background image

 

www.magnaflux.com 

Y-7 AC/DC Yoke Operating Instructions 

 

 

 

 

 

 

The Y-7 is an articulating leg magnetic yoke providing A.C. or D.C. magnetic fields for defect detection in 
ferromagnetic materials. The Y-7 has the capability of demagnetization, if required.  

Specifications: 

Power Source: 

  

The Y-7 is configured for 115 volt, 50/60Hz.1 phase, OR 220 volt, 50 Hz.1 phase. 

 

Current Draw: 

 115V version - 4 amps / 230V version - 2 amps.  

 
Mechanical Capacity:

    Minimum-maximum distance across poles: 1“ to 12".  

 
Magnetic Field:  

            A.C:  Constant level. D.C: Pulsed.  

 
Pull Force: 

 

The Magnaflux Y-7 yoke is designed to lift 10 lbs or 40 lbs as specified in  

 

ASTM E-444. Please refer to this specification for more details. 

 
Yoke Weight:   

 7.7 pounds.  

 
Controls: 

          

Electronic - Solid state contained within the housing.  

 
Manual:  

Momentary, push to test switch A.C./D.C. selector switch  

 
Duty Cycle:  

     

33% with 80 seconds "ON" maximum.  

Caution!  

1.  DO NOT connect the power cord to a battery for portable work-power source must be as 

noted on the nameplate.  

2.  DO NOT change the AC/DC selector switch while the unit is energized.  

 

Reviews: