Page 1: ...hickness Meter User Manual User manuals in various languages français italiano español português nederlands türk polski русский 中文 can be found by using our product search on www pce instruments com Last change 12 December 2019 v1 0 English ...

Page 2: ...Measurement 3 3 3 Measuring Range 4 3 4 Battery Indicator 4 3 5 Unit of Measurement 4 3 6 Display Flip 4 3 7 View Memory 5 3 8 SI and CO measurement mode 5 4 Standard Plastic Films Calibration 6 4 1 Accuracy Checking 6 4 2 Calibration 6 4 3 Initialisation 7 5 Shutdown 7 6 Warranty 8 7 Disposal 8 ...

Page 3: ...ns The case should only be opened by qualified PCE Instruments personnel Never use the instrument when your hands are wet You must not make any technical changes to the device The appliance should only be cleaned with a damp cloth Use only pH neutral cleaner no abrasives or solvents The device must only be used with accessories from PCE Instruments or equivalent Before each use inspect the case fo...

Page 4: ...8 g 1 lb without batteries 3 Operation 3 1 Start up The instrument is automatically turned on when the probe is pressed down to take a measurement and it is automatically turned off when no operation is taken within 90 seconds refer to Shutdown Important note The meter needs to warm up for a few minutes in order for the probe to reach environmental temperature Please disregard the first several re...

Page 5: ... in the memory excl current Mean 713μm average between the current previously saved values Min 625μm the minimum between the current previously saved values Max 800μm the maximum between the current previously saved values Sdev 22μm standard deviation of the current measurement 3 2μm NFe non magnetic base material Fe magnetic base material Important note Be sure not to slide along the coating or p...

Page 6: ...s too small A hardware error occurs probe is broken or dust dirt is on the probe 3 4 Battery Indicator The meter uses two AAA batteries as a power supply The battery icon in green colour represents a full battery Over time the green bar becomes shorter indicating the current capacity of the batteries The icon blinking in red indicates that the batteries are low Please recharge or replace the batte...

Page 7: ... or initialization is carried out the saved values will be cleared and the measurement counter returns to zero 3 8 SI and CO measurement mode By default the meter is in single measurement mode after start up and SI appears on the screen Press the and buttons at the same time to enter continuous measurement mode CO will appear on the screen In this mode the probe keeps moving on the measured surfac...

Page 8: ... is suggested 4 2 1 Zero In Press and hold the button for 2 seconds 000 will blink on the screen Take a normal measurement on an uncoated metal surface The reading is now calibrated to zero automatically After the Zero In procedure the meter can be used to accurately measure thickness However due to abnormal base materials or severe environmental circumstances errors may occur In such cases you ca...

Page 9: ... that the initialization is completed To ensure high measurement accuracy it is suggested to zero in after initialization All previously saved values and settings will be cleared after initialization the instrument is reset to factory default setting 5 Shutdown To save battery power the meter will shut down automatically after 90 seconds of inactivity Please refer to below figure The icon means th...

Page 10: ...d pollutants batteries must not be disposed of as household waste They must be given to collection points designed for that purpose In order to comply with the EU directive 2012 19 EU we take our devices back We either re use them or give them to a recycling company which disposes of the devices in line with law For countries outside the EU batteries and devices should be disposed of in accordance...

Page 11: ...3 975 114 Fax 49 0 2903 976 99 9971 Fax 44 0 2380 98703 9 Fax 39 0583 974 824 info pce instruments com info pce instruments co uk info pce italia it www pce instruments com deutsch www pce instruments com english www pce instruments com italiano The Netherlands China Hong Kong PCE Brookhuis B V PCE Beijing Technology Co Limited PCE Instruments HK Ltd Institutenweg 15 1519 Room 6 Building Unit J 21...