Page 1: ...Hearing Systems Cellion primax Guide for Hearing Care Professionals ...

Page 2: ...tivating 4 Charging 5 Connecting 5 Noting the serial number 6 Disconnecting 6 Replacing the housing 7 Disassembling 7 Assembling 9 Important safety information 11 General warnings 11 Disposal 12 Further information 13 Storage periods 13 Symbols used in this document 13 ...

Page 3: ...als It provides important information on hearing instruments with permanently built in power cells lithium ion batteries CAUTION It is important to read this guide the user guide and the safety manual thoroughly Follow the safety information to avoid damage or injury ...

Page 4: ...iece to the hearing instruments The procedure is the same as for any other of our RIC hearing instruments Activating Before first use or first fitting you have to activate the hearing instruments by putting them into the charger Wait until the first orange LED lights up Note Hearing instruments that are activated and not used may discharge faster than hearing instruments that have not been activat...

Page 5: ...r to the client A full charging cycle takes approximately 4 hours Follow the operating conditions in both the hearing instrument s and the charger s user guide Connecting You can either fit the hearing instruments wirelessly with ConnexxLink or use the CS44 connection beneath the push button X Use the included tool to lift the push button Do not use sharp tools X Plug the CS44 cable as shown in th...

Page 6: ...atus the hearing instruments may switch off after being disconnected from the programming device This safety feature avoids deep discharge of the power cell To turn the hearing instruments back on X Put them into the powered charger for a moment After removal from the charger the hearing instruments are automatically turned on X Or press the push button for several seconds ...

Page 7: ...the Hearing Care Professional Disassembling X Remove the receiver unit and the axles X Insert the included tool into the front end of the hearing instrument and detach the bottom housing You may feel some resistance since the hearing instrument sits firmly in the bottom housing ...

Page 8: ...8 X Use the included tool to remove the top housing Do not use sharp tools X To remove the push button first remove the axle then lift the push button ...

Page 9: ... insert the axle X Install the top housing Ensure that on both sides of the hearing instrument the notches of the top housing match the marked lines The top housing clicks noticeably in place X Install the bottom housing from back first then push in the front ...

Page 10: ... axles X Check the joint between top and bottom housing It must be even and very narrow If it is uneven or wide the silicone wrapper may not sit correctly Disassemble the hearing instrument again and improve the position of the silicone wrapper X Attach the receiver unit ...

Page 11: ...le battery is permanently built into your hearing instrument Incorrect use of the power cell or the hearing instrument can cause the power cell to burst Risk of injury fire or explosion X Follow the safety instructions for power cells in this section Do not use the power cell in explosive atmospheres Keep each power cell sealed in the original packaging until it is ready to be used Do not use powe...

Page 12: ... a cool dry place with good ventilation Keep the power cells away from foodstuffs Keep the power cells out of reach of children Do not place the power cells in your mouth and do not swallow them Do not deep discharge the power cells Observe the operating transport and storage conditions in the hearing instrument s user guide Disposal X Do not dispose of power cells or hearing instruments in fire o...

Page 13: ...e The production date is printed on the power cell Day DD month MM and last digit of the year Y of production is printed on the power cell Example 21056XX 21st of May 2016 Once a power cell is activated in order for it to remain in good condition it needs to be charged at least every 6 months Symbols used in this document Points out a situation that could lead to serious moderate or minor injuries...

Page 14: ...14 ...

Page 15: ......

Page 16: ...0 99T03 7600 EN Order Item No 109 423 16 Master Rev04 12 2016 Signia GmbH 12 2016 Legal Manufacturer Signia GmbH Henri Dunant Strasse 100 91058 Erlangen Germany Phone 49 9131 308 0 Manufactured by Signia GmbH under Trademark License of Siemens AG ...

Reviews: