background image

DM

9600

CHP

USER MANUAL

9122EN-CHP

Summary of Contents for DM 9600 CHP

Page 1: ...DM 9600 CHP USER MANUAL 9122EN CHP ...

Page 2: ... MRU No part of this manual my be published in any form print fotocopy electronic media or any other publication form without a written approval by the publisher All user trade marks and name mark descriptions even those which are not marked as such are properties of the respective owners Edition 2018 07 03 V01 ...

Page 3: ...o off function 11 5 Settings 12 5 1 Date and time 13 5 2 Bluetooth 13 6 Measurement operation 14 6 1 Preparation for each measurement 14 6 1 1 Power supply 14 6 1 2 Measuring with grid power supply Battery charging 14 6 1 3 Battery charge condition 14 6 1 4 Operating temperature 14 6 2 Performing a measurement 15 6 2 1 Auto Log 15 6 3 Pressure measurement 17 6 3 1 Performing a measurement 18 6 3 2...

Page 4: ...SV 25 8 4 2 Exporting measurements 25 8 4 3 Information about the data storage 25 9 Extras 26 9 1 Factory settings 26 9 2 Service values 26 9 3 Service menu 26 9 4 Device info 26 10 Appendix 27 10 1 Specifications 27 10 1 1 Accuracy of measurement 27 10 1 2 Data communication 28 10 1 3 Technical data 28 10 1 4 Device trouble shooting 28 10 2 Firmware update 29 10 2 1 How to install a new firmware ...

Page 5: ...g the instructions in this manual y y Always keep the manual near you when working with the analyzer to be able to read instructions as needed Please ensure to hand over all documents to when handing the analyz er over to others 1 1 Intended use The DM9600 CHP is designed for pressure measurements as well as temperature measurement of NON explosive gases The instrument was manufactured according r...

Page 6: ... and marketing high quality emission monitoring analyz ers MRU GmbH produces a wide range of instruments from standard analyzers up to tailor made industrial analyzers Plant 1 Sales Service R D Plant 2 Production MRU GmbH Fuchshalde 8 12 74172 Neckarsulm Obereisesheim GERMANY Fon 49 71 32 99 62 0 Zentrale Fon 49 71 32 99 62 61 Kundendienst Fax 49 71 32 99 62 20 Email info mru de Internet www mru e...

Page 7: ...ions 2 2 Safety precautions The used category s of safety precautions are here explained once more Identifies an immediate impending hazard that if ignored will result in severe bodily injuries or death DANGER Identifies an immediate impending hazard that if ignored may result in severe bodily injuries material damage or death WARNING Identifies a possibly dangerous situation that if ignored may r...

Page 8: ...s representatives 3 2 The measuring instrument The measuring instrument has a compact and robust fiber reinforced plastic enclosure It is entirely operated using the color touch screen Strong magnets on the backside are fixing the device on metallic parts 1 Touch screen 2 Pressure port P2 3 Pressure port P1 4 Mini USB interface 5 Temperature port Temp 1 6 Temperature port Temp 2 7 AUX port 8 IR in...

Page 9: ... 2 Scroll bar indicator for the main menu 3 Function keys such as scroll edit store 4 State of SD card Green symbol read and write is possible Yellow symbol only read SD card locked No symbol no card in reader 5 Main menu bars 6 Function key such as return 7 Battery charging condition Lithium Ion Battery 3 3 Menu structure The measuring instrument has 3 available main menus 1 6 5 3 7 2 3 4 ...

Page 10: ...main menus Menu Measure ment All available measurement options will be displayed and can be selected here This screen will only display installed and available options Menu Storage All available storage options will be displayed here Menu Extras All other options are displayed here for example setting to adapt your device ...

Page 11: ... Ö The device is ready for use 4 3 Switch off 4 3 1 Switch off f f Touch the Context menu on the display f f Push the Button Ö Ö Switch off the device now f f Push Yes switch off Ö Ö The device will switched off 4 3 2 Automatic switch off Auto off function The Auto off function powers down the device in the main menus Mea surement Storage und Extras if no key has been touched for 60 minutes During...

Page 12: ...king according to their regulations Language Select the operating language LCD brightness 5 100 The LCD brightness is depending on temperature and personal perspective of each user 50 brightness are normal at 68 F 20 C Keyboard beep ON OFF Turn the Keyboard beep ON or OFF Helping hints ON OFF Turn the feature Helping hints ON or OFF Temperature unit C F Select the units for pressure and tem peratu...

Page 13: ...letely drained f f Use the arrow buttons up down and left right to adjust the date and time f f In case the battery is discharged it might be necessary to correct the date and time settings 5 2 Bluetooth The Bluetooth mode can be viewed and selected in the Bluetooth menu Please use BT CL for connections to Android and BT LE for connections to IOS The full version of the MRU Bluebottle Software MRU...

Page 14: ...ged 6 1 3 Battery charge condition The battery symbol indicates the battery capac ity The battery symbol will start to flash every second in red once the battery reaches a remain ing operation time of 60 minutes depending on device configuration The device will shut down if the batteries are not charged in time to prevent a total battery discharge 6 1 4 Operating temperature The device will displa...

Page 15: ...lows you to define the measuring time and the measuring interval Auto Log settings Measuring time is the complete time for one measurement cycle f f Press the button measuring time to be able to set the measurement time using the arrows up and down to change the time Interval means after each interval duration a measurement is made f f Press the button interval to select how frequently you want to...

