background image

Technical Note 

P726T0001, valid for P-726 

ECo/CBo, 4/3/2017 
 

 

Physik Instrumente (PI) GmbH & Co. KG, Auf der Roemerstrasse 1, 76228 Karlsruhe, Germany 
Phone +49 721 4846-0, Fax +49 721 4846-1019, Email 

info@pi.ws

, 

www.pi.ws 

PIFOC® QuickLock Thread Option 

Installation instructions 

Introduction 

This technical note describes the assembly/disassembly and installation/deinstallation of P-726 

PIFOC® microscope objective nanofocussing positioners and scanners with the QuickLock 

feature. 
The QuickLock feature was developed to allow use of the same PIFOC® unit with any of the 

supported microscope and objective thread types. Previous PIFOC®s could each be used only 

with a single microscope thread size. 
All QuickLock PIFOC®s can be installed without rotating the unit, eliminating inconvenient cable 

windup. 
Note that PIFOC

®

s and QuickLock adapters have to be ordered separately. Mounting tools are 

included with QuickLock adapters. 

Design 

The QuickLock thread option provides interchangeable insert rings with matching threads for 

the microscope and the objective. The objective ring accepts the microscope objective and is 

screwed into the PIFOC® body (the required flat wrench is provided). The turret ring is screwed 

into the microscope and is clamped to the PIFOC® body under the knurled ring. 
 

 

NOTICE 

 

Possible damage due to excessive torque! 

Only moderate torque is necessary to secure the objective ring in the PIFOC® body. 

 

Only use the included flat wrench for tightening the objective ring. 

 

PIFOC® QuickLock Thread Option 

The QuickLock adapters are available with apertures of different inner diameters: 

QuickLock adapter / item no.  Size of adapter aperture 

Diameter of aperture 

P-726.04 

M28 x 0.75 mm 

23 mm 

P-726.05 

M32 x 0.75 mm 

27 mm 

P-726.06 

M26 x 1/36’’ 

21 mm 

P-726.11 

M25 x 0.75 mm 

21 mm 

P-726.12 

W0.8 x 1/36’’ 

16 mm 

 

Reviews: