background image

Quick reference guide USB-MESY

 HID 570.77A

 \ 10.2013 \

Installation and operation

Quick reference guide 

for USB-MESY HID 570.77A

 

Summary of Contents for 570.77A

Page 1: ...Quick reference guide USB MESY HID 570 77A 10 2013 Installation and operation Quick reference guide for USB MESY HID 570 77A ...

Page 2: ...Page 2 Quick reference guide USB MESY HID 570 77A 10 2013 ...

Page 3: ...soft Windows XP Windows Vista und Windows 7 Please check the scope of delivery before the beginning of the operation 1 USB MESY HID 570 77A 3 thermocouple sensors type K length 0 5 m 1 USB connecting cablel connector A Mini USB length 1 8 m 1 external power supply 5V 1A 1 installation CD ROM 1 quick reference guide USB MESY HID 570 77A Optional available accessories not included in the deliv ery t...

Page 4: ... USB device and will install the necessary driver automati cally Instruction Charge the internal accumulator of the MESY com pletely before the first use For this purpose connect the USB MESY HID with a free USB port of your PC The charge of a complete empty accumulator via the USB interface will take approx 10 hours You can use the external power supply to charge the MESY alternatively For this k...

Page 5: ...erence File Close Close the program Mesy for Windows File Save Save the actual temperature profile with the actual name File Save as Save the actual temperature profile with a free definable name File Delete Close and delete the actual measurement File End program Close the program Mesy Type Choose the used MESY type The following choices are available USB MESY 570 70 USB MESY 570 77A Serial MESY ...

Page 6: ...cel file 9 Opens a window to print the actual measurement values 10 Opens a window to adjust the measurement parameters like measurement interval and the measurement time for real time measurement 11 Opens a window to load a reference profile 12 Opens a window to manage reference profiles 13 Extracts a reference profile out of a measured profile 14 Scales the actual measurement automatically so th...

Page 7: ...ou hear a long beep to start the measurement The measurement begins Measurement values will be recorded with the predefined interval of 0 5 sec onds and will be stored in the memory of the MESY After you have recorded all desired measurement values press the button Measure Start Stop You will hear two long beeps and the USB MESY stops the measurement An internal temperature of more than 60 C will ...

Page 8: ...example 25 C and 100 C 4 2 Adjustment of the measurement box Thermocouple sensors are available with different accuracies With this function you can adjust the USB MESY HID for each connected sensor This adjustment is a two point adjustment Temperatures mea sured with the MESY will be compared with a second calibrated temperature value Press the button On Off until you hear two long beeps to switc...

Page 9: ...ntered value and the measured value of the USB MESY now Follow the same procedure for the channels 2 and 3 Now the channels will be adjusted with the tempera ture 100 C Insert the ends of the thermocouple sensors of all 3 channels in the pot with the 100 C medium Start with channel 1 Select channel 1 in the adjustment menu Enter the value 1000 temperature in 1 10 C in the field Adjust High Press t...

Page 10: ...the inscription 0 5 600 s Sample to switch off the MESY Use a ball pen for example to press the button Follow the same procedure for the channels 2 and 3 Close the window Adjustment If you press the button Start a measurement will start With a correct adjustment you should see the graphs of all channels at 25 C or at 100 C end of thermocouple sensors must be in the according fluid ...

Page 11: ...Page 11 Quick reference guide USB MESY HID 570 77A Quick reference guide USB MESY HID 570 77A 10 2013 ...

Page 12: ... Quick reference guide USB MESY HID 570 77A 10 2013 For any question please contact us under Tel No 49 0 4136 909 0 zur Verfügung Your SEF Systec Team C SEF Systec GmbH October 2013 Subject to technical change ...