background image

97.9°F
36.6°C

98.4°F
36.9°C

97.9°F
36.6°C

98.4°F
36.9°C

HEAT

POCKET

HEAT

POCKET

98.8°F
37.1°C

98.8°F
37.1°C

97.7°F
36.4°C

98.1°F
36.7°C

BATTERY COVER

ON/OFF BUTTON

DIGITAL DISPLAY WINDOW

PROBE TIP

Thank you for purchasing Samsung Healthy 

Living

®

 30 Second Digital Thermometer with 

beeper, memory and high temperature alarm.

Before using your new thermometer please read 

and follow instructions to insure most accurate 

body temperature readings. Information herein 

and product are not meant to be a substitute for 

the advice provided by a medical professional.  

Normal Body Temperature fluctuates during the 

course of a day and depends on location of 

temperature reading. The generally accepted 

"normal" body temperature is 98.6°F(37.0°C) for 

oral. An axillary or underarm reading is usually is 

1.0°F (0.5°C) lower 97.6°F (36.4°C) and a rectal 

temperature is usually 1.0°F higher (0.5°C) 99.6°F 

(37.5°C). Body temperature readings tend to be 

lowest in the morning hours from 2 a.m. to 4 a.m. 

while your body is resting and is highest during 

late afternoon hours from 4 p.m. because your 

body has been active throughout the day. 

Prior to each use clean the thermometer with a 

soft cloth moistened with rubbing alcohol (70% 

isopropyl alcohol) diluted with warm water, OR 

wash with cold, soapy water. An intact unit can be 

s u b m e r s e d  i n  5 0 %  w a t e r  a n d  5 0 %  a l c o h o l ;  

however do not soak for an extended period of 

time. Be sure to dry thermometer thoroughly using 

soft cloth. 
To make temperature taking more hygienic and to 

h e l p  p r e v e n t  c r o s s - c o n t a m i n a t i o n  i t  i s  

recommended to use Samsung Disposable Probe 

Covers (model # 

TSP-001

) each time temperature 

is taken.

When the "188.8" symbol is continuously flashing, 

the battery  needs to be replaced with 

an SG3 silver oxide type. 

1. Slide battery cap away from unit.
2. Remove the battery using a small Phillips-

head or jeweler's screwdriver and insert new 

battery with the positive (+) side facing up. 

3. Carefully slide battery cap back into its place, 

making sure the cap is securely fit to protect 

battery from moisture.     

1. Clean and properly disinfect the thermometer  

according to the instructions provided. 

2. Place a new or clean and undamaged probe 

cover carefully over the thermometer tip.

3 . S e l e c t  o r a l ,  a x i l l a r y  o r  r e c t a l  a s  y o u r  

measurement site.

4. Press On/Off button to activate thermometer.   

The digital display will show the following and 

the thermometer will beep. 

NOTE :

 Relax and wait approximately 5 minutes with your 

mouth closed before taking a reading if you've just 

consumed any hot or cold beverages, exercised or 

smoked. This will reduce chances of possible high or low 

temperature readings.

1. Place a new or clean and undamaged probe cover 

carefully over the thermometer tip.

2. Place the probe tip under the tongue as near as 

possible to the heat pocket.  The diagram below also 

shows the temperature variables by not placing the tip 

in the heat pocket areas. Hold the thermometer still in 

the same spot and mouth should remain closed for 

most accurate reading.

3. The current temperature is shown and the °F (°C) will 

continue to flash during the reading.

4. When peak temperature is reached, the °F (°C)  will 

stop a flashing and four quick beeps will sound.

5. To turn the unit off, press the On/Off button, otherwise 

the unit will automatically shut-off after approximately 

10 minutes to preserve battery life. 

This is the LCD display  test, which insures 

all  parts of the display are functioning 

properly. Continue to hold the button 

switch  down over 2 seconds, the display 

will flash alternatively °C/°F icons. Release 

button on  preferable mode of read-out. 

This display shows the last temperature 

reading after °C/°F  selection.

After the display of the last measured 

r e a d i n g ,  t h e  u n i t  w i l l   p e r f o r m  a  s e l f -

calibration  (standard temperature) and 

display 97.7°F for  about one second.

The "Lo" with the flashing °C (°F) indicates 

t h e  t h e r m o m e t e r  i s  r e a d y  t o  t a k e  a  

t e m p e r a t u r e   r e a g i n g  ( I f  " E R R "  i s  

d i s p l a y e d ,  a  m a l f u n c t i o n  h a s  b e e n  

detected and thermometer will need to be 

replaced.)

T h e  b e s t  w a y  t o  f i n d  o u t  y o u r  n o r m a l  b o d y  

temperature is to measure yourself when you are 

feeling well. Record your readings twice a day 

(early morning and late afternoon). The average 

o f  t h e  t w o  r e a d i n g s  i s  y o u r  n o r m a l  b o d y  

temperature. Any deviation from this may indicate 

an illness and you should consult a medical 

professional. 

This thermometer has a built-in alarm to warn 

user of higher than normal temperature. If reading 

exceeds 99.5°F (37.5°C) eight sets of three quick 

beeps will sound.

INSTRUCTIONS FOR USE

30 Second Digital Thermometer

with Accu-Fast

TM  

Technology

CLEANING INSTRUCTIONS &

USING DISPOSABLE PROBE COVERS 

BATTERY

BODY TEMPERATURE

How to know your Normal Body 

Temperature

HIGH TEMPERATURE ALARM 

To replace battery

,

Caution:

 Please properly dispose of the old battery away 

from small children and heat.

