background image

EL-SIE-2+

High Accuracy Temperature and Humidity 

USB Data Logger, EasyLog Cloud Compatible

Monitor the environment you live and work in with the EasyLog EL-SIE-2+. Configuration is simple, with no software to install 

on your PC or Mac – just connect the logger with a USB cable, and use your standard web browser to configure the device for 

logging. You don’t even need internet access to set up and use the EL-SIE-2+, it really couldn’t be easier.

The logging interval can be set between 10 seconds and 24 hours, with immediate, delayed, triggered or push-to-start logging. 

Alarms are fully user configurable, with functionality including cumulative alarms, pre-alarms, a delay before alarm triggering, 

and an alarm hold option, which continues showing the alarm condition even if the reading returns to an acceptable level.

The display shows current, maximum and minimum readings, and three coloured LEDs indicate device status at a glance.

Once logging is complete, re-connect to your computer and use your browser to view, analyse and save your data. You can also 

choose to upload your data to an Easy Log Cloud account, making the data accessible online for powerful graphing, analysis and 

report generation.  

Typical battery life is over 1 year using standard AAA alkaline batteries, and a wall mounting bracket is supplied with the device.

•  -18 to +55°C (-0.4 to +131°F) temperature measurement range

•  0 to 100%RH humidity measurement range
• 

Stores over 1,000,000 readings (As per spec table)

•  No software to install - configure using your normal web browser

•  Use with a PC or Mac

•  Display shows current, maximum and minimum readings

•  Status and alarm indicators
•  On-board alarm sounder

•  Data can be uploaded to the EasyLog Cloud

ACCESSORIES

BAT 1V5 AAA

2 x AAA 1.5V alkaline 

batteries

CABLE USB 3.1 C 1M

USB A to USB C Cable

INCLUDED IN THE BOX

BAT 1V5 AAA

2 x AAA 1.5V alkaline 

batteries

EL-SIE WALL BRACKET

Mounting bracket

CABLE USB 3.1 C 1M

USB A to USB C Cable

Temperature

Measurement range

-18 to +55°C (-0.4 to +131°F)

Resolution

0.01°

Accuracy

±0.2°C (±0.36°F) typical

Long term stability

<0.03°C (<0.054°F) / year

Measurement units

°C, °F or K

Relative Humidity

Measurement range

0 to 100%RH

Resolution

0.1%

Accuracy

±1.5% typical (0 to 80%RH)

Long term stability

<0.25%RH / year typical

Logging rate

10 seconds to 24 hours, user selectable

Start modes

Immediate, push to start, triggered start, 

parameter triggered

Memory capacity

Over 1,000,000 total readings >500,000 readings per 

channel

Sounder

Integrated alarm sounder

Power source

2 x AAA 1.5V battery

Battery life

>1 year (at 25°C with 10 minute logging rate)

Dimensions

93 x 42 x 17 mm (excluding bracket)

Operating temperature range

-18 to +55°C (-0.4 to +131°F)

Environmental rating

IP4X

SPECIFICATIONS

CALIBRATION CERTIFICATES AVAILABLE

Lascar offers a Traceable Calibration Certificate Service on Temperature Data Loggers. Using reference equipment 

which has been calibrated by a UKAS/NIST/CNAS accredited laboratory and using apparatus traceable to national or 

international standards. For more information please see 

www.lascarelectronics.com

.

Summary of Contents for EL-SIE-2+

Page 1: ...s over 1 000 000 readings As per spec table No software to install configure using your normal web browser Use with a PC or Mac Display shows current maximum and minimum readings Status and alarm indicators On board alarm sounder Data can be uploaded to the EasyLog Cloud ACCESSORIES BAT 1V5 AAA 2 x AAA 1 5V alkaline batteries CABLE USB 3 1 C 1M USB A to USB C Cable INCLUDED IN THE BOX BAT 1V5 AAA ...

Page 2: ...arameters to be configured Logger and channel names Measurement units Logging rate and start mode Up to 16 separate alarms high low pre alarm cumulative with thresholds delay and hold Display and LED indicator modes Once the logger is running you can plug it back into your computer and see the latest data device status and the event log You can also chose to stop the logger and change the configur...

Page 3: ...e is reached Low Battery The batteries are low and should be changed when possible Memory Full The memory is full and logging has stopped Memory 90 The memory is 90 full and data should be downloaded when possible Calibration Pending The calibration will expire in 30 days on the date shown can be DD MM YYYY or MM DD YYYY Calibration Expired The calibration has expired on the date shown can be DD M...

Page 4: ...creen Button Press Function Event Recorded USB n a n a n a n a Push to Start Any Long Start logging n a Triggered Start n a n a n a n a Delayed Start n a n a n a n a Channel Current reading Top Short Move to next channel or STOP LOG n a Long Clear alarm hold for all channels Clear Held Alarms Bottom Short Show Min reading for this channel n a Long Mute alarm sounder Mute Alarm Channel Min reading ...

Page 5: ...hows best performance when operated within the recommended normal temperature and humidity range of 5 to 55 C and 20 to 80 RH respectively Long term exposure to conditions outside the normal range especially at high humidity may temporarily offset the RH signal e g 3 RH after 60h kept at 80 RH After returning into the normal temperature and humidity range the sensor will slowly come back within th...

Page 6: ...d through exposure to a variety of compounds These products should not be kept in proximity to volatile chemicals such as solvents and other organic compounds Generally speaking if a material or compound emits a strong odour you should not keep your humidity data logger in close proximity to it If you would like more information please contact your local Lascar Electronics office High levels of po...

Reviews: