The Powerex Med Touch combination area and master alarm
panel digitally displays gas pressure (1 psi increments), monitors and
displays normal and alarm conditions from up to 128 remote medical
gas source signals, and provides alarm conditions as required by the
latest edition of NFPA 99 for up to 16 medical gases. The alarm is ETL
listed to UL 1069 and CSA C22.2 # 205 Signal equipment. Transducers
& DISS risers are included for area alarm gas boards. Transducers may
be installed remotely or in the back box. Pressure switches and DISS
union check valve connectors are sold separately.
Features & Benefits
•
Five year parts and one year labor warranty
•
2.85” LCD touch screen displays up to 8 normal and alarm
conditions for pressure switch sensor inputs
•
2.85” LCD touch screen displays up to 8 normal and alarm
conditions and provides a digital pressure display for up to 4
transducer sensor inputs
•
2.85” LCD touch screen displays normal and high/low alarm
conditions for one 4 – 20 mA sensor input
•
The LCD touch screen allows all alarm programming and set up to
be done without the use of tools
•
A green normal or red alarm condition for each gas services
confirms the condition for each individual gas service
•
Emergency preparedness instructions - Med Touch alarm panels
allow users to set up customized instructions for each alarm
signal, to appear on the screen when the signal is in alarm
•
Up to 128 signal points in a single panel
•
Wireless (minimal low voltage wiring) models available (wireless
feature for master signal points only)
•
Last event history (per gas board /signal point)
•
Made in the U.S.A.
•
NFPA and ISO pre-loaded gas ‘labels’
•
English, and English/French pre-loaded languages
•
Editable text and alarm ‘labels’
•
Self-contained unit - Designed for ease of installation and service
•
Self-diagnostic and error message display for ease of maintenance
•
Audio and visual alarm indicators
•
Bright easy to read LCD displays – clearly visible in both day and
night lighting conditions
•
Constant display and monitoring of each gas
•
User programmable high/low set points on transducer and 4 – 20
mA sensor input boards
•
Dry contacts provided on the CPU module for general fault
condition covering entire panel
•
Hinged frame for easy accessibility
•
Color coded displays
•
Alarm history display of previous alarm conditions
•
Hinged frame for easy accessibility
•
Color coded labeling
•
Wireless (minimal low voltage wiring) models available
Master Alarm
Area Alarm
Combination Alarm