solutions@shockwatch.com.au 1300 074 625 shockwatch.com.au
May 2022
1. If the data logger has not been
stored at room temperature, place
it in room temperature for 30 mins
before starting.
2. Press and hold the START/STOP
button for approximately 5 seconds.
The green light flashes for 5 secs.
3. Place the data logger into
the location to be monitored.
Recording will begin after the start
delay period shown on the logger.
4. If there has been no temperature
excursion, the green light will
flash green every 10 seconds.
If a temperature excursion has
occurred, the red light will flash
every 10 seconds.
5. To stop recording, press and hold
the START/STOP button for 5
seconds. The red light flashes for 5
secs.
6. To generate a trip report, tear or cut
away the waterproof plastic bag. Be
careful not to damage the device
when cutting away the bag. Insert
the logger into a PC USB port. The
green and red lights will alternately
flash when the PDF report is being
generated.
7. When the PDF report has been
generated, the green and red lights
will illuminate solid until the logger
is removed from the USB port.
RC-17 Temperature Data Logger
User Manual
STATUS
LED INDICATION
Does not Start
Red and green light flash simultaneously
Start-up Delay
Red and green light blinks alternately once
Start - Normal
Green light blinks once ( flashes 10 times every minute)
Start - Alarm
Red light flashes once (flashes 10 times every minute)
Start - Mark
Red and green light flash 3 times at the same time
Start - Mark overrun
Red and green light flash 3 times alternately
Stop - Normal
Green light flashes twice
Stop - Alarm
Red light flashes twice
OPERATION
ACTION
LED INDICATION
START Recording
Press the START/STOP Button for 5 seconds
The RECORD LED will blink 5 times
STOP Recording
Press the START/STOP Button for 5 seconds
The STATUS LED will blink 5 times
Status Indication
The status of the recording can be analyzed by the unit’s indicator light.