Page 16: ...rom the data storage menu f f First select a site The Auto Log can save results either on the internal memory or onto the SD Card f f Press the Store button and then select where the results should be stored f f Please make sure that there is enough storage space available f f Press the start button to start the Auto Log it will end automatically after the defined measurement time has elapsed ...

Page 17: ...e and the empty field f f Edit the text with the arrow keys f f And this way procedure with the name of the existing programs ATTENTION Damage to the device from improper operation Damage from exceeding the measuring ranges f fObey the measuring range of the pressure sensor NOTE The following Error message Pressure too high will be displayed and there will also be an additional acoustic warn signa...

Page 18: ...en saving results the max and min values are saved as well as the pressure average 6 3 2 Zeroing You can zeroing the pressure sensor at every measuring 6 3 3 Context menu You can configure the settings for each measuring auto Log Measurements are being logged only in graphic mode zero point The pressure value is set to zero store Single measurements are being stored print out Measurements are beei...

Page 19: ...s f f Scroll with the arrow keys between the displays Ö Ö On the following displays the temperature gradient will be presented graphically The temperature can be auto logged see auto Log function When sav ing results the max and min values are saved as well as the temperature average 6 4 2 Kontextmenü You can configure the settings for each measuring NOTE The measurement accuracy can only be grant...

Page 20: ...ment of the sensors at an MRU service department www mru eu are recommended for the preser vation of value 7 2 Care The Analyzer needs only low maintenance effort for long value preserva tion If the analyzer is not being used over a period it is recommended to charge the battery before storing it and then recharge the battery every 4 weeks ...

Page 21: ... viewed in the menu view measurements Once selected you will see an overview of the different measurement modes with the number of saved results for each mode In case of a logged data sequence the total duration interval and num ber of saved measurements are displayed Der Leseprozess der Note that the process of reading data may take some time when a lot of data records are stored f f Select Press...

Page 22: ...r modify sites f f Select new to add a new site Ö Ö The first line which must have a distinct site number for identification The device can also create a new site number the next available site number will be selected All other lines are user definable lines for address customer name and so on f f Select modify in the context menu Ö Ö The site will be displayed in red and is now editable f f Selec...

Page 23: ... sites In the menu Sites administration each site will be displayed with the distinct site number and 6 user definable fields You can page through the sites using the arrow buttons left and right 8 3 3 Delete sites You can delete single sites in the menu Sites administration by selecting delete in the context menu or you can delete all sites at once You will be asked to confirm your intention to d...

Page 24: ... start with a comma both would result in an empty site number which is not permitted will be transferred Each line and or data set a maximum of 7 lines will be transferred A maximum of 24 numbers and letters are allowed for each line everything longer than 24 letters numbers will be cut off Examples for sites Site Spelling 7 Lines A1 F1 A1 F2 A1 F3 A1 F4 A1 F5 A1 F6 A1 F7 A1 F8 A1 F9 2 Lines 1 and...

Page 25: ...0 however the name must be changed to anlagen CSV before it can be used for import 8 4 2 Exporting measurements This function is used to export measurements to a PC The created file has the file name TMPxxxxx csv the xxxxx in the file name are consecutive numbers containing zeros The created file has a header with information like Site number Date time measuring values and units all the informatio...

Page 26: ...sing the values in the service vales menu The service values of all installed sensors are displayed here Please contact your local MRU service partner in case you have a mal function of your device It s always handy to have the device next to you when make a service call most likely the technician will ask you for these values and the serial number of the device 9 3 Service menu The service adjust...

Page 27: ...rmocouple type 2 Range 40 C 1 200 C Accuracy 1 C 0 5 Velocity optional 10711 Only in combination with DM9600 75 hPa 912200 based on differential pressure mea surement with Pitot tube v Measuring range differential pressure 75 hPa Accuracy differential pressure 0 5 Pa 1 with 5 C temp change 30min meas time Measuring range 1 m s 100 m s Accuracy without error of Pitot tube 1 m s 0 v 2m s 0 2 m s 2 v...

Page 28: ...ness 20h Display 2 8 touch TFT Housing material PA6GF30 IP degree of protection IP30 Weight 340 g Size 83 x 38 x 180 mm Memory sites 500 Memory Measures 1 000 10 1 4 Device trouble shooting Fault indication Possible causes Repair Device shows no reaction Device doesn t react to any touch screen commands f fPress reset button Display notice Device too cold or A beep every 5 sec onds e g Device was ...

Page 29: ...0 48 We will need the following information from you in case something has gone wrong with the update f f Please note the version and the current firmware of your device 10 2 2 Performing and verification of an update Preparing a SD card Typically you will receive an update by email You will have to unzip the file in case you have received the file in a zip format you would do this on your PC Now ...

Page 30: ...ny keys during the update f f Start the device again after the update f f Confirm the information Firmware update was performed with OK How can I verify that the update was successfully f f Turn on the device Select the context menu extras device info Ö Ö In the 3rd line you will see the current installed firmware version for example 0 90 48 What can I do if the old firmware version number is stil...

Page 31: ...MRU 31 32 DM9600 CHP Version 01 Declaration of conformity 11 Declaration of conformity ...

Page 32: ......

Reviews: