CALIBRATION PROCEDURE
NI 9230/9232
This document contains the verification and adjustment procedures for the NI 9230 and NI 9232.
For more information on calibration, visit
Contents
Software.................................................................................................................................... 1
Documentation.......................................................................................................................... 2
Test Equipment......................................................................................................................... 3
Test Conditions......................................................................................................................... 3
Initial Setup............................................................................................................................... 3
Verification............................................................................................................................... 4
EEPROM Update ..................................................................................................................... 6
Reverification ........................................................................................................................... 6
Accuracy Under Calibration Conditions .................................................................................. 6
Worldwide Support and Services ............................................................................................. 7
Software
Calibrating the NI 9230/9232 requires the installation of NI-DAQmx on the calibration system.
Table 1 shows the necessary versions for each connector variation.
You can download NI-DAQmx from
. NI-DAQmx supports LabVIEW,
LabWindows
™
/CVI
™
, C/C++, C#, and Visual Basic .NET. When you install NI-DAQmx you
only need to install support for the application software that you intend to use.
Table 1.
Calibration System Software Requirements
Connector Variation
Version Requirements
Screw-Terminal
NI-DAQmx 14.5 or later
BNC
NI-DAQmx 15.5 or later