INSTALLATION GUIDE

NI PXI-1042HVDC DC Cable Kit

The PXI-1042HVDC DC cable is designed to connect the PXI-1042HVDC 
chassis to a DC power source. The cable consists of a DC power connector, 
18 feet of 10 AWG cable, and a high heat, impact resistant, water tight fuse 
holder with a 20 A fuse installed. For maximum DC power source 
protection, the fuse is located as close to the DC power source as is 
practical.

The user must install the appropriate connectors required to connect the 
cable to their DC power source.

Caution

Observe proper polarity when connecting to a DC power source. Refer to 

drawing CD_192057-01 for more detail.

Fuse Replacement

Caution

Disconnect the cable from the PXI-1042HVDC chassis and any power source 

before replacing the fuse or connecting the fuse holder to the DC cable.

Complete the following steps to replace the fuse:

1.

Unscrew the two halves of the fuse holder body by turning one half in 
a counter-clockwise direction.

2.

Replace the fuse with an F20A, 500 VDC/600 VAC, 1.5 in.

×

.41 in. 

(10 mm

×

38 mm) Midget, fast-acting fuse.

3.

Align the mating tabs on one half of the fuse holder body with the slots 
on the other half, and screw the two halves back together in a clockwise 
direction. Hand tighten only.

Reducing Cable Length

National Instruments recommends using the cable at its designed length. If 
it is necessary to shorten the cable for a particular application, the following 
guidelines should be observed.

Caution

Do 

not

 leave the fuse holder out of the cable assembly.