background image

With meter  

powered off, press 

HOLD/ENT and  

ON/OFF at the  

same time. First 

release ON/OFF 

button, then  

HOLD/ENT button.

After buffer 

selection, whole 

LCD segment 

lights up and 

goes back to 

measurement 

mode.

Figure 1: Buffer Selection Sequence

Figure 5: User Reset Sequence

With meter 

powered off, press 

CAL and ON/OFF at 

the same time. First 

release ON/OFF 

button, then  

CAL button.

After reset, whole 

LCD segment lits 

up and goes back to 

measurement mode.

Before You Begin 

Your instrument has been factory calibrated and usually works well out of  

the box. However, after extended periods of non-use, it’s best to soak the  

sensor in electrode storage solution, pH buffer for 10 minutes or so. A brief 

rinse with deionized (DI) water is OK, but avoid soaking or storing in deionized 

water as this will shorten the pH electrode life. Prior to taking measurements, 

periodic calibration with certified standards is recommended for best accuracy.

pH Buffer Set Selection

pHTestr features USA (pH 4.01, pH 7.00 and pH 10.01) or NIST (pH 4.01,  

pH 6.86, and pH 9.18) standards. Select either one to suit your requirements. 

1.  While pressing the HOLD/ENT button, switch on the unit by pressing the 

ON/OFF button.

2.  Release the HOLD/ENT button. The display will flash either USA or NIST.

3.  Press CAL button to toggle between the two buffer set standards.

4.  Press the HOLD/ENT button to confirm the selection of the buffer set.

pH Calibration

Calibration should be done regularly, preferably once a week. You can calibrate 

up to three points using either the USA or the NIST buffer set standards.

1. Press ON/OFF button to switch unit on.

2.  Dip electrode about 2 to 3 cm into the pH standard buffer solution.

3.  Press the CAL button to enter calibration mode. The ‘CAL’ indicator will be 

shown. The upper display will show the measured reading based on the last 

calibration while the lower display will indicate the pH standard buffer solution.

Note: To abort calibration, press the CAL button. 

4.  Allow about 2 minutes for the reading to stabilize before pressing the  

HOLD/ENT button to confirm the first calibration point. The upper display 

will be calibrated to the pH standard buffer solution and the lower display 

will then search for the next pH standard buffer solution.

5.  Repeat with other buffers if necessary. Rinse electrode before dipping into 

next buffer.

Note: The calibration mode allows you to perform up to three calibration points 

before returning to the measurement mode automatically. However, if you opted 

to have only one or two calibration points, simply skip the remaining calibration 

points by exiting to the measurement mode by pressing the CAL button. 

Operating Instructions

  

pHTestr

®

 30 Pocket Tester

Large Screen 

Waterproof pH/Temperature Tester

Double-Junction Sensor

To do a 1 point calibration  

only, press CAL button at this 

point to exit to the measurement 

mode. Otherwise, proceed to 

second buffer for a second 

buffer for a second point 

calibration.

To do a 2 point calibration only, press 

CAL button at this point to exit to the 

measurement mode. Otherwise,  

proceed to third buffer for a third  

buffer for a third point calibration.

After the third point calibration, the meter will 

automatically return to the measurement mode. 

At any point, an error message ‘Er. 1’ will be 

displayed momentarily if the confirmed pH value 

is not within the pH calibration window.

Figure 2: Example of pH Calibration Sequence

pH Measurement

1

. Press the ON/OFF button to switch the unit on.

2.  Dip the electrode about 2 to 3 cm into the test solution. Stir and let the  

reading stabilize. For semi-solid samples using pHSpear, rotation is ideal  

for sample contact.  

 

CAUTION: Testing dry samples is not accurate and can lead to sensor damage 

or breakage. Soils must be wet and free of particulates that may scratch the 

glass sensor. Excessive force into dry samples can cause glass breakage.

3.  Note the pH value or press HOLD/ENT button to freeze the reading. To 

release the reading, press HOLD/ENT again.

4.  Press ON/OFF to turn off unit. If you do not press a button for 8.5 minutes, 

the unit will automatically shut off to conserve batteries. 

Figure 4: Example of HOLD Function

HOLD Function

This feature lets you freeze the display for a delayed observation.

1.  Press HOLD/ENT button to freeze the measurement. A ‘HOLD’ indicator will 

be displayed and the measurement will be frozen.

2.  Press HOLD/ENT again to release the measurement. The ‘HOLD’ indicator 

will not be displayed anymore indicating the held measurement is released.

ATC - Automatic Temperature Compensation 

Through its built-in temperature sensor, the measurement error due to the 

changes in electrode sensitivity due to changes in temperature is compensated  

to give the actual pH reading of the sample measured. 

Note: To exit this program without confirming the calibration, press the  

CAL button before the automatic confirmation takes place.

• Aquaculture
• Aquariums
• Boilers
• Car washes
• Ecology studies

Applications

• Food sectors 
• Hydroponices
• Labs 
• Pools
• Sanitation plants

• Spas
• Steam generators
•  Water & wastewater 

treatment

•  

Water quality testing

• More!

User Reset

You can reset the pH calibration to the factory default by using the user reset  

function. Buffer set selection and temperature user calibration are not affected 

 

by the user reset function.

1. Switch off the unit.

2.  While pressing the CAL button, press and release the ON/OFF button to 

enter the ‘User Reset’ selection menu. The screen will display ‘rSt’ on the 

bottom display with a flashing ‘nO’ selection.

3. Use the CAL button to toggle between ‘nO’ and ‘YES’ selection.

• nO deactivates reset selection

 

• YES activates the reset selection

4. Press the HOLD/ENT button to confirm the selection made.

5.  If you have selected ‘YES’, the unit will show ‘CO’ momentarily and proceed 

to the measurement mode with the calibration reset back to factory default 

value.

6.  If ‘nO’ is selected, the unit will proceed to the measurement mode without 

any calibration reset performed.

1065O100_MAN_35634-30.indd   1

8/25/2017   7:15:07 AM

Reviews: