Summary of Contents for BH2-TET

Page 1: ...QUADRUPLE OBSERVATION TUBE BH2 TET REPAIRMANUAL OLYMPUS...

Page 2: ...For adjustment of BI and TRC part refer to the following manuals 1 BH2 B130 Repair Manual 2 BH2 TR30 Repair Manual BH2 B130 Repair Manual 0 Adjustment part explained in this manual Adjustment of optic...

Page 3: ..._ _ ________ A 2 4 SPECIFICATIONS _ _____________ ___ A 3 B DISASSEMBLY PROCEDURE 1 DISASSEMBLY PROCEDURE B 2 C ASSEMBLY AND ADJUSTMENT 1 ASSEMBLY OFEACH PART c 2 2 TENTATIVE ASSEMBLY c 3 3 ADJUSTMENT...

Page 4: ...BHZ TET A OUTLINE OF PRODUCT CONTENTS 1 OUTLINE A 2 2 FEATURES _ _ A 2 3 USING CONDITIONS ________ __________ A 2 4 SpE IFI ATIONS ____ ____ _ __ __ A 3 A l...

Page 5: ...th TV monitor image 3 Use of mold preventive agent suppresses generation of mold offering a good observation image 13 USING CONDITIONS 1 Unusable with BH2 accessories BH2 DO including MD01 I BH SPS in...

Page 6: ...djustment ring difference correction correction range 5 diopter 6 Photomicrographic PM mount system mount 7 TV camera system mount Straight tube C mount type TV camera is mounted through MTV3 Connecti...

Page 7: ...BH2 TET B DISASSEMBLY PROCEDURE CoNTENTS 1 DISASSEMBLY PROCEDURE _ B 2 c B l...

Page 8: ...3 PCS 0 Remove the PLATE 0 stuck with a double coated tape of the CONNECTINGTUBE with care taken not to bend it Remove the MIRROR MOUNT Screw CUK3x8SA 3 PCS 69 Remove the MIRROR stuck with the silico...

Page 9: ...the CONNECTINGTUBE Screw ANU3x4SA 2 PCS Remove the RING from the PRISM MOUNTa and take off the lens 0 Remove the LENS FRAME from the CONNECTINGTUBE Adjustable spanner TETKC02 Remove the RINGafrom the...

Page 10: ...BH2 TET C ASSEMBLY AND ADJUSTMENT CONTENTS 1 ASSEMBLY OF EACH PART __ __ __ ___ _ _ _ c 2 2 TENTATIVE ASSEMBLY _ c 3 3 ADJUSTMENT OF STRAIGHT TUBE C 6 4 ADJUSTMENT OF CONNECTING TUBE C 8 C l...

Page 11: ...he SCREW 01 and apply the adhesive to the mounting surface of PRISM MOUNTa and the PRISM Adhesive Araldite rapid OT1315 l 2 Bond the MIRROR totheMIRROR MOUNTa 1 Place the MIRROR on the MIRROR MOUNTawi...

Page 12: ...EASSEMBLY 2 l Screw the LENS FF AME in the CONNECTINGTUBE Adjustable spanner TETKC02 2 2 Insert the PRISM MOUNT all the way into the CONNECTINGTUBE Do not fix it Secure the MIRROR MOUNTOtentative ly t...

Page 13: ...5mm 2 5 Secure the SPACER firmly Place the SPACEROagainst the mount part of CONNECTINGTUBE even Screw AB3x14SA 3 PCS 2 6 Fit the FILTER into the SPACER 0 and secure it with the SPRING 2 7 Place the SP...

Page 14: ...BH2 TET C ASSEMBLY AND ADJUSTMENT 2 8 Assemble the CONNECTINGTUBE tentatively Screw AB3x35SA 4 PCS _ c 5...

Page 15: ...placement of optical tube length from the eyepiece scale scale corresponds to O lmm in if the displacement reduces in the direction of pulling out the helicoid ring of the eye 6 Increase or decrease t...

Page 16: ...STRAIGHT TUBE 4 Adjust the center cross hairs of the eyepiece to the center of the cross micrometer by loosening the screw and displacing the MOUNTO 5 Tighten the screw securely 6 Check the optical ax...

Page 17: ...pter TETKCOl 85 DAchlOx universal standard eyepiece KNOO48 2 Look through the centering tele scope and twist the CONNECTING TUBEato minimize displacement of the exit pupil 3 Tighten the screw to fix t...

Page 18: ...t the displace ment of the optical axis between BINOCULAR TUBE and CONNECTINGTUBE is less than 5 graduations 0 5mm 8 If the displacement is over 5 graduations repeat adjustment from 4 1 4 2 Adjustment...

Page 19: ...airs of the eyepiece fit into the SLEEVE R to the center of the cross micrometer move the stage After the adjustment take care I 3 Put the eyepiece to the CONNECTING TUBE 4 Adjust the center cross hai...

Page 20: ...BH2 TET C ASSEMBLY AND ADJUSTMENT 4 4 4 5 Stick dn the PLATEa 4 6 Mount the C MOUNT Apply the adhesive to the fixing screw of SPACER Adhesive Shellac OT1131 i c 11...

Page 21: ...BH2 TET D JIGS TOOLSAND ADHESIVES CONTENTS 1 ADHESIVES _ D 2 2 JIGSAND TOOLS D 3 D 1...

Page 22: ...BH2 TET D JIGS TOOLSAND ADHESIVES 1 ADHESIVES _ No Name OT1315 Adhesive Araldite rapid OT1538 Adhesive SE9156 OT1131 Adhesive Shellac D 2...

Page 23: ...AND TOOLS 6 No Name TETKC02 Adjustable spanner TETKCOl Eyepiece adapter KNOO48 Universal standard eyepiece KNoo29 Centering telescope Others Microscope stand adjusted FT36 Or IWO0251 l 100 cross micro...