The HOBO External Temp/RH data logger records temperature and relative humidity in an indoor 
environment with its external sensor. Using HOBOware®, you can easily configure the logger 
alarm to trip for specific high or low sensor readings. Or, you can set up burst logging in which the 
logger records data at a different interval during certain conditions. The logger can also calculate 
minimum, maximum, average, and standard deviation statistics. This compact data logger has a 
built-in LCD screen to monitor the current temperature, relative humidity, logging status, battery 
use, and memory consumption in between readouts. 

Specifications 

Temperature Sensor 

 

Range 

-20° to 70°C (-4° to 158°F) 

Accuracy 

±0.21°C from 0° to 50°C (±0.38°F from 32° to 122°F), see Plot A 

Resolution 

0.024°C at 25°C (0.04°F at 77°F), see Plot A 

Response Time 

6 minutes in air moving 1 m/s (2.2 mph) 

Drift 

<0.1°C (0.18°F) per year 

RH Sensor 

 

Range 

1% to 100% RH at -40° to 75°C (-40° to 167°F), non-condensing; 
exposure to conditions below -20°C (-4°F) or above 95% RH may 
temporarily increase the maximum RH sensor error by an additional 
1% 

Accuracy 

±2.5% from 10% to 90% RH typical to a maximum of ±3.5% including 
hysteresis at 25°C (77°F); below 10% and above 90% ±5% typical 

Resolution 

0.05% 

Response Time 

5 minutes to 90% in airflow of 1 m/s (2.2 mph) 

Drift 

<1% per year typical 

Cable Length 

1.83 m (6 ft) 

Logger 

 

Logger Operating Range 

Logging: -20° to 70°C (-4° to 158°F); 0 to 95% RH (non-condensing) 
Launch/Readout: 0° to 50°C (32° to 122°F) per USB specification 

Logging Rate 

1 second to 18 hours, 12 minutes, 15 seconds 

Logging Modes 

Fixed interval (normal), burst, or statistics 

Memory Modes 

Wrap when full or stop when full 

Start Modes 

Immediate, push button, date & time, or next interval 

Stop Modes 

When memory full, push button, or date & time 

Restart Mode 

Push button 

Time Accuracy 

±1 minute per month at 25°C (77°F), see Plot B 

Power Source 

One 3V CR2032 lithium battery and USB cable 

Battery Life 

1 year, typical with logging rate of 1 minute and sampling interval of 
15 seconds or greater 

Memory 

128 KB (84,650 measurements, maximum) 

Download Type 

USB 2.0 interface 

Full Memory Download Time 

20 seconds 

LCD 

LCD is visible from 0° to 50°C (32° to 122°F); the LCD may react slowly 
or go blank in temperatures outside this range 

Size 

3.66 x 8.48 x 1.52 cm (1.44 x 3.34 x 0.6 in.) 

Weight 

30 g (1.06 oz) 

Environmental Rating 

IP50 

 

The CE Marking identifies this product as complying with all relevant 
directives in the European Union (EU). 

HOBO External  
Temp/RH Data Logger 

UX100-023 

Included Items: 

 

Command™ strip 

 

Double-sided tape 

 

Hook & loop strap 

Required Items: 

 

HOBOware 3.4 or later 
(download at 
www.onsetcomp.com/ 
hoboware-free-download) 

 

USB cable 

 

 

Summary of Contents for HOBO UX100-023

Page 1: ...e 90 5 typical Resolution 0 05 Response Time 5 minutes to 90 in airflow of 1 m s 2 2 mph Drift 1 per year typical Cable Length 1 83 m 6 ft Logger Logger Operating Range Logging 20 to 70 C 4 to 158 F 0...

Page 2: ...d on the LCD screen followed by definitions of each symbol in the table below LCD Symbol Description The logger is waiting to be launched Press and hold the Start Stop button for 3 seconds to launch t...

Page 3: ...t Stop button for 1 second The LCD will then remain on for 10 minutes The LCD screen refreshes every 15 seconds regardless of the logging interval selected in HOBOware If you choose a logging interval...

Page 4: ...ogging If you select the Turn LCD off checkbox the LCD will not show the current readings status or other information while the logger is logging You will however be able to temporarily turn the LCD s...

Page 5: ...le then the logger would switch to burst logging mode again and record data every 10 seconds Once the temperature rises back to 32 F the logger will then return to normal mode logging every 5 minutes...

Page 6: ...ght maximum minimum average and standard deviation series four for temperature and four for RH with values calculated and logged every 5 minutes based on the 30 second sampling To set up statistics 1...

Page 7: ...der an RH Sensor Replacement Kit HUM RHPCB 2 and follow these instructions 1 Remove the tape fastening the sensor cap to the receptacle Discard the tape 2 Grasp the cap and pull firmly to remove it Di...

Page 8: ...ly installing the battery WARNING Do not cut open incinerate heat above 85 C 185 F or recharge the lithium battery The battery may explode if the logger is exposed to extreme heat or conditions that c...

Reviews: