background image

OPERATING INSTRUCTIONS

OAKTON

®

35630-00, 35630-02

Portable pH/CON 10

and Waterproof 

pH/CON 10 Meter

Printed in Singapore. 06/00

68X230403 00702-09 rev 06/00

pH/Conductivity/ C Meter

HOLD MODE

ON

OFF

CAL

MEAS

 C

MEAS

pH

ATC

pH/Con 10 Series

ENTER

12. Warranty

OAKTON warrants this meter to be free from significant deviations in material 
and workmanship for a period of one year from date of purchase. OAKTON
warrants this probe to be free from significant deviations in material and 
workmanship for a period of six months from date of purchase. If repair or 
adjustment is necessary and has not been the result of abuse or misuse within 
the warrantied time period, please return—freight prepaid—and correction will 
be made without charge. OAKTON alone will determine if the product problem 
is due to deviations or customer misuse.

Out-of-warranty products will be repaired on a charge basis.

13. Return of items

Authorization must be obtained from our Customer Service Department before
returning items for any reason. When applying for authorization, please include
data regarding the reason the items are to be returned. For your protection, items
must be carefully packed to prevent damage in shipment and insured against possi-
ble damage or loss. We will not be responsible for damage resulting from careless or
insufficient packing. A restocking charge will be made on all unauthorized returns.

NOTE: We reserve the right to make improvements in design, construction, and

appearance of products without notice.

pH/Conductivity/

°

C Meter

pH/CON 10 Series

WATERPROOF

MODE

HOLD

ENTER

RANGE

ON

OFF

CAL

MEAS

ATC

READY

MEAS

 C

µ

S

Summary of Contents for 35630-00

Page 1: ...will be made without charge OAKTON alone will determine if the product problem is due to deviations or customer misuse Out of warranty products will be repaired on a charge basis 13 Return of items Au...

Page 2: ...ting 19 9 Error Messages 20 10 Specifications 21 11 Accessories 22 23 12 Warranty 24 13 Return of Items 24 2 3 pH Conductivity C Meter HOLD MODE ON OFF CAL MEAS C MEAS pH ATC pH Con 10 Series ENTER 1...

Page 3: ...nding graphic indicator on the LCD ON OFF Powers and shuts off the meter HOLD Freezes the measured reading To activate press HOLD while in measurement mode To release press HOLD again MODE Selects the...

Page 4: ...s a notched 6 pin connector to attach the probe to the meter NOTE Do not substitute other probes or electrodes For a replacement probe see the Accessories section page 22 NOTE Keep connector dry and c...

Page 5: ...bration See section 4 5 on pages 14 15 for temperature calibration Preparing the Meter for Calibration The pH conductivity temperature probe included with this meter is designed for use with this mete...

Page 6: ...ate the pH standard buffer solution See figure NOTE If using a pH buffer other than pH 7 press the or keys to scroll up or down until the secondary display value is the same as your pH buffer value pH...

Page 7: ...containers 2 If necessary press the MODE key to select Conductivity Mode The S or mS indicator will appear on the right side of the display 3 Rinse your probe with deionized water then rinse the probe...

Page 8: ...and the secondary display shows you what the temperature value was initially See figure 4 5 ATC CAL Temp C B A B pH ATC CAL pH pH A 4 Compare the primary display reading to a NIST traceable thermomete...

Page 9: ...electrode has dehydrated soak it for 30 minutes in OAKTON electrode storage solution or a 2M 4M KCl solution 2 Switch on the meter The MEAS annun ciator appears on the top center of the LCD The ATC i...

Page 10: ...a Clean probe See Probe Care Maintenance p 18 18 7 Probe Care and Maintenance Under typical operating conditions the probe will need to be replaced every 6 to 12 months In extreme applications the pr...

Page 11: ...Error Messages LCD Indicates Cause Solution Display Err Unrecognized input Wrong input in Release key Select annunciator from keypad selected mode valid operations depending on mode CAL Err Calibratio...

Page 12: ...D 00653 04 Electrode storage solution 1 pint bottle Keeps electrode bulb moist for faster more accurate readings Shpg wt 1 1 lbs 510 gms WD 00653 06 Electrode cleaning solution 1 pint bottle Removes b...