background image

 

 

Part functions: 

(See figure A.)

 1

     Micro USB port:

 Used to insert or remove Quick disconnect cable.  

Cover during washing or rinsing. 

 2

     Quick disconnect cable:

 Keyboard cable can be removed from Micro 

USB port to disconnect keyboard. 

 3

    Port cover:

 Used to cover Micro USB port during washing or cleaning.

  
  

Installation:

 (See figure B.) 

 

  Cleaning Instructions 

 

   Clean keyboard with approved cleaning method and approved 

disinfectant or soap and water.

 

  

Approved cleaning methods keyboard

 

•    You can wipe the product with a soft cloth that is non-abrasive like a  

microfiber cloth, a disinfecting wipe, a soft brush similar to a paint  

brush in bristle stiffness or a sponge, but not a scourer of any sort.

 

•    You can spray the product with soap and water or an approved 

disinfectant. 

 

•    You can rinse the product with soap and water or an approved  

 disinfectant. Before rinsing, remove cable (2) from port (1) and 

cover port (1) with cover (3). Make sure cover (3) is fully inserted 

(See figure C). 

  

  

Approved cleaning methods drape

 

•    You can wipe, spray or rinse the product as described under 

“Approved cleaning methods keyboard“.

 

•     You can immerse the product in soap and water or an approved 

disinfectant.

 

•     You can expose the product to Hydrogen Peroxide Vapour (HPV).

 

•    You can sterilize the product in an autoclave or thermo disinfector 

(up to 134 ºC, tested 200 cycles).

  

Approved disinfectants 

 

•   Chlorine based, up to 100,000 ppm (10%) bleach content  

 

•   Alcohol based 

 

•   Glutaraldehyde based 

 

•   Phenol based 

 

•   Formaldehyde based, up to 370,000 ppm (37%) formaldehyde content 

 

•   Oxidizing, such as Hydrogen Peroxide, up to 30,000 ppm (3%) content 

 

•   Quaternary Ammonium, up to 20,000 ppm (2%) content. 

  

Restrictions

 

 

•     Only clean using the cleaning methods and disinfectants described 

under cleaning instructions in this manual.

 

•    Please note that the product must not be immersed. 

 

•     Always follow the instructions provided by the cleaning product 

manufacturer. 

 

•    Do not use petroleum based cleaning fluids such as Acetone or 

abrasive cleaning aids and products.  

 

•    Do not expose the product to temperatures below 32 ºF or above 158 °F. 

(below 0 ºC or above 70 ºC.) 

 

 

•    Do not use temperature or pressure sterilization.

PS/2

USB

 B

 A

Its Cool Flat

 (Replacement drapes sold separately)

 2

  3

 1

Its Cool

 3

 C

Its Cool Flat

Its Cool

• Operating Manual

Reviews: