NOVASUB - DDG-P QUICKSTART GUIDE - R1.1
WWW.NOVASUB.COM
R1.0 1 of 2
Manufactured by Seascape.nl
Action
Description
Needed
Installation
•
Connect both probe cables and probes to the SCC-DDG-P side
panel
•
Connect external 100-230 vac power
2x UDS-2 pressure
sensor with cables
Power Up
Press Power button to start the DDG
Calibration
preparations
•
Place both sensors vertical together
•
Connect the measuring tap 0 on the measuring point of the sensors
•
Make sure that both cables are long enough so that the pressure
sensors can be lowered to the max. calibration point
Measuring tape
Initial calibration
(Do the initial
calibration on each
new project
location)
1. Press Menu and go to Maintenance
Admin settings
Login
Admin
2. Type password: OK+--+OK and OK again
3. Go to Menu Calibrations in Admin settings Menu
4. Select P1 and P2 and press OK
5. With Both pressure sensor in the air above and out of the water
press, OK
6. Lower down both Pressure sensors to the max. calibration depth
and read the exact depth at the measuring tape water surface line.
7. Enter the value in millimetres , use + and
– buttons to select value
and press OK to go to next digit. Press OK after last digit.
8. Make sure pressure sensors are on the entered water depth, Press
OK to set the WCcalMax.
9. Reboot the DDG-P
Reference point
NAP-ref.
(Reference point of
the project to which
p1 if referenced to)
1. Press Menu and go to Maintenance
Admin settings
Login Admin
2. Type password: OK+--+OK and OK again
3. Go to Menu Calibrations in Admin settings Menu
4. Select Config NAP ref and press OK
5. Enter the value in millimetres , use + and
– buttons to select value
and press OK to go to next digit. Press OK after last digit.
Installing P1
P1 needs to be placed vertical in water approx. 0,5 m below water level
and deep enough that the water level never gets below the depth
sensor.
•
Measure the exact distance from bottom of P1 to NAP-ref. point
Measuring tape
Level P1 settings
1. Login in Admin settings Menu
2. Go to Menu Calibrations in Admin settings Menu
3. Select Config Level P1 and press OK
4. Enter the value of Level-P1 in millimetres , use + and
– buttons to
select value and press OK to go to next digit. Press OK after last
digit.
P-offset
The P-offset is fixed for the standard supplied UDS-2 sensor.
Do not
change
.
Logging
1. T
he “Dive monitor” shows “ Press OK to start log” on the display
2. Start the next log by pressing OK
3. To log a point Press OK
4.
Continue Pressing “OK” to log a new point
5.
Stop logging by pressing “MENU” and “OK”
6. Logfile: X shows the file number and Point: x shows the point
number in the logfile. Each point is stored in the file with all data.