background image

www.logtagrecorders.com

TREL30-16

Ultra-Low 

Temperature 

Recorder with 

Display

Product User Guide

Document Release Version: 1.1

Published 22. November 2020

Copyright © LogTag Recorders, 2004-2020

LogTag Recorders

Summary of Contents for TREL30-16

Page 1: ...logtagrecorders com TREL30 16 Ultra Low Temperature Recorder with Display Product User Guide Document Release Version 1 1 Published 22 November 2020 Copyright LogTag Recorders 2004 2020 LogTag Record...

Page 2: ...11 Audible Alarm 11 Alarm Configuration Options 13 Advanced Configuration Settings 14 Display Overview 16 Real Time Clock 18 Adjusting the Display Clock 18 Starting the Logger 19 Push button start 19...

Page 3: ...Minimum Maximum Temperatures 33 Resetting the Trip Minimum Maximum Temperatures 33 Stopping the TREL30 16 35 Automatically 35 Push button stop 35 Resetting the Logger 37 Hibernating a TREL30 16 37 Tec...

Page 4: ...the improper storage and use of the device l for any problems with refrigeration units l for the bad quality of the monitored goods if any l for incorrect readings if the device was used with a low ba...

Page 5: ...is font describes features of the product This text describes certain aspects of the product where incorrect use of a feature may lead to inadvertent loss of data This text contains important informat...

Page 6: ...riggered when a number of readings are outside preset Alarm thresholds and a day alarm indicator appears on the display Logged temperature data can be downloaded via a standard LogTag Interface to the...

Page 7: ...ta directly on the display It is also used to place an inspection mark in the data listing Display The display shows at a glance if alarm events have occurred for both the current day and up to 29 day...

Page 8: ...he PC l An external probe of the ST10 series l For configuration a PC running Windows 7 or later and LogTag Analyzer installed Optional Items In addition to the above following items are useful access...

Page 9: ...le You can check if your logger is already configured or not by briefly pressing the REVIEW MARK button The logger is already configured to start If your logger is configured to start it will show eit...

Page 10: ...g Analyzer software which can also be used for downloading and analyzing data Standard Configuration Options The standard configuration options include settings such as User ID start method pre start...

Page 11: ...r Start delay If you configure the TREL30 16 to start after a delay period the logger will not immediately record temperature readings after you have pressed START CLEAR STOP but start a countdown tim...

Page 12: ...s re triggered Please see more information about re triggering alarms in The buzzer will temporarily turn off when you are reviewing data 1 The interval at which the alarm sounds will increase to 8 se...

Page 13: ...configured in the Alarm Settings tab during configuration of the logger with LogTag Analyzer The screen shows an example where l the upper alarm is triggered when the temperature is 70 0 C or above fo...

Page 14: ...or additional configuration settings These settings decide how some of the elements are displayed on the unit s own display and set certain options specific to the TREL30 16 Figure 2 TREL30 16 advance...

Page 15: ...l allowing the user to stop the logger with the REVIEW MARK button l allowing the user to reset the logger with the START CLEAR STOP button l enabling the buzzer for the audible alarm For detailed in...

Page 16: ...oday is 00 DAYS days in the past are represented between yesterday 01 DAYS and 29 DAYS During configuration of the logger in LogTag Analyzer you can also enable this to show the total number of days f...

Page 17: ...s and minutes 24 hour format l If DELAY is shown the time value represents a start delay or the time remaining until a date time start will occur l If DURATION is shown the time value represents the d...

Page 18: ...ck LogTag Analyzer can display the logged readings in a number of different time zones regardless of where it was configured Adjusting the Display Clock The display clock of the logger can be set to t...

Page 19: ...ta If a display clock adjustment is made while recording data the next log taken will be identified in the downloaded data with a time change mark Changes to the display clock do not affect the intern...

Page 20: ...button within two seconds STARTING will also disappear and the symbol will be shown The TREL30 16 now records temperature data The logger will not start if you l Release the button before READY disapp...

Page 21: ...AY are shown press and hold the START CLEAR STOP button STOPPED will now show Release the button when STARTING disappears STOPPED and DELAY will both disappear and READY will show together with the cu...

