background image

UROSHIELD

®

QUICK GUIDE

Please read the complete UroShield manual before using

the device and pay attention to provided warnings.

Clip-on

Actuator

Urinary catheter

(Not included)

The UroShield can be used with Foley catheters made

of any material in sized 14 - 22 French, up to 30 days

Driver

ON/OFF

button

Important Notes

UroShield should only be installed on a newly

inserted catheter

The UroShield should not be used to treat an

active urinary infection

Gently pull on the catheter until

you feel slight resistance.

Peel off the protective strips

from the inside of both halves

of the actuator.

Place the actuator on the catheter

about 3 cm from the point the catheter

exits the body, with the connection

cable towards the patient.

Align the catheter with the

actuator grooves and close it until

both sides snap together.

Connect the actuator cable to

the device cable in correct

alignment.

Connect the supplied charger to the

driver charging port and plug the

charger to the mains socket.

Press the device ON/OFF button for about 2 seconds until beeping sound is heard.

The device requires a few seconds to initialize and then displays the information screen.

1

2

3

4

5

6

7

Reviews: