BiNAPS

Nasal  Airflow  Pressure  and  Snore  Transducer

User Manual

Salter Labs

30 Spur Road

El Paso, TX 79906

 

Toll Free: 1-800-421-0024

www.salterlabs.com

®

®

®

Nasal Airflow Pressure and Snore Transducer
CLEANING    &    MAINTENANCE

The BiNAPS device is reusable and must be cleaned between patients.

1.

Wipe the outside surface of the BiNAPS with a damp cloth and mild

 

detergent, or an approved disinfectant. Refer to the table below. Follow the

 

manufacturer’s instructions for use and recommended exposure time.

2.

DO NOT immerse or soak the BiNAPS device in liquid.

3.

DO NOT sterilize.

4.

Ensure devices are completely dry before reusing.

5.

Salter Labs sleep diagnostic cannulas are disposable and for single-patient

 

use. Do not clean or sterilize or re-use. 

WARRANTY

 Features and specifications are subject to change without notice.

121321 Rev.

 F

MT Promedt Consulting GmbH

Altenhofstrasse 80

D-66386 St. Ingbert

Germany

®

Salter Labs

 warrants this product to be free from defects in materials and

 

  

workmanship as long as 

 used with  

Salter Labs 

cannulas.

 

Salter Labs  cannulas 

must be used to protect the BiNAPS Lifetime Warranty. 

 

Should the BiNAPS 

module fail prematurely, the sole liability of 

Salter Labs

  is limited to repair or 

replacement of the product with no charge for parts or labor. Under no 

circumstances shall 

Salter Labs

 be liable for any loss or damage, direct, incidental 

or consequential, including property damage or personal injury arising form the use 

of or the inability to use the product. This warranty is rendered void and Salter Labs 

cannot be held liable for conditions resulting from damage, marginal performance,  

or malfunctions caused by: misuse, abuse, neglect, improper line voltage, power 

fluctuations or any adverse environmental conditions, tampering, unauthorized 

modifications, adjustment or repair to the product or its accessories. This warranty 

is in lieu of all other warranties, expressed or implied and is extended only to the 

original purchaser.  IMPORTANT NOTICE: The use on non-Salter Labs cannulas 

will void the warranty.

EC  REP

  

Sensor  Technology ............................................. Piezo  Pressure  Transducer

  Output  Types .......................................................................... Airflow  &  Snore
  Size ..................................................................................... 2.26in X 1.37in X .68in
  Weight ..................................................................................................... 30 grams
  Power Source ................................................................................. None required
  Output  Connectors ....................................................................... 1.5 mm Safety
  Airflow  Sensor  Cannula  Input  Connector ......................... Female luer lock
  Input  Pressure  Range ...................................................................... ± 25cm H

2

0

  Signal  Output  Levels  (Typical) ....................................  1000µV peak to peak
  Operating  Temperature ........................................ 5º C (40º F) to 40º C (104º F)
  Storage Temperature ............................................. 5º C (40º F) to 40º C (104º F)
  Operating/Storage  Humidity ................................. 15 - 95% Non-condensing
  Test Lead Recepticle ............................................................ 1.5 mm Plug (Male)

PRODUCT SPECIFICATIONS