Installation

Checking the pressure readings:

1.  Place a handpiece coupling onto an air pressure gauge, if available.
2.  Place the gauge onto the dental unit tubing. 
3.  Shut off the water and place your foot onto the rheostat.
4.   Adjust the drive air, chip air & water on the unit, as necessary, to reach the recom-

mended pressure readings. 

Pressure readings for the STAT

IS

 SL & ML:

• 

Drive air = 2.1-4.2 bar (30-60 p.s.i.)

• 

Recommendation = 2.8 bar (40 p.s.i.)

• 

Return air < 0.5 bar (7 p.s.i.)

• 

Water = 0.8-2.5 bar (11-36 p.s.i.)

• 

Spray air = 1.0-4.0 bar (14-57 p.s.i.)

 

 The minimum amount of water required for spray cooling is  
50 mL per min. An insufficient amount of water may cause  
overheating of the tooth.

Checking the STAT

IS

 SL & ML coupling:

All 5 O-rings must be in place and should not be damaged. Missing or damaged O-rings 
should be replaced, or they may result in premature failure or negatively affect handpiece 
performance.

Important notes for burs & diamonds:

Diamond and carbide use: Only use burs that have shafts compliant with IS0 1797-1 
type 3, made from steel or tungsten carbide and that meet the following criteria:

• 

Shaft diameter from 1.59 -1.60 mm.

• 

Overall length up to max. 25 mm.

• 

Shaft clamping length of at least 11 mm.

• 

Cutting edge diameter up to max. 2 mm.

• 

Short shank burs can be used with the STAT

IS

 SL 101.1 model.

STAT

IS

®

 Handpieces

 

- Quick Reference Guide

SciCan GmbH

Wangener Strasse 78 
88299 Leutkirch  GERMANY 

Tel.: +49 (0)7561 98343 - 0 
Fax: +49 (0)7561 98343 - 699

Manufactured by:

STAT

IS

 ML

S10000

STAT

IS

 SL

S10001

SD-469 CA EN FR R1. 

STAT

IS

 is a registered trademark and 

Your Infection Control Specialist

 is a trademark of SciCan Ltd.

Please consult Operator’s Manual before using the devise(s).

6 pin, MULTIflex style 

coupling with LED bulb,

anti retraction valve.

Connex LED

S10010

6 pin, coupling with 

light connection, anti 

retraction valve.

Connex L

S10002

5 hole, coupling with 

fiber optic glass rod, 

anti retraction valve.

Connex GR

S11002

Did you select the right coupler for your tubing?

Reviews: