Operation Manual

 

 

   

 
 
 
 
 
 

 
 
 
 
 
 
 
 
 

Ver. 1.7

 

Summary of Contents for SCIP5-Mini

Page 1: ...Operation Manual Ver 1 7...

Page 2: ...eir possible applications please read the operating manual carefully prior your use And handle the device with care during transportation unpack testing operation maintains storage and scrape Catalog...

Page 3: ...qualifications of the person operating the unit To ensure that the person who operate the unit is trained in the relevant work application by a trained person The operative must receive regular traini...

Page 4: ...terials Fire hazard Recommended fluids only chemical dangers may occur depending on the bath medium Prevent water getting into hot bath oils The safety temperature limit must always be set to at least...

Page 5: ...ly Please FILL the warranty card if you found any damage of the device and accessories notify the distributor and inform us immediately Setting up Functional Module Control Panel Pump Inlet Pump Outle...

Page 6: ...re the pump plug is security fixed without any external circulation Filling and draining Connect the inlet and outlet to the external reactor if needed Attention Ensure the drain valve is fully closed...

Page 7: ...s for 1 seconds to turn on off the circulation pump press for 5 seconds to entry setting mode Menu press for start stop circulation and temperaturecontrol press for turn up the setting temperature pre...

Page 8: ...on B to jump to the next setting item Press button A to exit the setting mode and display the working temperature Press button C or D to enter the lowest temperature setting interface Press button C o...

Page 9: ...restart of functions depending on previous settings Set values set in A1or B cannot be changed Over temperature protection Over temperature O T Press button B to jump to the next setting item Press b...

Page 10: ...to exit the setting mode and display the operation temperature Press button C or D to enter the interface for viewing the value Press button B to jump to the temperature calibration point 1 deviation...

Page 11: ...ibration value is stable and reliable and then press the button B to save the deviation value Show device serial number To show the 1st 4 Digi of the serial number High Serial number HiSn Press button...

Page 12: ...the liquid level of the bath vessel Er05 Over temperature Check the over temperature setting value Check the Temperature sensor and the heater Please contact us if the corrective action described fai...

Page 13: ...tubes Thermal insulation sleeve for DN12 tubes PT 100 external temperature sensor Warranty The warranty period of device is 12 months not include damage to the product resulting from accident misuse...