FRONT PANEL

02

General Description

  This compact digital multimeter is designed to measure
  AC and DC Voltages, AC and DC current, Resistance, 
  Diode and to perform audible Continuity checks with
  accuracy and ease.
  Small and lightweight, with a carrying case and test leads
  wound on its Body, this instrument will provide you years 
  of satisfactory service.
  Auto power-off function extends the battery life. If no
  key-inputs happen around 30 minutes, this meter will be
  turned off automatically.

1.  Select Button 
    Momentary-type push 
    switch for measuring
    functions select.
2.  HOLD Button
    Momentary-type push 
    switch for data hold.
3.  Function Switch
    Rotary switch for 
    selecting functions.

03

M320C

AUTO RANGING

 POCKET-SIZED DIGITAL MULTIMETER

OPERATOR'S INSTRUCTION MANUAL

Maintenance

• Before opening case, always disconnect test beads 
  from all energized circuits.
• For continuous protection against fire, replace fuse only
  with ratings; F 400mA/300V (Quick Acting).  
• Never use the meter unless the back cover is in place 
  and fastened completely.
• Do not use abrasives or solvents on the meter. To clean
  it uses only a damp cloth and mild detergent.

Safety Symbols

05

Specifcation

Accuracy is guaranteed for 1 year,
23°C±5°C,less than 75% RH.

DC Voltage

AC Voltage

Overload protection: 300V DC or rms AC for all ranges
Input Impedance:10MΩ
Frequency range: 50Hz to 400Hz, 50 to 60Hz for 300V
range.
Response: Average responding, calibrated in rms of 
a sine wave

DC Current

 
Overload Protection: F 400mA/300V fuse      

Overload protection: 300V DC or rms AC for all ranges 
Input Impedance: 10MΩ

Accuracy

 ±0.8%  of  rdg ±4 dgts

  ±0.8%  of  rdg ±4 dgts

 ±0.8%  of  rdg ±4 dgts

Range
4V
40V
300V

Resolution
1mV
10mV
1V

Range

Accuracy

Resolution

40mA

±2.0% of rdg ±3dgts 

0.01mA

400mA

±2.0% of rdg ±3dgts

0.1mA

Range

Accuracy

Resolution

400mV

 ±0.5% of rdg ±3dgt

0.1mV

4V
40V

 ±0.5% of rdg ±3dgt

±0.8% of rdg ±3dgt 

1mV
10mV

300V

1V

±0.8% of rdg ±3dgt 

AUTO-RANGE
DMM

300V    400mA MAX

CAT III 300V 

M320C

HOLD    SELECT

OFF

mA

V

Hz

5

1

4

3

2

AUTO-RANGE
DMM

300V    400mA MAX

CAT III 300V 

M320C

HOLD    SELECT

OFF

mA

V

Hz

AC Current

Overload Protection:  F  400mA/300V  fuse.

Range

Accuracy

Resolution

40mA

±3.0%  of  rdg  ± 4dgts

0.01mA

400mA

±3.0%  of  rdg  ± 4dgts

0.1mA

Audible Continuity Test

        Built-in buzzer sounds when resistance 
        is less than 50±20Ω. 

 Overload Protection:250V rms ac

Range

   

 

Description

4

.  Test Leads

    Red test lead for 
    positive (+) and 
    black test lead for
    negative (-)
5.  LCD Display
    3 ¾ digits, 7 segment, 
    maximum 3999 counts.

Safety information

This meter has been designed according to EN61010-1,
EN61010-2-030,EN61010-2-033, 300V CAT III and 
pollution2.
Follow all safety and operating instructions to ensure the 
meter is used safely and is kept in good condition With 
proper use and care, your digital multimeter will give you 
years of satisfactory service.

During use

• Never exceed the protection limit indicated in the 
  specifications for each range of measurement.
• Never use the meter to measure voltages that might 
  exceed 300V above earth ground in category III 
  installations
• Always be careful when working with voltages above 
  60v dc or 30v ac rms. keep fingers behind the probe 
  barriers while measuring.
• Do not perform resistance measurements on live circuits.
• Inspect test leads and probes for cracks, breaks or 
  crazes in the insulation before using the meter
• If the equipment is used in a manner not specified by
  manufacturer, the protection provided by equipment may
  be impaired.

 

Important safety information ,refer to the 
instruction manual.

Fuse must be replaced with ratings specified 
in the manual.

 

AC (Alternating Current)

 

DC (Direct Current)

Earth(ground) TERMINAL

Equipment protected throughout by double insulation.

CAT III:

              Applicable to test and measuring circuits connected 
to the distribution part of the building's low-voltage MAINS 
installation.Example:fixed equipment switchboards,circuit 
breakers,wiring,including cables,bus bars,junction boxes,
switches,sockets,utput terminals on devices for industrial 
use and other equipment.

CAT III

300 V

European union directives

CONFORMS TO UL STD. 61010-1, 61010-2-030,
61010-2-033, 61010-031; CERTIFIED TO CSA STD.
C22.2 No.61010-1,61010-2-030, 61010-2-033, 61010-031.