a-AB23EC Quick Start Guide
Switch On/Off
On
Off
Switch EC/TDS/SAL Mode
Main screen
EC
TDS
Switch Endpoint Mode
Auto end point mode
Continuous reading mode
Standard Solution Setup
Temperature Setup
Note:
Predefined standard solution: 10 μS/cm, 84 μS/cm, 1413 μS/cm,12.88 μS/cm
Select the solution group
Confirm
Confirm
Confirm
Calibration
Measurement
Store/Recall Data
Calibrating
Calibration result
(shown for 3 seconds)
Insert a probe into the
standard solution.
Insert the probe into a sample.
Measuring
Completed
Store measurement data:
After the measurement is
completed, press the
Store
button.
Recall measurement data:
press and hold the
Store
button to recall.
Recall calibration data:
After the calibration is
completed, press and hold
the
Cal
button to recall.
(
°C
/
°F)
Setup MTC temperature
Confirm
Setup temperature compensation
SAL
Setup reference temperature
Completed
TDS Factor Setup
Confirm