Page 22: ...Starting the Logger TREL30 16 User Guide 22 Hibernating the logger using LogTag Analyzer will abort any previously configured Date Time start...

Page 23: ...splay screens At 1 29 pm the display shows the following l Alarm events were recorded 7 19 20 25 days ago l These alarms were cleared by an inspector as the display currently shows the OK tick l The c...

Page 24: ...1 day as the display time has passed through midnight 00 00 Marking a reading with an inspection mark When you press the REVIEW MARK button while the TREL30 16 is recording the next reading taken will...

Page 25: ...tly on l Keep holding the button after the symbol remains permanently on for more than 2 seconds In this example the paused function was activated and the paused symbol is shown on the display For mor...

Page 26: ...pecially marked in the graph and data listings How long PAUSED is displayed depends on when between readings you press the button It will show longer if the button is pressed just after the logger tak...

Page 27: ...on the display and in the list Power Save When Power Save is enabled the display will automatically switch off if none of the buttons have been pressed for 30 seconds This function is appropriate in...

Page 28: ...the total number of days for which statistics were collected are shown on both the Recording Display and the Stopped Display Figure 4 TREL30 16 recording showing 18 days of data has been collected Fi...

Page 29: ...e current day s maximum statistic l The Today marker flashes and 00 DAYS is shown to indicate that today s data is being displayed l Today no temperature values were recorded above the upper alarm thr...

Page 30: ...sterday was 58 9 C Pressing the button now displays yesterday s minimum statistic l still flashes and 01 DAYS still shows as we are still looking at yesterday s data l Yesterday no temperature values...

Page 31: ...minimum temperature recorded on the day before yesterday was 73 5 C Data from 7 days ago After pressing the button for a few times skipping days 3 to 6 the maximum statistic from seven days ago is di...

Page 32: ...by the upper alarm threshold marker l The duration was 5 hour 30 minutes l An alarm event was generated Neither duration from day 7 or 8 by itself would cause an alarm event however the combined dura...

Page 33: ...tton Instead of today s statistic data the first screen shows the maximum temperature recorded during the trip to date identified by the symbol User resettable Pressing the REVIEW MARK button again sh...

Page 34: ...iewing the Minimum temperature screen l Press and hold REVIEW MARK After 1 second the following screen sequence will show and then repeat once During the cycle the buzzer will sound a series of short...

Page 35: ...16 automatically stops recording temperature when the maximum number of readings specified during configuration has been reached Your unit can also be set up to stop automatically when it is downloade...

Page 36: ...or more than 2 seconds after the STOPPED symbol remains permanently on If an ALARM is present and the Clear alarm by pressing STOP button function is enabled you will need to clear the alarm first bef...

Page 37: ...now ready to be started again If the button is released while STOPPED is still blinking or you wait until the READY symbol disappears the display shows STOPPED again and the logger remains stopped Th...

Page 38: ...TREL30 16 can also be activated again by pressing and holding the and buttons together for a period of 6 seconds During this time the text not rdY appears and the READY symbol flashes Release the but...

Page 39: ...Hibernating a TREL30 16 TREL30 16 User Guide 39...

Page 40: ...T90 in moving air 1m s method as detailed in EN12830 2018 Clock accuracy Quartz crystal locked real time clock rated accuracy 25ppm 25 C equiv to 2 5 seconds day Rated temperature coefficient is 0 034...

Page 41: ...the battery When an alarm is triggered the alarm should be cleared and reset as soon as possible Size 93mm H x 54 5mm W x 8 6mm T Weight 43 g excluding probe Calibration Factory calibration using inst...

Page 42: ...lot l Remove the battery using a small bladed screwdriver as shown Insert the new battery Only use batteries from a reputable manufacturer Check the temperature range of the battery and make sure it c...

Page 43: ...ity value that lies above the upper or below the lower alarm threshold value Alarm Threshold Value The temperature or humidity value at which a reading is regarded as an alarm reading This can be an u...

Page 44: ...s are regarded as acceptable Non Alert Range Target temperature humidity range where the device does not trigger an alert P Primary Alarm The alarm threshold closest to the non alarm range in a multi...