background image

Home 

Use

Bubble Humidifier Instructions for Use 
and Cleaning Instructions

Instructions for Use

1. Wash hands and remove bubble humidifier from package.

2. Unscrew lid and add water to the Maximum Level marking 

on the side of the bottle. Sterile, distilled or boiled water is 

recommended.

3. Replace lid and ensure lid is screwed on tight.

4.  Attach bubble humidifier to oxygen source, i.e. flowmeter, 

concentrator, liquid tank.

5. Test for leaks (A). Turn the flow to 6 LPM and block the  

outlet port on the bubble humidifier. An alarm should sound in  

5 seconds or less. If not, check connections.

6. Attach oxygen tubing to the outlet port (B) and adjust flow to 

prescribed setting.

7. Apply interface, i.e. nasal cannula or oxygen mask, to patient.

8. If water level drops to Minimum level marking on bottle, add 

more water.

The Bubble Humidifier is a disposable, single-patient 
use device for use in health care facilities and home care. 
The humidifier is used to provide additional humidity to 
supplemental oxygen therapy and is available with either a  
3 PSI or 6 PSI safety pop-off valve. 

7100 Bubble Humidifier- 3 PSI Pop-off   |   7600 Bubble Humidifier- 6 PSI Pop-off

MAXIMUM

Pressure 

relief alarm

Pressure 

relief port

Maximum

water level

Minimum

water level

Diffuser

Oxygen 

outlet port

MINIMUM

Oxygen source

O

2

 Source

MAXIMUM

MINIMUM

(A)

MAXIMUM

MINIMUM

Oxygen

Tubing

To

Patient

O

2

 Source

(B)

Reviews: