background image

101 N. Alloy Dr. 
Fenton, MI 48430

 

 

Ph (810) 

714-5811

Fax    (810) 714-5711

CustomerService@lmicorporation.com

 

 

Research, Development and Manufacturing of Precision Measuring Systems

 

 

Form:  CA 033 

September 21, 

2004 

R:\Quality\Calibration Instructions\CA 033.doc                                             Rev. A 

Page 1 of  7 

 

Configuration and Mastering Instruction for the 

LMI 200-S or LMI 200-SB to the LMI 440 or ASI DataMyte 501 

 

This process will outline: 
 
 Section 

 

 

 

 

 

 

 

         Pages 

I.

 

Configuration of the LMI 440 or ASI DataMyte 501    

2-4 

 

 

 

II.

 

Mastering the LMI 200-S or 200-SB 

 

 

 

5-6 

III.

 

How to verify the mastering of the LMI 200-S or 200-SB 

 

 
 
 

 

 

Required Equipment: 

1

 

LMI 440 or ASI DataMyte 501 

2

 

LMI 200-S or 200-SB transducer 

3

 

LMI 210 master block 

4

LMI 6009 4 pin – 4 pin cable

2

3

Summary of Contents for 200-S

Page 1: ...ge 1 of 7 Configuration and Mastering Instruction for the LMI 200 S or LMI 200 SB to the LMI 440 or ASI DataMyte 501 This process will outline Section Pages I Configuration of the LMI 440 or ASI DataMyte 501 2 4 II Mastering the LMI 200 S or 200 SB 5 6 III How to verify the mastering of the LMI 200 S or 200 SB 7 Required Equipment 1 LMI 440 or ASI DataMyte 501 2 LMI 200 S or 200 SB transducer 3 LM...

Page 2: ... manual for further details 1 Press menu to turn on the collector 2 Press 6 to highlight Gage 3 Press enter It is recommended to assign simple user name to the gage files such as 200 probe lmi probe etc This will help to identify different setups 4 To assign a gage file name press the or to highlight gage G4 in the Gage List and press enter on the collector The alphanumeric screen will then appear...

Page 3: ...ermines how the collector will interpret the signal from the gage Failure to set this screen properly may cause undesired results If no changes to the screen are required configuration is complete and press the menu key If changes are required continue to step 7 7 If changes in this screen are needed press the or to highlight the different selections then press enter to toggle through the choices ...

Page 4: ...Rev A Page 4 of 7 9 Press the menu key to return to the Main Menu If any changes were made in the Gage Configuration screen a save gage notification will appear If the changes are intentional highlight Save to current gage and press enter If changes are not intended highlight Cancel and press enter and reset Configure Gages per step 6 Gage configuration is complete ...

Page 5: ...G INSTRUCTIONS LMI suggests that this process be performed at the start of every shift 1 Connect the transducer to Gage Port 4 of the data collector If G5 was selected in gage configuration use Gage Port 5 2 Press Menu to turn on the DATAMYTE 501 3 Press to highlight Gage 4 Press enter 5 From the gage list use the or keys on the data collector to choose gage file G4 and press on the collector ...

Page 6: ...appears 7 Place the transducer into the top or Lo step of LMI 210 Master Block Verify Master Lo is highlighted on the collector press enter 8 Place the transducer into the bottom or Hi step of the LMI 210 Master Block Verify Master Hi is highlighted on the collector press enter 9 Position the transducer into the calibration block s center or Master Verify Master Zero is highlighted on the collecto...

Page 7: ...00 into the Master position of the 210 block The value on screen needs to read 0 00 03mm 2 Insert the 200 into the Lo or top step of the 210 block and observe the value 3 Insert the 200 into the Hi or bottom step of the 210 block and observe the value 4 Add the raw value of step 2 and 3 together this example shows 4 49 5 51 10 00 The sum of steps 2 and 3 needs to equal 10 00mm 03 If at any time th...

Reviews: