NOX W7 LINK KITS

Instructions for Use

Introduction

The Nox W7 LINK KITs come in three different versions and can be

used with the Nox recording devices when configured for an ambulatory 

study. Note there is a slight difference in the configuration dependent on 

if the Nox T3 recorder or Nox A1 recorder is used (Refer to section 

Instuctions for Use).

 Nox W7 LINK KIT - A

    To be used with: 

         •  Resmed AirSe

nse™10

• Nox W

7 LINK KIT - R 

    To be used w

ith: 

        

  •  Radiometer TCM4, TCMCombiM, TCM40, TCM TOSCA

         •  Nonin RespSense

         •  Resmed S9™

• Nox W7 L

INK KIT - S 

    To be used with: 

         •  SenTec Digital Monitor (Software SMB07.0x and SMB08.00)

The Nox W7 LINK KIT

 packages include:

• 

Nox W7 LINK (Uncord Serial Port Plug F2M01SXA) 

• Power Adapter (FW7721M/05) 

• USA Plug Adapter 

• European Plug Adapter 

• UK Plug Adapter 

• Nox USB Mini Cable 

• DB9 Male to Male Mini Null Modem

• Nox User Instruction Leaflet 

Intended Use

The Nox W7 LINK is a Medical Device Data System (MDDS) and allows for 

electronic conversion and wireless transfer of medical device data between 

an auxiliary medical device and a Nox recording device. The Nox W7 LINK 

is connected to a serial port of an auxiliary medical device. Before the data 

is transmitted over Bluetooth® link to the applicable Nox recording device 

it is converted from serial to Bluetooth format.

Scope

This leaflet is designed to instruc

t users on how to use the Nox W7 

LINK with supported auxiliary devices.

Warnings and Cautions

• Warning: Do not use damaged equipment, sensors or accessories. 

  This may result in bad performance of the Nox W7 LINK or patient/

  operator injury.

• Warning: The 

Nox W7 LINK

 is not certified to be used for continuous 

   monitoring where failure to operate can cause injuries or death of the 

   patient.

• Warning: The RF function of the Nox W7 LINK might influence the 

  performance of other devices in close proximity and lead to potential 

  serious harm.

• Warning: Only use the medical grade power adapter provided by Nox 

  Medical as part of the Nox W7 LINK KIT package to power the Nox W7 

  LINK. The use of other power adapters may result in electric shock or 

  overheating of the power adapter/burns.

• Warning: Use of the Nox W7 LINK with auxiliary devices that have not 

  been validated by Nox Medical may result in electric shock. Always use 

  the Nox W7 LINK with validated auxiliary devices. Auxiliary devices need 

  to comply with applicable IEC 60601-1 safety standards for medical 

  electrical equipment.

• Warning: Do not use the Nox W7 LINK KITs in a Magne

ti

c Resonance 

  Imaging (MRI) environment. That might lead to serious patient harm.

Instructions for Us

e

1. Preparing a Study with the Nox W7 LINK

When configuring a No

x recording device for an ambulatory study with the 

Nox W7 LINK in the Noxturnal software, the applicable Nox W7 LINK is 

added by entering its Bluetooth address in the relevant field. Two different 

addresses are used, depending on if the Nox A1 recorder is being used or 

the Nox T3 recorder:
Nox A1 Recorder: The Nox W7 LINK’s Bluetooth address (BDA) is used and 

has 12 digits. The BDA can be found on the Nox W7 LINK’s label.
Nox T3 Recorder: The Nox W7 LINK’s serial number (SN) is used and has 9 

digits. The SN can be found on the Nox W7 LINK’s label. 

2. Connecting to a Resmed S9™/AirSense™10 Device

The Nox W7 LINK is connected to the serial interface output located on 

the Resmed Embletta Adapter. The Nox W7 LINK has thumbscrews which 

can be used for more secure placement. For thumbscrews use a 4mm flat 

screwdriver.

Note: The supported modes are AutoSet and CPAP.

3. Connecting to a Radiometer Device

The Nox W7 LINK KIT - R includes a DB9 male to male mini null modem. 

