INSTALLATION INSTRUCTIONS
NI TB-2666
Terminal Block for the NI PXI-2566
This guide describes how to install and connect signals to the
National Instruments TB-2666 terminal block. Refer to the
NI Switches
Getting Started Guide
to determine when to install the terminal block.
Introduction
The TB-2666 terminal block installs in front of the PXI-2566 switch
module and has screw terminals that provide access to each of the 16 relays.
Connections for the trigger output and trigger input signals also are
available.
Caution
This terminal block is rated for Measurement Category I and intended to carry
signal voltages no greater than 150 V. This module can withstand up to 800 V impulse
voltage. Do not use this module for connection to signals or for measurements within
Categories II, III, or IV. Do not connect to MAINS supply circuits (for example, wall
outlets) of 115 or 230 VAC. Refer to the
NI Switches Getting Started Guide
for more
information on measurement categories.
When hazardous voltages (>42.4 V
pk
/60 VDC) are present on any relay terminal, safety
low-voltage (
≤
42.4 V
pk
/60 VDC) cannot be connected to any other relay terminal.
Conventions
The following conventions are used in this guide:
This icon denotes a note, which alerts you to important information.
This icon denotes a caution, which advises you of precautions to take to
avoid injury, data loss, or a system crash.