Page 1: ...Polarizing Microscope INSTRUCTION MANUAL MODEL...

Page 2: ...ing the color and brightness of specimens The polarizing microscope goes a step further it can bring to light the chemical structure of even the smallest specimen The polarizing microscope therefore i...

Page 3: ...mm scale 10X cross hair with registration pin Objectives Achromat PS 4x N A 0 10 PS 10X N A 0 25 w centering device PS 40x N A 0 65 w centering device Condenser N A 1 20 iris diaphragm Sliding mechan...

Page 4: ...ens swing out nob nalyzer sliding plate entering screw bjective lens holder objective lens rotating scale vernier scale condenser clamp screw iris diaphragm lever filter holder mirror base BODY PARTS...

Page 5: ...lever the rotating scale is divided into 5 calibration When the polarizer is clicked into position at 0 its oscillation is longitudinal The iris diaphragm can be adjusted by means of a lever NOTE When...

Page 6: ...stage clip holes for fitting stage round clip mechanical stage holes for fitting stage clips vernier scale stage rotating scale stage rotation clamp Objectives Strain free objectives arc screwed firml...

Page 7: ...Moreover it rotates through 90 1 div 5 by means of a rotating lever At 0 the oscillation is latitudinal In other words at 0 the polarizer and analyzer are crossed Nicols Bertrand Lens The Bertrand le...

Page 8: ...are positioned at 0 and when it is inserted into the other groove B the cross line agrees with the diagonal position B A insertion grooves for registration pin eyepiece tube Eyepieces The flat lOx eye...

Page 9: ...t since dust together with moisture can cause molding Dust on the optical system especially the polarizer and analyzer causes double refraction which in turn dims contrast The POS therefore should be...

Page 10: ...clicked into position at 0 the oscillation direction of each coincides with the cross line of the eyepiece lengthwise for the polarizer and vvidthwise for the analyzer This condition is called crossed...

Page 11: ...r Conoscope First produce a conoscopic image by sliding the lower condenser lens out of the optical path and by inserting a large aperture objective 40X and Bertrand lens according to instructions giv...

Page 12: ...determined by checking the dial rotation angle against the correction table First of all the specimen is blotted out of vision by bringing the stage to a crossed Nicols position The stage is rotated...

Page 13: ...re meter Has smooth stable shutter performance Magnification on the film plane is about l 3x of micro scope total magnifications Opaque Illuminator Model LSP The Olympus Model LSP is designed to use w...

Page 14: ...enser and reglector these parts should be regularly cleaned with a soft brush Eyepieces should always be capped when not in use Special grease has been applied to the upper and lower moving parts of t...