BiNAPS
Nasal Airflow Pressure and Snore Transducer
User Manual
Salter Labs
30 Spur Road
El Paso, TX 79906
Toll Free: 1-800-421-0024
www.salterlabs.com
®
®
®
Nasal Airflow Pressure and Snore Transducer
CLEANING & MAINTENANCE
The BiNAPS device is reusable and must be cleaned between patients.
1.
Wipe the outside surface of the BiNAPS with a damp cloth and mild
detergent, or an approved disinfectant. Refer to the table below. Follow the
manufacturer’s instructions for use and recommended exposure time.
2.
DO NOT immerse or soak the BiNAPS device in liquid.
3.
DO NOT sterilize.
4.
Ensure devices are completely dry before reusing.
5.
Salter Labs sleep diagnostic cannulas are disposable and for single-patient
use. Do not clean or sterilize or re-use.
WARRANTY
Features and specifications are subject to change without notice.
121321 Rev.
F
MT Promedt Consulting GmbH
Altenhofstrasse 80
D-66386 St. Ingbert
Germany
®
Salter Labs
warrants this product to be free from defects in materials and
workmanship as long as
used with
Salter Labs
cannulas.
Salter Labs cannulas
must be used to protect the BiNAPS Lifetime Warranty.
Should the BiNAPS
module fail prematurely, the sole liability of
Salter Labs
is limited to repair or
replacement of the product with no charge for parts or labor. Under no
circumstances shall
Salter Labs
be liable for any loss or damage, direct, incidental
or consequential, including property damage or personal injury arising form the use
of or the inability to use the product. This warranty is rendered void and Salter Labs
cannot be held liable for conditions resulting from damage, marginal performance,
or malfunctions caused by: misuse, abuse, neglect, improper line voltage, power
fluctuations or any adverse environmental conditions, tampering, unauthorized
modifications, adjustment or repair to the product or its accessories. This warranty
is in lieu of all other warranties, expressed or implied and is extended only to the
original purchaser. IMPORTANT NOTICE: The use on non-Salter Labs cannulas
will void the warranty.
EC REP
Sensor Technology ............................................. Piezo Pressure Transducer
Output Types .......................................................................... Airflow & Snore
Size ..................................................................................... 2.26in X 1.37in X .68in
Weight ..................................................................................................... 30 grams
Power Source ................................................................................. None required
Output Connectors ....................................................................... 1.5 mm Safety
Airflow Sensor Cannula Input Connector ......................... Female luer lock
Input Pressure Range ...................................................................... ± 25cm H
2
0
Signal Output Levels (Typical) .................................... 1000µV peak to peak
Operating Temperature ........................................ 5º C (40º F) to 40º C (104º F)
Storage Temperature ............................................. 5º C (40º F) to 40º C (104º F)
Operating/Storage Humidity ................................. 15 - 95% Non-condensing
Test Lead Recepticle ............................................................ 1.5 mm Plug (Male)