Product Information Card

RHTemp2000

RHTemp2000

Temperature and Humidity Recorder with LCD

Product Quick Reference Card

Part Number

RHTemp2000

Temperature Sensor

Semiconductor

Temperature Range

–20 to +60°C

Temperature Resolution

0.1°C

Calibrated Accuracy

+0.5°C (0 to 50°C)

Humidity Sensor

Internal Semiconductor

Humidity Range

0 to 95%RH

Humidity Resolution

0.1%RH

Humidity Accuracy

+3%RH (+2%RH @ 25°C Typical)

Memory

131,071/channel

Sample Rate

2 seconds to 24 hours

Required Interface Package

IFC110 or IFC200

Baud Rate

115,200

Typical Battery Life

1 year with display off, 30 days with

continuous LCD and no backlight

Operating Environment

–20 to +60°C, 0 to 95%RH (non-condensing)

Material

Black Anodized Aluminum

Dimensions

4.8” x 3.3” x 1.25” (112mm x 84mm x 32mm)

Approvals

CE

Specifications subject to change.

See MadgeTech’s terms and conditions at www.madgetech.com

Device Maintenance

Battery Replacement

Materials:

3/32” HEX Driver (Allen Key)

Replacement Battery (U9VL-J)

- Remove the cover from the device by unscrewing the four 

screws.

- Remove the battery from its compartment and unsnap it 

from the connector.

- Snap the new battery into the terminals and verify it is secure.

- Replace the cover taking care not to pinch the wires. Screw 

the enclosure back together securely.

Note: Be sure not to over tighten the screws or strip the threads.

Battery Warning

WARNING: FIRE, EXPLOSION, AND SEVERE BURN HAZARD.  DO NOT 

SHORT CIRCUIT, CHARGE, FORCE OVER DISCHARGE, DISASSEMBLE, 

CRUSH,  PENETRATE  OR  INCINERATE.  BATTERY  MAY  LEAK  OR 

EXPLODE IF HEATED ABOVE 60°C (140°F).

Recalibration 

The RHTemp2000 standard calibration is one point at 25°C and 

two points at  25%RH and 75%RH.

Pricing:

Recalibration traceable to NIST 

$90.00

Recalibration 

 

 

$70.00

Additional:

As Found Data 

 

 

$15.00

Additional Points   

 

$15.00 per point

To send the devices back, visit www.madgetech.com, select 

Services

 

then 

RMA Process

.