background image

PHILIPS  SEM  525  INSTRUCTIONS

Start-up:

Turn  master  mains  switch  on  (if  not  already  on)
Turn  the  water  cooler/circulator  timer  on  to  HOLD  (counter  clockwise)
Press  the  SEM  mains  button  ON  (lower  left  panel,  listen  for  the  mechanical

vacuum  pump  to  come  on)

Press  the  vacuum  ON  button

The  red  high  vacuum  light  (left  most  panel)  will  be  lit.    After  about  30
minutes,  this  light  indicator  will  turn  off  indicating  a  sufficient  vacuum  to
operate  the  microscope.    However,  to  ensure  maximum  filament  life,  do  not
turn  the  beam  on  until  the  vacuum  gauge  reads  well  below  “10”.

Turn  o n  the  exposure  light  column  switch.
Turn  o n  the  “marker”  switch.
Turn  ON  the  detector  and  adjust  the  “Black  Level”.

Set  the  detector  to  GREY  SCALE  and  adjust  brightness  and  contrast  for  the
display  screens  (once  set,  this  does  not  change  appreciably  over  time).

If  using  the  Polaroid  camera,  press  and  hold  the  photo  start  button  and
press  the  scan  rate  2000  and  32  buttons.    This  sets  the  scan  rate  for  the
camera  screen.    The  photo  scan  can  be  stopped  by  pressing  the  interrupt
b u t t o n .

Adjust  operating  parameters  appropriate  to  the  specimen  characteristics.

Select  accelerating 

voltage:

For  maximum  resolution  and  high  magnification  work,  use  25  kV  or  above
with  gun  bias  setting  of  “1”  and  a  spot  size  of  20  or  10.    This  electron  beam
will  penetrate  well  into  most  samples  and  require  a  lower  secondary
electron  detector  gain  setting.    The  detector  bias  may  be  turned  to  negative
to  detect  backscattered  (BS)  electrons  if  desired,  but  there  will  be  a  lot  of  BS
electrons  with  this  kV  setting.    A  better  BS  image  will  be  obtained  using  a
lower  accelerating  voltage.

Summary of Contents for SEM 525

Page 1: ...E and adjust brightness and contrast for the display screens once set this does not change appreciably over time If using the Polaroid camera press and hold the photo start button and press the scan rate 2000 and 32 buttons This sets the scan rate for the camera screen The photo scan can be stopped by pressing the interrupt button Adjust operating parameters appropriate to the specimen characteris...

Page 2: ...isions the following should be expected at filament saturation Bias 1 25 kv saturated filament current 40 µA Bias 2 10 kv saturated filament current 35 µA To lower the saturated filament current increase the filament to Wehnelt distance Select spot size Generally speaking use a large spot size 200 or 500 for low magnification work under 1000 x magnification Smaller spot size is used for higher mag...

Page 3: ... 500 and minimum magnification turn up the filament current until the current no longer rises on the gauge Adjust the detector gain to provide a circular illumination Center this circle on the screen using the gun shift then use the gun tilt controls to give the brightest illumination Changing or inserting a new specimen Be sure the filament current and high tension is off and the filament has coo...

Page 4: ...magnification control to maximum lowest lens current setting Turn off the light column switch Turn off the marker switch The SEM can be left in this setting as standby mode for the day Do not leave the SEM on overnight To completely shut down the SEM overnight or longer Press the vacuum OFF button Press the power OFF button Set the cooling water timer to at least 30 minutes to cool the diffusion p...

Reviews: