display still 

blanks, the batteries are exhausted.  Replace the batteries.

(2)Make sure that the Function Selector Switch is set to the appropriate range.

 

Also  make  sure  that  data  hold  function  is  not  enabled.    If  inappropriate  range  is 

selected, desired measurement cannot be made.

(3)Install Test Lead to the Holster on the side of body

 

It is possible to measure with seeing the LCD Display keep 

Test Lead installing to the Holster.

AC/DC

Battery Voltage Warning

Auto Range

Data Hold

Unit

Bar Graph

Buzzer

Manual

Range

Wind the lead around the folder.

Test Leads

AC/DC Clamp Sensor

6.HOW TO MAKE MEASUREMENT

6-1 Current Measurement

   

DANGER

● In order to avoid possible shock hazard, never make measurement on circuits with a 

maximum  voltage  difference  of  600VAC/DC  or  greater  between  conductors 

(300VAC/DC or greater between a conductor and ground). 

● Do not make measurement with the test leads connected to the circuit under test. 

Never make measurement with the battery compartment cover removed.

●Keep your fingers and hands behind the barrier during measurement.

   

CAUTION

● When handling the clamp sensor, exercise caution not to apply excessive shocks or 

vibration to the sensor. 

● 

M

aximum  measurable  conductor  size  is  MODEL2000A  6mm  /  MODEL2001A 

10mm in diameter.

 6-1-1 DC Current Measurement

(1)The Function Selector Swith to the "  A" position.

 ("DC" and "AUTO" marks are shown on the top of the display.)

(2) Turn the 0(Zero) ADJ knob to set the reading of the multimeter to zero. (If this zero 

adjustment is made incorrectly, measurement errors will result.) 

(3)  Adjust one of the conductors to the center of the clamp sensor's arrow. 

  (  When  the  position  of  the  conductor  is  not  at  the  center  of  the  arrow,  the  error 

occurs.)

     Measured value is shown on the display.

Note:  When current flows from the upside to the underside of the instrument, the polarity 

of the reading is positive (+).  Otherwise, the polarity of the reading is negative (-). 

 6-1-2 AC Current Measurement

(1)Set the Function Selector Switch to "  A."

 ("AC" and "AUTO" marks are shown on the top of the LCD.)

(2) Adjust one of the conductors to the center of the clamp sensor's arrow.

   (  When  the  position  of  the  conductor  is  not  at  the  center  of  the  arrow,  the  error 

occurs.)

     Measured value is shown on the display.

Note:  Unlike DC current measurement, zero adjustment is not necessary.  There is not 

polarity indication either.

6-2 Voltage Measurement

   

DANGER

● In order to avoid possible shock hazard, never make measurement on circuits with a 

maximum  voltage  difference  of  600VAC/DC  or  greater  between  conductors 

(300VAC/DC or greater between a conductor and ground). 

● Do not make measurement with the battery compartment cover removed.

●Keep your fingers and hands behind the protective fingerguard during measurement.

 6-2-1 DC Voltage Measurement

(1)Set the Function Selector Switch to "  V."

  ("DC" and "AUTO" marks are shown on the top of the LCD.)

(2) Connect the red test lead to the positive (+) side of the circuit under test and the 

black test lead to the negative (-) side.  Measured voltage value is shown on the 

display.

  When the connection is reversed, "-" is shown on the display.

Conductor

MODEL2000A MODEL2001A

Adjust a

conductor to

the center of

the arrow.

Correct

Wrong

that the 

value  is 

, the buzzer 

value.  

with a 

 

under test. 

under 

Measured 

is shown on 

input 

input 

leads to 

7.OTHER FUNCTIONS

7-1  Auto-Power-Save Function                                         

NOTE

A small amount of current is consumed even in the power-save state.  Make sure 

to set the Function Selector Switch to the OFF position when the instrument is not 

used.
This  function  helps  to  avoid  unwanted  exhaustion  of  the  batteries  because  of 

leaving  the  instrument  powered  on  and  extend  battery  life.    The  instrument 

automatically  shifts  to  the  power-save  state  about  10  minutes  after  the  last 

Function Selector Switch or other switch operation.  

To  return  to  the  normal  state:    Turn  the  Function  Selector  Switch  or  press  the 

Data  Hold  Button  twice  to  exit  the  power-save  state  and  enable  measurement 

functions.

7-2 Data Hold Function

This  is  a  function  to  freeze  a  measured  value  on  the  display.    Press  the  Data 

Hold Button once to hold the current reading.  In this data hold state, the reading 

is held even if input varies.  "H" and "  " marks are shown on the LCD instead 

of "AUTO" mark.  

To exit the data hold state, press the Data Hold Button again.

7-3 Range Hold Function

The  instrument  defaults  to  auto-ranging  ("AUTO"  is  shown  on  the  LCD).  

Pressing  the  Range  Hold  Button  enables  manual  selection  among  measurement 

ranges("  " mark is shown on the LCD instead of "AUTO" mark)

Press the Range Hold Button to select a higher range.  

To  switch  from  manual  range  selection  to  auto-ranging,  press  down  the  Range 

Hold  Button  for  about  one  seconds,  or  turn  the  Function  Selector  Switch  to 

another position before setting it back to the current range. 

8.BATTERY REPLACEMENT

   

WARNING

● In order to avoid possible shock hazard, always disconnect the test leads from 

the circuit under test and set the Function Selector Switch to the OFF position 

before trying to replace the batteries.

  

 CAUTION

● Do not mix new and old batteries. 

● Install  batteries  in  the  orientation  as  shown  inside  the  battery  compartment, 

observing correct polarity.

When the battery voltage warning mark "BATT" is shown on the top left corner of the 

LCD,  replace  the  batteries.    Note  that  the  display  blanks  and  "BATT"  mark  is  not 

shown if the batteries are completely exhausted.

 (1) Set the Function Selector Switch to "OFF."

 (2) Remove the instrument from the holster.

 (3) Loosen  the  battery-compartment-cover-fixing  screw  on  the  lower  back  of  the 

instrument. 

 (4) Replace the batteries with two new R03 (UM-4) 1.5V batteries.

 (5) Put the battery compartment cover back in place and tighten the screw.

Screw

Battery

Compartment

Cover

Batteries

92-2318A

6-18

DISTRIBUTOR

Kyoritsu  reserves  the  rights  to  change  specifications  or 
designs  described  in  this  manual  without  notice  and  without 
obligations.

Test  Lead  Cap:Test  leads  can  be  used  under  the  CAT.II  and  III  environments  by 

attaching  a  Protective  cap  as  illustrated  below.  Use  of  our  Protective  cap  offers 

different lengths suitable for the test environments.

Barrer

Protective fingerguard: It  is  a  part  providing  protection  against  electrical  shock  and 

ensuring the minimum required air and creepage distances.

Protective Cap

Exposed metal part

Uncapped condition for CAT.II environment

Capped condition for CAT.III environments

This  instrument  satisfies  the  marking  requirement  defined  in  the  WEEE 

Directive.  This  symbol  indicates  separate  collection  for  electrical  and 

electronic equipment.
This marking means they shall be sorted out and collected as ordained in 

DIRECTIVE.

This  directive  is  valid  only  in  the  EU.  When  you  remove  batteries  from 

this product and dispose them, discard them in accordance with domestic 

law concerning disposal. Take a right action on waste batteries, because 

the collection system in the EU on waste batteries are regulated.

www.

.com

information@itm.com

1.800.561.8187

Reviews: