PoolPro PS6FC

E

 Quick Start Guide

 

Handheld Conductivity, Mineral / Salt Concentration, TDS, ORP, Free Chlorine, pH and Temperature Meter

 

 

 

PS6QSI Rev: 07-16 

This  Quick  Start  Guide  is  for  the  Myron  L

®

  Company  Pool  Pro  Model  PS6FC

E

.    Download  the  full  PS6FC

E

  Operation  Manual  (P/N  PS6FCEOM)  from 

www.myronl.com  to  get  a  more  detailed  set  of  instructions  for  taking  measurements,  storing  and  recalling  measurements,  changing  instrument  settings, 
maintenance procedures, calibration procedures, Troubleshooting Guide and the FACTORY CAL reset procedure. 

 

 

QUICK REFERENCE INSTRUCTIONS

 

 

Press and release any one of these buttons to turn on the PS6FC

E

 and begin taking measurements.  

 

Press and release the CAL/MCLR button to calibrate (CAL) the active measurement parameter or press and hold to clear displayed memory location (MCLR). 

 

Press and release to store (MS) a measurement or move upward on the list of stored measurements (Location 1 to 100) while in memory recall mode. 

 

Press and release to view the most recent stored measurements or move downward on the list of stored measurements (Location 100 to 1).  Also used 
to enter Settings Mode. 

 

 

PS6FC

E

 

 FEATURES 

1.

 

TEMPERATURE SENSOR – 

Houses thermistor for making temperature measurements. 

2.  CONDUCTIVITY CELL

 – 

Contains electrodes that generate a flux field in defined area for Conductivity, TDS, and 

MIN/SALT, 

 

3.  PH SENSOR 

(user replaceable) – 

Measures test sample’s hydrogen ion concentration

. The reference junction is 

located under the glass bulb. 

4.

 

pH SENSOR PROTECTIVE CAP

  – Snaps in place to protect pH Sensor when not in use and seals in Storage 

Solution. 

5.

 

LCD DISPLAY – 

Displays measurements, units of measure, active measurement mode, current solution mode setting, 

 low battery warning (LOBATT), and other information (see below).

 

6.   MEASUREMENT KEYS

 – Press and release any of these to turn on the PS6FC

E

 

 and begin taking measurements. 

7.  FUNCTION KEYS

 – 

 

Press and release to store the current measurement (MS), recall a stored measurement (MR). 

Also used to move between memory locations and to select various mode settings

8.  CAL/MEMORY CLEAR KEY

 – Use to calibrate (CAL) the active measurement parameter, enter  or clear the contents 

of the currently displayed memory location (MCLR). 

9

BLUDOCK SYMBOL 

– Indicates that this PoolPro PS6FC

E

 is equipped with Myron L

®

 Company’s optional wireless 

transceiver for wireless data download. 

 

LCD DISPLAY 

1.  MAIN VALUE DISPLAY 

– Displays the current measurement value or settings menu information. 

2.  UNITS OF MEASURE ICONS

 – Appropriate icon appears to show the units of measure for the current measurement. 

3.

 

TEMPERATURE UNITS OF MEASURE

 – Displays °C (Celsius) or °F (Fahrenheit) based on setting selected 

for Temperature readings. 

4.

 

MEASUREMENT ICONS

 – Appropriate icon appears to show what type of measurement is being made. 

5.

 

BUFFER ICON 

– Appears when the PS6FC

E

 is in pH Calibration mode to indicate that the instrument is 

expecting a pH buffer solution. 

6.

 

TEMPERATURE DISPLAY

– Displays the temperature of the sample solution, memory location number  

(1 to 100) or buffer during pH calibration. 

7.

 

SOLUTION MODE ICONS

 – Appropriate icon appears to indicate the current solution temperature 

compensation mode setting for Conductivity, Min/Salt and TDS measurements. 

8.

 

LOBATT ICON

 – Appears when the PS6FC

E

 

’s battery requires replacement. 

9.  CAL ICON – 

Appears when the PS6FC

E

 

 is in Calibration mode. 

10.

 

MEMORY ICON

 – Appears to indicate that the values and icons being displayed are for a measurement stored in the PS6FC

E

 

’s memory and are NOT a 

live measurement. 

 

PS6FC

E

 

 DEFAULT SETTINGS 

Temperature Units:

 °C  (Temperature is displayed in Degrees Celsius)     

Solution Modes: Conductivity – 

KCl

,  TDS – 

442

; MIN/SALT –

NaCl

OPERATING INSTRUCTIONS 

MEASUREMENT SETUP: 

Before you take a reading, make sure the 

PS6FC

E

 

 is clean, calibrated, and if measuring COND or TDS, that the 

desired solution mode has been selected (See Section II of Operating 
Instructions, below). The sample solution must also be within the specified 
range. 

I. MEASUREMENT 

1.

  Rinse and empty the appropriate sensor (conductivity or pH / ORP) 3 

times with sample to be measured. 

2.

  Fill with test sample. 

3.

  Press appropriate Measurement Key: 

COND to measure CONDUCTIVITY 
0-9999 MICROSIEMENS (µS/cm), 10-200 (mS/cm) 

 

 
TDS: Measures TOTAL DISSOLVED SOLIDS 
0-9999 parts per million (ppm), 10-200 (ppt) 

 

MIN / SALT: Same as TDS but displays in  
10-200 parts per thousand (ppt). 

 

displays as PPT of NaCl 

ORP: Measures Oxidation Reduction Potential (REDOX) ± 999 mV  

pH: Measures 0-14 pH 

NOTES:  

  When finished making Conductivity, TDS, or MIN/SALT measurements, 

rinse the cell cup with clean water (preferably DI, RO, or distilled). 

  When finished measuring pH, ORP or FC

E

, rinse sensor well with clean 

water (preferably DI, RO, or Distilled), refill it with Myron L Storage 
Solution then reinstall protective cap. 