Connect the null modem to the Nox W7 LINK, use thumbscrews for secure

placement. For thumbscrews use a 4mm flat screwdriver. 

Connect the applicable serial port extension cable (available from 

Radiometer) to the Radiometer device and connect the Nox W7 LINK 

to the serial interface located on the extension cable.

4. Connecting to

 a Nonin RespSense Device

The Nox W7 LINK KIT - R includes a DB9 male to male mini null modem. 

Conne

ct the null modem to the Nox W7 LINK. Connect the Nox W7 Link 

to the serial interface output located on the side of th

e Nonin RespSense. 

Use the thumbscrews for secure placement. For thumbscrews use a 

4mm flat screwdriver.  

5. Connecting to SenTec Digital M

onitors

The Nox W7 LINK KIT - S includes a DB9 mal

e to male mini null modem. 

The null modem is connected to the serial interface output located at the 

back of the SenTec device. Connect the Nox W7 LINK to the null modem. 

Use thumbscrews for more secure placement. For thumbscrews use a 4mm 

flat screwdriver.

6. Starting a Study

a. Connect the Nox W7 LINK to the applicable auxiliary device as instructed 

    above (Step 2, 3, 4 or 5) and make sure the Nox Mini USB cable is 

    connected to the NOX W7 LINK and the Power Adapter (FW7721M/05).

b. Turn on the auxiliary device. The blue LED on the link starts blinking.    

c. Turn on the Nox recording device.

d. The green LED on the NOX W7 LINK starts blinking, indicating Bluetooth 

    connection with the Nox recording device. 

    Note: Establishing Bluetooth connection may take up to 2 minutes.

e. Now a Bluetooth connection has been established between the Nox 

    recording device and the auxiliary device. The auxiliary device’s signals 

    are being recorded.

 

States of the Nox W7 LINK’s LEDs

The NOX W7 LINK is equipped with two LEDs. The green LED indi

cates the 

status of the transmission. It blinks when the device is communicating with 

the Nox recording device. It is steady when the device is on standby. The 

blue LED indicates that the auxiliary device is streaming data to the device.

Maintenance

When the Nox W7 LINK is not in use, disconnect it from auxiliary 

devices and store in a dry place. 

Environmental Conditions

Operation:

           

Temperature:                0 °C to +40 °C (+32 °F to +104 °F)

Relative Humidity:         15-95% (non-condensing)

Atmospheric Pressure:   700 hPa - 1060 hPa

Storage: 

                 

Temperature:                -20 °C to +70 °C (-4 °F to +158 °F)

Relative Humidity:         10-95% (non-condensing)

Atmospheric Pressure:   700 hPa - 1060 hPa  

Cleaning

The Nox W7 LINK tolerates cleaning with a soft cloth moistened with a 

mild soap solution.

Technical Specifications

Nox W7 LINK

• Physical Size: 36x17x60 (WxHxL) mm

• Voltage Range: 5-30V DC

• Current Consumption (max): 300mA

• Baud Rate: Configured up to 230400 baud/s

• Operating Frequency: 2402-2480 MHz

• Transmit Power: -8 to 16 dBm

• Sensitivity at 0.1% BER: -90 dBm

• Range: 10m in line of sight when used with Nox recording devices

Mini Null Modem 

• DB9 Male to Male Mini Null Modem Adapter

Power Adapter (FW7721M/05) 

• Input voltage 100 to 240 V AC (± 10%)

• Input current 200 to 100 mA

• Frequency 50 to 60 Hz

• Efficiency 80% typ. at full load

• Output 5V DC/1.5A

• Exchangeable plug system: European, UK, USA

Nox USB Mini Cable 

• Mini-B plug

• Standard-A plug

Disposal

According to the regulation in Europe on 

Waste of Electrical and Electronic 

Equipment (WEEE) the equipment may not

be disposed of as unsorted municipal waste. 

The equipment shall be collected separately 

and returned to the appropriate collection 

system available. Please contact your 

sales representative regarding take-back or 

recycling of the equipment.