Serial Label 

X2-Cell Quick Start 

For best results, setup your new system in the office/lab. Collect a few data 

points and become familiar with the system prior to field deployment. 

1

 

Go to 

V2.WQDataLIVE.com/Getting-Started

 

2

 

Create a new account or sign into an existing account and enter the 

Claim Code

 found on the serial label to access the device’s data 

3

 

 Place the X2 in an area with 

strong cellular coverage and 

connect the antenna 

4

 

Remove the center power 

plug and required outer port 

plugs (1 for each sensor to 

connect) 

5

 

Connect sensors to any outer 

port (8 pin) 

6

 

Connect power (5 to 24VDC) 

to the center port (6 pin) 

7

 

Wait up to 60 seconds for the blue cellular status LED to turn on  

 

If the blue LED repeatedly blinks three

 

 times, the cellular strength is 

sufficient to connect the X2 with WQData LIVE 

 

If the blue LED repeatedly blinks once, cell coverage is low. Move the 

X2 to a location with better cell coverage and cycle power 

8

 

Allow up to 5 minutes for sensor detection and data acquisition. The 

detected X2 profile is sent to WQData LIVE along with new data 

 

The blue LED will be on (solid) when setup is completed 

9

 

Refresh the web browser and confirm that all connected sensors and 

their parameters are displayed on WQData LIVE 

 

 

 

 

 