TAKING AN ORAL TEMPERATURE

Approximate reading time: 30 seconds 

NOTE :

 Rectal temperatures are the most accurate 

method for infants and small children. Otherwise this 

method should only be used when it is difficult to take an 

oral or axillary measurement. Rectal temperatures are 

generally 1.0°F (0.5°C) higher than oral readings. For 

hygienic reasons, once thermometer is used rectally, do 

not use orally. 

1. Place a new or clean and undamaged probe 

cover carefully over the thermometer tip.

2. Have your child lie stomach down on your lap.

3. Apply a water-soluble lubricant to the tip of the 

thermometer and around rectum.

4. Hold your child still and gently insert the probe 

maximum of 1 inch

 into the rectum. If child 

i s  l e s s  t h a n  6  m o n t h s  o l d ,  g e n t l y  i n s e r t  a  

maximum of 1/4 to 1/2 inch

 into the rectum.

5. When peak temperature is reached, the °F (°C) 

will stop flashing and four quick beeps will 
sound.

6. To turn the unit off, press the On/Off button, 

otherwise the unit will automatically shut-off 
approximately after 10 minutes to preserve 
battery life. 

TAKING A RECTAL TEMPERATURE

Approximate reading time: 30 seconds 

NOTE :

 Axillary temperatures are generally 1.0°F 

(0.5°C) lower than oral readings. 

1. Wipe underarm area with a dry towel.

2. Place a new or clean and undamaged probe cover 

carefully over the thermometer tip.

3. Place the probe tip under the arm so that the tip is 

t o u c h i n g  t h e  s k i n  w i t h  t h e  t h e r m o m e t e r  

perpendicular (vertical) to the body. 

4. Position arm across the chest so the probe tip is 

well covered by the arm. This insures the ambient 

temperature does not affect the probe.

5. When peak temperature is reached, the °F (°C) will 

stop flashing and four quick beeps will sound. 

6 .  T o  t u r n  t h e  u n i t  o f f ,  p r e s s  t h e  O n / O f f  b u t t o n ,  

otherwise the unit will automatically shut-off 

approximately after 10 minutes to preserve battery 

life. 

TAKING AN AXILLARY 

TEMPERATURE (UNDERARM)

Approximate reading time: 50 seconds 

SPECIFICATIONS

Model Number
Display
Display Resolution
Measurement Range 
Measurement Location
Temperature Scales 
Accuracy

High Temperature Alarm

Completed Measurement
Probe Covers
Battery Type
Battery Life
Automatic Shut-off
Memory
Waterproof
Ambient Range

Storage Range 

Dimensions (approx.)
Weight (approx.)

TSR-301S
Liquid Crystal Display
4 Digits (1/10th of a degree increments)
89.6°F to 109.4°F (32.0°C to 43.0°C)
Oral, Axillary, and Rectal
°F or °C
± 0.2°F between 89.6°F and 109.4°F 

(±0.1°C between 32.0°C and 43.0°C)
> 99.5°F (37.5°C) eight sets of three 

quick beeps will sound
Four quick beeps will sound
TSP-001
SG3 Alkaline
Approximately 200 hours
Approximately after 10 minutes
Last temperature recall
Yes
41.0°F to 104.0°F (5.0°C to 40.0°C)

10% to 80% RH
14.0°F to 140.0°F (-10.0°C to 60.0°C) 

10% to 80% RH 
5.2" x 0.83" x 0.43"
0.47 oz including battery

Specifications are subject to change without notice.

Manufacturer's One Year

Limited Warranty

Manufacturer will, at its option, repair or replace your thermometer with 

the same or comparable model for a period of one year from the date of 

purchase in the event of a defect in materials or workmanship.  The 

warranty only extends to the original retail purchaser.  Purchase receipt 

or other proof of original purchase is required before warranty 

performance.

This warranty extends solely to failures due to defects in materials or 

workmanship occurring during normal use.  It does not cover normal 

wear of the product.  This warranty is void if: the housing product has 

been removed, if the manufacturer's label, logo or serial number have 

been removed or if the product fails to function properly as a result of 

accident, misuse, abuse, neglect, mishandling, misapplication, 

defective batteries, faulty installation, set-up, adjustments, improper 

maintenance, alteration, maladjustments of control, modification, power 

surges, commercial use of product, service by anyone other than an 

authorized service center or acts beyond the manufacturer's control.

This warranty and remedy provided herein are exclusive and in lieu of 

all other express and implied warranties and unless stated herein, any 

statements or representations made by any other person or entity are 

void.  Neither the manufacturer, distributor nor their affiliates shall be 

liable for loss, inconvenience, injury, or damage including direct, 

special, incidental, or consequential damages resulting from the use, 

improper use, failure to heed any warning, disclaimer, or instruction or 

inability to use this product, whether arising out of breach of this 

warranty or any other legal theory.  All express and implied warranties, 

including the warranties of merchantibility are limited to the applicable 

warranty period set forth above.  Manufacturer and distributor assume 

no responsibility for injuries, damages or penalties incurred or suffered 

resulting from the use of this product in an improper manner or location 

other than for which it is intended or from failure to heed any warning, 

disclaimer, or instruction.

SAMSUNG AMERICA, INC.

Printed in Taiwan

info@SamsungHealthcare.com

www.SamsungHealthcare.com

1.800.261.0130

Question or Comments:

105 Challenger Road 6

th

 Fl.

Ridgefield Park, NJ 07660

1

2

3

4

5

6

7

8

9

Reviews: